Tue Nov 24 15:35:49 2020 | 
Jan | Calibration | DSSD X/Y channel mapping | 2021 | 
During the detector test test measurements with alpha source the allocation of the 16 X- and 16 Y-channels has been checked. 
 
For the following final allocation, it is always assumed that the (horizontal) y-strips are placed to face the beam directly, while the (vertical) x-strips are on the backside. 
 
Now, all cables from the preamp to the ADC/TDCs are either labeled BLACK (= X-strips, pos. signals) or labeled RED (= Y-strips, neg. signals). These red or black connections should be kept consistently in order to ensure a well known orientation of the DSSD during the experiment. The test run Si2_run006.lmd was taken with this final assignment and serves as reference. 
 
BLACK LABEL > X1 to X16 > pos. MSCF > ADC/TDC ch  0-15 >   50 Ohm resistor at preamps HV-input 
RED LABEL   > Y1 to Y16 > neg. MSCF > ADC/TDC ch 16-31 > neg. bias voltage at preamps HV-input 
 
Additionally, the RED label indicates the section on the preamp to which negative bias voltage should be applied. | 
| 
  287  
 | 
Tue Feb  2 13:59:00 2021 | 
Jan | Detectors | DSSD installation and alignment | 2021 | 
The DSSD has been exchanged and aligned in 
November 2020. After bakeout at 140°C 
(externally, 120°C at internal temp. Sensor) 
for more than a week, the vacuum in the setup 
is roughly 4.5e-10 mbar.
After the bakeout the detector had to be 
realigned, it was lower by 1-2 mm. Using the 
line laser the realignment was done by 
touching only the screws on the far part of 
the base of the flange. These screws have been 
thightend by about a 1/4 turn after releasing 
the headless positioning screw accordingly.  | 
| 
  288  
 | 
Wed Mar 31 13:52:04 2021 | 
Jan | Detectors | BaF Na22 Testing | 2021 | 
On 30. and 31.03.2021 we did some testing with the 6 BaF detectors.
Trigger & threshold:
With a threshold at ~250 keV, each detector has an trigger rate of about 0.5kHz due to internal activity.
Energy spectra
All detectors showed the Na22 lines (511 + 1275 keV) and also the lines from internal activity.
The energy resolution was on the order of 8.5 - 11 keV, depending on the detector.
Time spectra
Raw TDC spectra showed self-stop peak.
TDC diff spectra (between detectors) show one main (sharp) correlation peak about 8ns = 30ch*0.293ns/ch off the zero. 
At very low intensity a time structure around this peak is visible in a region of 120ns around the main peak. This is not understood, the energy spectra look similar for all these time regions (above/on/below the main peak).
Additionally, the coincidence E-specta between detector 1 and 2 (gating on the main time peak), nicely showed only the 511 keV line from Na22 and some gras.
Overnight test run:
Det. 1, 2, 3, 4 with the Na22 source
lxg1275:/data.local3/e127/lmd_2021/test/
e127b_run0021.lmd
e127b_run0022.lmd
e127b_run0023.lmd
e127b_run0024.lmd
e127b_run0025.lmd
e127b_run0026.lmd
e127b_run0027.lmd
e127b_run0028.lmd
e127b_run0029.lmd
e127b_run0030.lmd
e127b_run0031.lmd
e127b_run0032.lmd
e127b_run0033.lmd
e127b_run0034.lmd
Gain matched HV settings: https://elog.gsi.de/esr/E127/289
Issues to solve:
- 2 missing HV channels
- Go4 Analysis not stable
- trigger not on Baf but on Si  | 
| 
  289  
 | 
Wed Mar 31 13:57:27 2021 | 
Jan | Detectors | BaF HV settings | 2021 | 
The setting below is a start value for final gain matching with a full ADC range of 12MeV: 
 
| 
Det.Nr. |  ADC/TDC ch  |  HV  |  MSCF Gain
 |  | 
1  |  20  |  2400V  |  6
 |  | 
2  |  21  |  1780V  |  6
 |  | 
3  |  22  |  2350V  |  6
 |  | 
4  |  23  |  2130V  |  6
 |  | 
5  |  24  |  2050V  |  6
 |  | 
6  |  25  |  1780V  |  6
 |  
  | 
| 
  290  
 | 
Wed Apr 14 09:42:50 2021 | 
Jan | Detectors | BaF HV gain matching | 2021 | 
The preliminary gain matching settings for a full ADC range of ~12MeV: 
 
