Fri May 14 13:58:52 2021, Jan Glorius, Calibration, run0044 - Xray145 calib 210Pb d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
90deg TFA diff set from 50 to 200.
Also threshold is slightly increased.
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 13:58:33 - 14.05.2021
Stop time: 16:46:00 - 14.05.2021
file name: e127b_run0044.lmd
avrg. rate: 40Hz
dead-time: 0%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Fri May 14 11:17:48 2021, Jan Glorius, Calibration, run0043 - Xray145 calib 210Pb d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Threshold slightly increased
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 11:17:23 - 14.05.2021
Stop time: 13:47:42 - 14.05.2021
file name: e127b_run0043.lmd
avrg. rate: 40Hz
dead-time: 0%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Fri May 14 10:28:49 2021, Jan Glorius, Calibration, run0042 - Xray145 calib 241Am_hi d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Threshold slightly increased
Detector: GEM1800 - 145 deg
Source: 241Am_hi
Distance: 305mm
Start time: 10:27:58 - 14.05.2021
Stop time: 11:14:03 - 14.05.2021
file name: e127b_run0042.lmd
avrg. rate: 200Hz
dead-time: 1.5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Fri May 14 09:35:13 2021, Jan Glorius, Calibration, run0041 - Xray145 calib Cd109 d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Threshold slightly increased
Detector: GEM1800 - 145 deg
Source: 109Cd
Distance: 305mm
Start time: 09:34:52 - 14.05.2021
Stop time: 09:47:42 - 14.05.2021
file name: e127b_run0041.lmd
avrg. rate: 600Hz
dead-time: 5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Fri May 14 09:06:55 2021, Jan Glorius, Calibration, run0040 - Xray145 calib Ba133_hi d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Threshold slightly increased
Detector: GEM1800 - 145 deg
Source: 133Ba_hi
Distance: 305mm
Start time: 09:06:33 - 14.05.2021
Stop time: 09:31:32 - 14.05.2021
file name: e127b_run0040.lmd
avrg. rate: 850Hz
dead-time: 6%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 21:22:31 2021, Jan Glorius, Calibration, run0039 - Xray145 calib Pb210 d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Rate is increasing slowly 200 > 400 Hz, more noise coming in.
Also possible source holder movement!
Resolution still good.
Take this run with care!
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 21:22:00 - 13.05.2021
Stop time: 08:56:31 - 14.05.2021
file name: e127b_run0039.lmd
avrg. rate: 220Hz > 400Hz !!!!
dead-time: 3%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 14:16:58 2021, Jan Glorius, Calibration, run0038 - Xray145 calib Pb210 d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Rate is increasing slowly, from ~70 to ~220 Hz in ~7 hours.
Also the threshold seems to be lower now.
Also possible source holder movement!
Resolution still good.
Take this run with care!
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 14:16:18 - 13.05.2021
Stop time: 21:09:12 - 13.05.2021
file name: e127b_run0038.lmd
avrg. rate: 70Hz > 220Hz!!!!
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 13:24:44 2021, Jan Glorius, Calibration, run0037 - Xray90 calib Pb210 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 13:24:26 - 13.05.2021
Stop time: 14:06:35 - 13.05.2021
file name: e127b_run0037.lmd
avrg. rate: 70Hz
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 12:21:41 2021, Jan Glorius, Calibration, run0036 - Xray90 calib Ba133_lo d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 133Ba_lo
Distance: 167.5mm
Start time: 12:21:21 - 13.05.2021
Stop time: 13:22:04 - 13.05.2021
file name: e127b_run0036.lmd
avrg. rate: 300Hz
dead-time: 2%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 12:06:51 2021, Jan Glorius, Calibration, run0035 - Xray90 calib Cd109 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 109Cd
Distance: 167.5mm
Start time: 12:06:24 - 13.05.2021
Stop time: 12:16:46 - 13.05.2021
file name: e127b_run0035.lmd
avrg. rate: 1800Hz
dead-time: 12%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 11:51:42 2021, Jan Glorius, Calibration, run0034 - Xray90 calib Am241_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 11:49:53 - 13.05.2021
Stop time: 12:02:24 - 13.05.2021
file name: e127b_run0034.lmd
avrg. rate: 600Hz
dead-time: 5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 15:28:55 2021, Jan Glorius, Calibration, run0033 - Xray145 calib Cd109 d=305mm, 2021
|
Efficiency calibration in the lab
Detector: GEM3000 - 145 deg
Source: 109Cd
Distance: 305mm
Start time: 15:28:01 - 12.05.2021
Stop time: 16:01:57 - 12.05.2021
file name: e127b_run0033.lmd
avrg. rate: 600 Hz
dead-time: 5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 14:55:04 2021, Jan Glorius, Calibration, run0032 - Xray90 calib Cd109 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 109Cd
Distance: 167.5mm
Start time: 14:54:36 - 12.05.2021
Stop time: 15:10:18- 12.05.2021
file name: e127b_run0032.lmd
avrg. rate: 1800Hz
dead-time: 12%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 14:16:27 2021, Jan Glorius, Calibration, run0031 - Xray90 calib Am241_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 14:16:07 - 12.05.2021
Stop time: 14:34:00- 12.05.2021
file name: e127b_run0031.lmd
avrg. rate: 600Hz
dead-time: 4%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 13:36:24 2021, Jan Glorius, Calibration, run0030 - Xray90 calib Am241_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
AMPLIFIER JUMPED!
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 13:35:42 - 12.05.2021
Stop time: 14:05:08 - 12.05.2021
file name: e127b_run0030.lmd
avrg. rate: 600Hz
dead-time: 4%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 11:47:41 2021, Jan Glorius, Calibration, run0029 - Xray90 calib Pb210 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 11:46:19 - 12.05.2021
Stop time: 13:25:46 - 12.05.2021
file name: e127b_run0029.lmd
avrg. rate: 50Hz
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 11:29:23 2021, Jan Glorius, Calibration, run0028 - Xray90 calib Ba133_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 133Ba_hi
Distance: 167.5mm
Start time: 11:32:50 - 12.05.2021
Stop time: 11:44:29 - 12.05.2021
file name: e127b_run0028.lmd
avrg. rate: 2500Hz
dead-time: 17%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 11:34:11 2021, Jan Glorius, Calibration, run0027 - crap, 2021
|
|
Wed May 12 10:16:10 2021, Jan Glorius, Calibration, run0026 - Xray145 calib Ba133_hi d=305mm, 2021
|
Efficiency calibration in the lab
Detector: GEM3000 - 145 deg
Source: 133Ba_hi
Distance: 305mm
Start time: 10:15:51 - 12.05.2021
Stop time: 10:59:35 - 12.05.2021
file name: e127b_run0026.lmd
avrg. rate: 850 Hz
dead-time: 5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Tue May 11 16:49:23 2021, Jan Glorius, Calibration, run0025 - Xray145 calib Pb210 d=305mm, 2021
|
Efficiency calibration in the lab
Detector: GEM3000 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 16:49:03 - 11.05.2021
Stop time: 09:21:13 - 12.05.2021
file name: e127b_run0025.lmd
avrg. rate: 50 Hz
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off.
Check for rate spikes |