  107   Wed Jun 1 08:39:50 2022 Kristian, KonstantinCalibrationWavemeter readingWe recorded files with high wavemeter readout
rate and saved it at the atplaser network
drive (Q:) - Bismuth - wavemeter folder.
  
  5   Tue Apr 26 16:28:14 2022 CarstenDAQCable of Bunch HF disconnected and 1.995Mhz signal connected for DAQ tests see subject.   
  8   Wed May 4 22:45:12 2022 Carsten, SebastianDAQQ-Switch Signal (ch15) fixed The "dead" Q-Switch signal was converted 
two times... This explains why in the 
Thorium experiment the Q-Switch signal 
  
  12   Fri May 6 10:30:23 2022 SebastianDAQlaser busy / dye out signal fixedSolution for a working laser busy / dye out
signal was to disable the FCU and set the
FCU status to "False" (see attached picture).
 CobraFix_LaserBusy.PNG 
  13   Fri May 6 10:35:06 2022 SebastianDAQlaser step complete cabling changedI changed the laser step complete cabling
from a quad coinc (see attached picture)
to a Logic Fan In / Fan out (same as laser
 stepComplete_beforeChange_QuadCoincCabling.PNG 
  20   Mon May 9 17:01:02 2022 CarstenDAQdirectories / paths / comments / reduction factors /startup.scom files for the 2 DAQs [U][B]DAQ1 (original):[/B][/U]
-user atplaser
-runs on lxg1050
  
  21   Mon May 9 17:14:02 2022 CarstenDAQNTCAP DAQThe NTCap DAQ computer in the HKR (Schottky
logbook) is started, and all preparations
for data taking (mounting of NAS etc.) are
  
  23   Mon May 9 19:05:17 2022 WilfriedDAQQ-Switch signals + 0.5 ms and +1 ms in VUPROMWe have generated two additional signals which
are fed into two VUPROM channels:
Q-Switch + 0.5 ms --> channel 9
  
  31   Tue May 10 22:29:45 2022 Jan GloriusDAQDAQ2 shaper settingsAttached are the MSCF shaper settings applied
on Tue May 10 at 22:30.
 setting_e128.v1 
  32   Tue May 10 22:37:10 2022 Jan GloriusDAQDAQ2 cabling docu   e128_daq2_cabling.ods 
  43   Wed May 11 14:48:31 2022 KonstantinDAQVoltage range of the Keysight DMM changed from 10V to 1VDuring the setup of the ion beam the voltage
range of the Keysight DMM was changed from
10V to 1V. Additionally, the name of the
 VoltageRangeChanged.png 
  46   Wed May 11 18:13:04 2022 Wilfried, Phillip, MaxDAQlaser scan/step completeThe scaler channel laser scan complete and
laser step complete is interchanged in go4.
Carsten will fix this on the software side.
  
  52   Wed May 11 23:36:24 2022 Ruijiu Chen, Carsten, GuyDAQstart NTCAP.Carrier frequency:244MHz
IQ rate: 10M Samples/seconds
Reference level: -25
  
  54   Thu May 12 03:25:37 2022 PatrickDAQLaser scan complete signal offAs in ID:46, the 'laser scan complete' again
is not coming in.
It is not very useful to repeatedly fix cable
  
  55   Thu May 12 10:10:35 2022 Simon, Christopher, Wilfried, SebastianDAQ1ms gated PMT north / middle added to DAQWe added on DAQ1 and DAQ2 1ms gated PMT North
and PMT Middle signals.
Now we "supress" the signal by a factor of
  
  61   Thu May 12 14:04:09 2022 KonstantinDAQGo4 Laserstep histograms adaptedIn order to correctly display the gated PMT
signals [I]PMT Middle[/I] and [I]PMT North[/I]
some modifications in the Go4 code were necessary.
 Go4Modifications.PNG 
  63   Thu May 12 15:34:47 2022 Max, Kristian, SebastianDAQlaser scan, laser step and laser busy checkedWe checked all three signals in the Messhütte.
Laser scan and laser step are OK.
The Laser busy signal is for large increments
  
  76   Mon May 16 22:19:54 2022 PatrickDAQKeep go4 running (in theory), MBSTo accumulate more data into the opened go4,
do not restart MBS, but just
  
  78   Tue May 17 00:44:01 2022 Patrick, KristianDAQChanged PMT North CFD channelIn run 8134 - 8135 PMT North showed a super
high background rate. After the scan we checked
the signal cables.
  
  81   Fri May 27 10:39:53 2022 KristianDAQPMT North noiseDuring the last run, PMT North showed some