| 
Det.Nr. |  ADC/TDC ch  |  HV  |  I  |  comment
 |  | 
1  |  20  |  2400V  |  390uA  |  CAEN HV ch0
 |  | 
2  |  21  |  1763V  |  287uA  |  CAEN HV ch1
 |  | 
3  |  22  |  2340V  |  380uA  |  CAEN HV ch2
 |  | 
4  |  23  |  2175V  |  354uA  |  CAEN HV ch3
 |  | 
5  |  24  |  2017.5V  |  ?  |  emetron HV 
 |  | 
6  |  25  |  1780V  |  ?  |  not yet biased
 |  
  
 
Attached is a 22Na-spectrum taken with detectors 1 - 5. 
The used MSCF settings are also attached. 
 
Detector 6 does not yet have a HV channel available. | 
| Attachment 1: BaF_Na22_gain_match_v1.png
 | 
 
 | 
| Attachment 2: baf_gain_match_v1.dat
 | 
atpnuc006:mrcc:mscf1:getGainCommon 15
atpnuc006:mrcc:mscf1:getShapingTimeCommon 3
atpnuc006:mrcc:mscf1:getThresholdCommon 6
atpnuc006:mrcc:mscf1:getPzCommon 6
atpnuc006:mrcc:mscf1:getGain1  6
atpnuc006:mrcc:mscf1:getGain2  6
atpnuc006:mrcc:mscf1:getGain3  6
atpnuc006:mrcc:mscf1:getGain4  6
atpnuc006:mrcc:mscf1:getShapingTime1 3
atpnuc006:mrcc:mscf1:getShapingTime2 3
atpnuc006:mrcc:mscf1:getShapingTime3 3
atpnuc006:mrcc:mscf1:getShapingTime4 3
atpnuc006:mrcc:mscf1:getThreshold1 6
atpnuc006:mrcc:mscf1:getThreshold2 6
atpnuc006:mrcc:mscf1:getThreshold3 6
atpnuc006:mrcc:mscf1:getThreshold4 6
atpnuc006:mrcc:mscf1:getThreshold5 6
atpnuc006:mrcc:mscf1:getThreshold6 6
atpnuc006:mrcc:mscf1:getThreshold7 6
atpnuc006:mrcc:mscf1:getThreshold8 6
atpnuc006:mrcc:mscf1:getThreshold9 6
atpnuc006:mrcc:mscf1:getThreshold10 6
atpnuc006:mrcc:mscf1:getThreshold11 6
atpnuc006:mrcc:mscf1:getThreshold12 6
atpnuc006:mrcc:mscf1:getThreshold13 6
atpnuc006:mrcc:mscf1:getThreshold14 6
atpnuc006:mrcc:mscf1:getThreshold15 6
atpnuc006:mrcc:mscf1:getThreshold16 6
atpnuc006:mrcc:mscf1:getPz1    6
atpnuc006:mrcc:mscf1:getPz2    3
atpnuc006:mrcc:mscf1:getPz3    3
atpnuc006:mrcc:mscf1:getPz4    1
atpnuc006:mrcc:mscf1:getPz5    5
atpnuc006:mrcc:mscf1:getPz6    5
atpnuc006:mrcc:mscf1:getPz7    6
atpnuc006:mrcc:mscf1:getPz8    6
atpnuc006:mrcc:mscf1:getPz9    6
atpnuc006:mrcc:mscf1:getPz10   6
atpnuc006:mrcc:mscf1:getPz11   6
atpnuc006:mrcc:mscf1:getPz12   6
atpnuc006:mrcc:mscf1:getPz13   6
atpnuc006:mrcc:mscf1:getPz14   6
atpnuc006:mrcc:mscf1:getPz15   6
atpnuc006:mrcc:mscf1:getPz16   6
atpnuc006:mrcc:mscf1:getSingleChMode 1
atpnuc006:mrcc:mscf1:getRcMode 1
atpnuc006:mrcc:mscf1:getAutoPZ 0
atpnuc006:mrcc:mscf1:getMultiplicityHi 8
atpnuc006:mrcc:mscf1:getMultiplicityLo 1
atpnuc006:mrcc:mscf1:getSumTrgThresh 100
atpnuc006:mrcc:mscf1:getBlrOn  0
atpnuc006:mrcc:mscf1:getCoincTime 100
atpnuc006:mrcc:mscf1:getThreshOffset 1
atpnuc006:mrcc:mscf1:getShaperOffset 100
atpnuc006:mrcc:mscf1:getBlrThresh 10
atpnuc006:mrcc:mscf1:getECLDelay 0
 
 | 
| 
  291  
 | 
Thu Apr 22 10:56:20 2021 | 
Jan | Detectors | target-chamber det. distances | 2021 | 
This is the lookup table for the distance of each x-ray detector to the target. 
See attached foto for explanation. 
 
It is the same values as in 2020: https://elog.gsi.de/esr/E127/36  
 
| 
Detector  |  A (dA) [mm]  |  B (dB) [mm]  |  C (dC) [mm]  |  calculate  |  result [mm]
 |  | 
35°  |  136.0 (1.0)  |  20.0 (0.2)  |  450.0 (0.5)  |  C + B - A  |  334.0 (?)
 |  | 
90°  |  76.0 (0.5)  |  20.0 (0.2)  |  447.0/2 = 223.5 (1.0)  |  C + B - A  |  167.5 (?)
 |  | 
145°  |  165.0 (1.0)  |  20.0 (0.2)  |  450.0 (0.5)  |  C + B - A  |  305.0 (?)
 |  
  
Errors are estimated after measurement, final errorbars have to be double checked! | 
| Attachment 1: E127_xray_distances.jpg
 | 
 
 | 
| 
  361  
 | 
Fri May 21 10:00:51 2021 | 
Jan | Accelerator | target bump | 2021 | 
The target overlap was optimized on xray and si rates.
The setting now is -10mm in SC9  | 
| 
  363  
 | 
Fri May 21 11:55:24 2021 | 
Jan | Detectors | SI MSCF settings v1 | 2021 | 
attached are the settings for all MSCFs.
Settings are optimized for DSSD only, not BaF!  | 
| Attachment 1: setting_test.1621590878
 | 
1621590878
Fri May 21 11:54:38 CEST 2021
e127pi:mrcc:mscf1:getGainCommon 5
e127pi:mrcc:mscf1:getShapingTimeCommon 3
e127pi:mrcc:mscf1:getThresholdCommon 8
e127pi:mrcc:mscf1:getPzCommon  100
e127pi:mrcc:mscf2:getGainCommon 5
e127pi:mrcc:mscf2:getShapingTimeCommon 3
e127pi:mrcc:mscf2:getThresholdCommon 8
e127pi:mrcc:mscf2:getPzCommon  100
e127pi:mrcc:mscf1:getGain1     4
e127pi:mrcc:mscf1:getGain2     4
e127pi:mrcc:mscf1:getGain3     4
e127pi:mrcc:mscf1:getGain4     4
e127pi:mrcc:mscf1:getShapingTime1 1
e127pi:mrcc:mscf1:getShapingTime2 1
e127pi:mrcc:mscf1:getShapingTime3 1
e127pi:mrcc:mscf1:getShapingTime4 1
e127pi:mrcc:mscf2:getGain1     9
e127pi:mrcc:mscf2:getGain2     4
e127pi:mrcc:mscf2:getGain3     8
e127pi:mrcc:mscf2:getGain4     8
e127pi:mrcc:mscf2:getShapingTime1 2
e127pi:mrcc:mscf2:getShapingTime2 2
e127pi:mrcc:mscf2:getShapingTime3 2
e127pi:mrcc:mscf2:getShapingTime4 2
e127pi:mrcc:mscf1:getThreshold1 4
e127pi:mrcc:mscf1:getThreshold2 4
e127pi:mrcc:mscf1:getThreshold3 4
e127pi:mrcc:mscf1:getThreshold4 1
e127pi:mrcc:mscf1:getThreshold5 4
e127pi:mrcc:mscf1:getThreshold6 4
e127pi:mrcc:mscf1:getThreshold7 4
e127pi:mrcc:mscf1:getThreshold8 4
e127pi:mrcc:mscf1:getThreshold9 4
e127pi:mrcc:mscf1:getThreshold10 5
e127pi:mrcc:mscf1:getThreshold11 5
e127pi:mrcc:mscf1:getThreshold12 5
e127pi:mrcc:mscf1:getThreshold13 3
e127pi:mrcc:mscf1:getThreshold14 4
e127pi:mrcc:mscf1:getThreshold15 5
e127pi:mrcc:mscf1:getThreshold16 5
e127pi:mrcc:mscf1:getPz1       128
e127pi:mrcc:mscf1:getPz2       128
e127pi:mrcc:mscf1:getPz3       128
e127pi:mrcc:mscf1:getPz4       128
e127pi:mrcc:mscf1:getPz5       128
e127pi:mrcc:mscf1:getPz6       128
e127pi:mrcc:mscf1:getPz7       128
e127pi:mrcc:mscf1:getPz8       128
e127pi:mrcc:mscf1:getPz9       128
e127pi:mrcc:mscf1:getPz10      128
e127pi:mrcc:mscf1:getPz11      128
e127pi:mrcc:mscf1:getPz12      128
e127pi:mrcc:mscf1:getPz13      128
e127pi:mrcc:mscf1:getPz14      128
e127pi:mrcc:mscf1:getPz15      128
e127pi:mrcc:mscf1:getPz16      128
e127pi:mrcc:mscf2:getThreshold1 255
e127pi:mrcc:mscf2:getThreshold2 255
e127pi:mrcc:mscf2:getThreshold3 0
e127pi:mrcc:mscf2:getThreshold4 255
e127pi:mrcc:mscf2:getThreshold5 255
e127pi:mrcc:mscf2:getThreshold6 255
e127pi:mrcc:mscf2:getThreshold7 255
e127pi:mrcc:mscf2:getThreshold8 255
e127pi:mrcc:mscf2:getThreshold9 255
e127pi:mrcc:mscf2:getThreshold10 255
e127pi:mrcc:mscf2:getThreshold11 0
e127pi:mrcc:mscf2:getThreshold12 255
e127pi:mrcc:mscf2:getThreshold13 255
e127pi:mrcc:mscf2:getThreshold14 255
e127pi:mrcc:mscf2:getThreshold15 0
e127pi:mrcc:mscf2:getThreshold16 255
e127pi:mrcc:mscf2:getPz1       128
e127pi:mrcc:mscf2:getPz2       128
e127pi:mrcc:mscf2:getPz3       87
e127pi:mrcc:mscf2:getPz4       255
e127pi:mrcc:mscf2:getPz5       255
e127pi:mrcc:mscf2:getPz6       255
e127pi:mrcc:mscf2:getPz7       255
e127pi:mrcc:mscf2:getPz8       255
e127pi:mrcc:mscf2:getPz9       255
e127pi:mrcc:mscf2:getPz10      255
e127pi:mrcc:mscf2:getPz11      85
e127pi:mrcc:mscf2:getPz12      255
e127pi:mrcc:mscf2:getPz13      255
e127pi:mrcc:mscf2:getPz14      255
e127pi:mrcc:mscf2:getPz15      85
e127pi:mrcc:mscf2:getPz16      255
e127pi:mrcc:mscf1:getSingleChMode 0
e127pi:mrcc:mscf2:getSingleChMode 0
e127pi:mrcc:mscf1:getRcMode    1
e127pi:mrcc:mscf2:getRcMode    1
e127pi:mrcc:mscf1:getAutoPZ    0
e127pi:mrcc:mscf2:getAutoPZ    0
e127pi:mrcc:mscf1:getMultiplicityHi 8
e127pi:mrcc:mscf2:getMultiplicityHi 8
e127pi:mrcc:mscf1:getMultiplicityLo 4
e127pi:mrcc:mscf2:getMultiplicityLo 4
e127pi:mrcc:mscf1:getSumTrgThresh 10
e127pi:mrcc:mscf2:getSumTrgThresh 10
e127pi:mrcc:mscf1:getBlrOn     1
e127pi:mrcc:mscf2:getBlrOn     1
e127pi:mrcc:mscf1:getCoincTime 128
e127pi:mrcc:mscf2:getCoincTime 128
e127pi:mrcc:mscf1:getThreshOffset 100
e127pi:mrcc:mscf2:getThreshOffset 100
e127pi:mrcc:mscf1:getShaperOffset 100
e127pi:mrcc:mscf2:getShaperOffset 100
e127pi:mrcc:mscf1:getBlrThresh 25
e127pi:mrcc:mscf2:getBlrThresh 25
e127pi:mrcc:mscf1:getECLDelay  0
e127pi:mrcc:mscf2:getECLDelay  0
atpnuc006:mrcc:mscf1:getGainCommon 15
atpnuc006:mrcc:mscf1:getShapingTimeCommon 3
atpnuc006:mrcc:mscf1:getThresholdCommon 6
atpnuc006:mrcc:mscf1:getPzCommon 6
atpnuc006:mrcc:mscf1:getGain1  6
atpnuc006:mrcc:mscf1:getGain2  6
atpnuc006:mrcc:mscf1:getGain3  6
atpnuc006:mrcc:mscf1:getGain4  6
atpnuc006:mrcc:mscf1:getShapingTime1 3
atpnuc006:mrcc:mscf1:getShapingTime2 3
atpnuc006:mrcc:mscf1:getShapingTime3 3
atpnuc006:mrcc:mscf1:getShapingTime4 3
atpnuc006:mrcc:mscf1:getThreshold1 6
atpnuc006:mrcc:mscf1:getThreshold2 6
atpnuc006:mrcc:mscf1:getThreshold3 6
atpnuc006:mrcc:mscf1:getThreshold4 6
atpnuc006:mrcc:mscf1:getThreshold5 6
atpnuc006:mrcc:mscf1:getThreshold6 6
atpnuc006:mrcc:mscf1:getThreshold7 6
atpnuc006:mrcc:mscf1:getThreshold8 6
atpnuc006:mrcc:mscf1:getThreshold9 6
atpnuc006:mrcc:mscf1:getThreshold10 6
atpnuc006:mrcc:mscf1:getThreshold11 6
atpnuc006:mrcc:mscf1:getThreshold12 6
atpnuc006:mrcc:mscf1:getThreshold13 6
atpnuc006:mrcc:mscf1:getThreshold14 6
atpnuc006:mrcc:mscf1:getThreshold15 6
atpnuc006:mrcc:mscf1:getThreshold16 6
atpnuc006:mrcc:mscf1:getPz1    6
atpnuc006:mrcc:mscf1:getPz2    3
atpnuc006:mrcc:mscf1:getPz3    3
atpnuc006:mrcc:mscf1:getPz4    1
atpnuc006:mrcc:mscf1:getPz5    35
atpnuc006:mrcc:mscf1:getPz6    35
atpnuc006:mrcc:mscf1:getPz7    35
atpnuc006:mrcc:mscf1:getPz8    35
atpnuc006:mrcc:mscf1:getPz9    35
atpnuc006:mrcc:mscf1:getPz10   35
atpnuc006:mrcc:mscf1:getPz11   6
atpnuc006:mrcc:mscf1:getPz12   6
atpnuc006:mrcc:mscf1:getPz13   6
atpnuc006:mrcc:mscf1:getPz14   6
atpnuc006:mrcc:mscf1:getPz15   6
atpnuc006:mrcc:mscf1:getPz16   6
atpnuc006:mrcc:mscf1:getSingleChMode 0
atpnuc006:mrcc:mscf1:getRcMode 1
atpnuc006:mrcc:mscf1:getAutoPZ 0
atpnuc006:mrcc:mscf1:getMultiplicityHi 8
atpnuc006:mrcc:mscf1:getMultiplicityLo 1
atpnuc006:mrcc:mscf1:getSumTrgThresh 100
atpnuc006:mrcc:mscf1:getBlrOn  0
atpnuc006:mrcc:mscf1:getCoincTime 100
atpnuc006:mrcc:mscf1:getThreshOffset 1
atpnuc006:mrcc:mscf1:getShaperOffset 100
atpnuc006:mrcc:mscf1:getBlrThresh 10
atpnuc006:mrcc:mscf1:getECLDelay 0
 
 | 
| 
  364  
 | 
Fri May 21 12:21:57 2021 | 
Jan | Runs | Run 0053 | 2021 | 
open file : e127b_run0053.lmd 
 
beam: 124Xe54+ 
energy: 7 MeV/u 
cooler voltage: 
 
purpose:  
Illumination of the detector to see if the gain settings are ok. 
Spectra look reasonable, rates are low. 
 
Detector position (Si): -25 mm 
GE01DD1IG position (Scraper): out/not used 
 
run start: 12:21:34 21-05-2021 
run stop:  14:05:27 21-05-2021 
 
_______________________________________________________________________________ 
 
ON Rates 
Si_bdt:     15 Hz 
Xray_35:    20 Hz 
Xray_90:    200 Hz 
Xray_145:   70 Hz 
BaF_OR:     --- 
 
Si voltage:          100.0 V   
Si leakage current:  360 uA 
 
 
SIS particles before:            
ESR current at injection:       3e7 
ESR particles after decel.:     5e6 
Target ON density:              9e13 | 
| 
  365  
 | 
Fri May 21 14:06:23 2021 | 
Jan | Runs | Run 0054 | 2021 | 
open file : e127b_run0054.lmd 
 
beam: 124Xe54+ 
energy: 7 MeV/u 
cooler voltage: 
 
purpose:  
Illumination of the detector to see if the gain settings are ok. 
Spectra look reasonable, rates are low. 
 
Detector position (Si): -25 mm 
GE01DD1IG position (Scraper): out/not used 
 
run start: 14:06:04 21-05-2021 
run stop:  15:14:48 21-05-2021 
 
_______________________________________________________________________________ 
 
ON Rates 
Si_bdt:     31 Hz 
Xray_35:    145 Hz 
Xray_90:    10 Hz 
Xray_145:   56 Hz 
BaF_OR:     --- 
 
Si voltage:          100.0 V   
Si leakage current:  350 uA 
 
 
SIS particles before:            
ESR current at injection:       3e7 
ESR particles after decel.:     5e6 
Target ON density:              9e13 | 
| 
  366  
 | 
Fri May 21 15:15:59 2021 | 
Jan | Runs | Run 0055 | 2021 | 
open file : e127b_run0055.lmd 
 
beam: 124Xe54+ 
energy: 7 MeV/u 
cooler voltage: 
 
purpose:  
Illumination of the detector to see if the gain settings are ok. 
Spectra look reasonable, rates are low. 
 
Detector position (Si): -25 mm 
GE01DD1IG position (Scraper): out/not used 
 
run start: 15:14:48 21-05-2021 
run stop:  15:10:10 21-05-2021 
 
_______________________________________________________________________________ 
 
ON Rates 
Si_bdt:     21 Hz 
Xray_35:    143 Hz 
Xray_90:    7 Hz 
Xray_145:   63 Hz 
BaF_OR:     --- 
 
Si voltage:          100.0 V   
Si leakage current:  350 uA 
 
 
SIS particles before:            
ESR current at injection:        
ESR particles after decel.:      
Target ON density:              10e13 | 
| 
  369  
 | 
Fri May 21 19:36:46 2021 | 
Jan | Runs | Run 0056 | 2021 | 
open file : e127b_run0056.lmd 
 
beam: 124Xe54+ 
energy: 7 MeV/u 
cooler voltage: 
 
purpose:  
Illumination of the detector to see if the gain settings are ok. 
Spectra look reasonable, rates are low. 
 
Detector position (Si): -25 mm 
GE01DD1IG position (Scraper): out/not used 
 
run start: 19:19:52 21-05-2021 
run stop:  20:08:08 21-05-2021 
 
_______________________________________________________________________________ 
 
ON Rates 
Si_bdt:     81 Hz 
Xray_35:    514 Hz 
Xray_90:    20 Hz 
Xray_145:   127 Hz 
BaF_OR:     --- 
 
Si voltage:          100.0 V   
Si leakage current:  340 uA 
 
 
SIS particles before: 9e7           
ESR current at injection:  4e7      
ESR particles after decel.:   4e6   
Target ON density:              9e13 | 
| 
  370  
 | 
Fri May 21 20:09:00 2021 | 
Jan | Runs | Run 0057 | 2021 | 
open file : e127b_run0057.lmd 
 
beam: 124Xe54+ 
energy: 7 MeV/u 
cooler voltage: 
 
purpose:  
Illumination of the detector to see if the gain settings are ok. 
Spectra look reasonable, rates are low. 
 
Detector position (Si): -25 mm 
GE01DD1IG position (Scraper): out/not used 
 
run start: 20:08:08 21-05-2021 
run stop:  21:01:58 21-05-2021 
_______________________________________________________________________________ 
 
ON Rates 
Si_bdt:     73 Hz 
Xray_35:    584 Hz 
Xray_90:    20 Hz 
Xray_145:   130 Hz 
BaF_OR:     --- 
 
Si voltage:          100.0 V   
Si leakage current:  340 uA 
 
 
SIS particles before: 9e7           
ESR current at injection:  2.5e7      
ESR particles after decel.:   4e6   
Target ON density:              9e13 | 
| 
  371  
 | 
Fri May 21 21:02:53 2021 | 
Jan | Runs | Run 0058 | 2021 | 
open file : e127b_run0058.lmd 
 
beam: 124Xe54+ 
energy: 7 MeV/u 
cooler voltage: 
 
purpose:  
Illumination of the detector to see if the gain settings are ok. 
Spectra look reasonable, rates are low. 
 
Detector position (Si): -25 mm 
GE01DD1IG position (Scraper): out/not used 
 
run start: 21:01:58 21-05-2021 
run stop:  22:00:05 21-05-2021 
_______________________________________________________________________________ 
 
ON Rates 
Si_bdt:     94 Hz 
Xray_35:    581 Hz 
Xray_90:    15 Hz 
Xray_145:   123 Hz 
BaF_OR:     --- 
 
Si voltage:          100.0 V   
Si leakage current:  340 uA 
 
 
SIS particles before: 7.4e7           
ESR current at injection:  2.8e7      
ESR particles after decel.:   2.5e6   
Target ON density:              9.8e13 | 
| 
  412  
 | 
Sun May 23 23:44:37 2021 | 
Jan | Runs | Run 0088 - background BaF/Xray | 2021 | 
open file : e127b_run0088.lmd 
 
no beam 
 
purpose:  
BG  measurement for Xray and BaF 
 
run start: 23:44:09 23.05.2021 
run stop:   
_______________________________________________________________________________ 
 
ON Rates 
Si_OFF_bdt:     Hz 
Xray_ON_bdt:  60Hz 
BaF_OR:     3000Hz | 
| 
  438  
 | 
Tue May 25 07:51:48 2021 | 
Jan | Accelerator | ESR stacking cycle length | 2021 | 
The full cycle length of ESR with 30 stacks is about 7:45 minutes, out of which 15.5 sec are with target switched ON.
This is without CMB/HTD in parallel.  | 
| 
  439  
 | 
Tue May 25 08:36:15 2021 | 
Jan | General | lessons learned | 2021 | 
1. before using stoch. cooling in ESR, boot and update the windows PC
2. better diagnostics for low energy/intensity beam-target overlapp in ESR is needed (either a clever analysis or a new detector)
more to come  | 
| 
  445  
 | 
Tue May 25 11:58:30 2021 | 
Jan | Analysis | 118(p,g) counts until run0115 | 2021 | 
The two attached plots show the collected Si data up to run0115.
Within this ~15h we have about ~70 counts in the peak region.
This translates to ~4.5 counts per hour.
And shall be compared to the 1 count/hour we had last year.  | 
| Attachment 1: run91to115_Si_map.png
 | 
 
 | 
| Attachment 2: run91to115_Si_projection.png
 | 
 
 | 
| 
  469  
 | 
Wed May 26 14:28:23 2021 | 
Jan | Analysis | Xrays from 118Te & 124Xe | 2021 | 
Attached is a plot of the uncalibrated energy spectra taken with the 90 degree xray detector for 118Te (red) and 124Xe (blue).
Shown is the region covering the K-alpha and K-REC peaks.
Calculating the ratios (Te/Xe) for K_alpha and K-REC energies from theory one gets:
K_alpha (theory): 0.9236  
K-REC   (theory): 0.9308
Without energy calibration we get the following ratio (using channel numbers):
K_alpha (exp): 0.9293  
K-REC   (exp): 0.9344
Taking into account that the channel numbers are not calibrated, the agreement is still very good.
In summary we can be sure that we have 118Te stored at 7AMeV.  | 
| Attachment 1: xray90_124Xe_118Te.png
 | 
 
 | 
| 
  496  
 | 
Thu May 27 13:26:35 2021 | 
Jan | Accelerator | FRS degrader failure | 2021 | 
Maybe, our degrader in FRS was removed by hand at (~13:40 26/05/21)
It is put back now (~13:10 27/05/21)
All runs taken in this period have to be double checked!
However, the xray lines from this night show Te for k_alpha and K-KREC
It is also possible that the readout of this degrader motor was just wrong due to its state as hand-controlled.  |