ID |
Date |
Author |
Category |
Subject |
11
|
Thu May 5 15:24:34 2022 |
Sebastian | Laser | laser busy problem |
Current status:
Laser busy signal works correctly (5V if the grating moves, 0V if not) if the laser is operated by the Sirah software.
When the laser is controled by the "Thorium Main VI" the laser busy signal is always 0V. |
15
|
Fri May 6 16:41:57 2022 |
Patrick, Rodolfo | Laser | Wavemeter |
We coupled the laser through a multi-mode fiber into the wavemeter.
With the main amplifier enabled, we got enough power behind the fiber (0.5 - 1.5 mW) to use the wavemeter with 34 ms exposure time (1s / 30Hz to prevent jumps in the wavemeter readout).
Pictures below:
1: First test scan with 34ms exp. time.
2: Comp. too large laser power (2. scan duoblet) vs. attenuated power (right scan doublet). The latter has less hickups.
3: Laser scan with attenuated power in go4.
The measured frequencies in the first scan step are unstable in any case.
Spikes (rather "second wavelength the laser is regulrarly jumping to") can be removed by adjusting the fiber coupling with the mirrors (change of interference pattern) or the L/2-waveplate (change of coupled power).
However, during (and after) the scan the spikes seem to reappear. |
Attachment 1: Screenshot_(4).png
|
|
Attachment 2: Bi2022_laserscan_test.PNG
|
|
18
|
Mon May 9 13:48:20 2022 |
Rodolfo | Laser | Laser for H-like 209Bi |
I have changed the dye to Rhodamin 610 + Rhodamin 640. The Cobra wavelength can be tuned now from 600 to 630 nm.
The current wavelength is set to 605 nm. |
27
|
Tue May 10 11:59:45 2022 |
Simon, Christopher, Kristian, Sebastian | Laser | Labview VI checkups |
The lower limit (to position) of the Cobra control VI has to be one increment smaller.
This means for moving from 601,95 to 601,74 the "to position" has to be 601.735.
We also checked the displayed HV values in Go4 with the shown values of the HV VI.
Both values in go4 and in the VI show the same value and are changing "simultaneously" (depending on the refresh time of go4 and Labview). |
28
|
Tue May 10 15:26:35 2022 |
Rodolfo, Max | Laser | Laser stabilisation running |
It is now possible to stabilize and move the laser with both stages.
The software and hardware settings are attached.
For synchronizing with the laser pulse we use following frequency generator settings:
period: 200us
high: 5V
low: 0V
duty: 50%
The frequency generator is triggered with the external Q-switch trigger signal, that is in coincidence with the ESR HF.
To use the stabilization one has to use the external and NOT the q-switch mode of the laser. |
Attachment 1: MRC_Settings_for_Bismuth_10-05-2022.PNG
|
|
Attachment 2: MRC_hardware_Setting.PNG
|
|
35
|
Wed May 11 01:09:07 2022 |
Patrick, Sebastian, Volker, Guy | Laser | Beam overlap |
We overlapped the laser with the ion beam.
The ion beam positions are:
device | move in dir. | scratch (mm) | gone (mm) | gone + 4.5 (mm) | x0 (mm)
|
GECEDS1VU | up | -3.0 | -2.7 | 1.8 |
|
| | | | |
|
GECEDS1HA | right | -21.9 | -20.9 | -16.4 | -4.7
|
| left | -12.0 | -13.1 | |
|
| | | | |
|
GECEDS2VU | up | -3.8 | -2.9 | 1.6 |
|
| | | | |
|
GECEDS2HA | right | -19.2 | -18.4 | -13.9 | in [-0.1, 4.3]
|
| left | -9.0 | -9.7 | |
|
We used the "gone + 4.5" column where the scraper is in the middle of the ion beam to adjust the laser.
Picture 1: Final beam position.
Picture 2: Final MRC settings.
Note 1: By switching to the "real" bunching frequency, the laser went out of the stabilization. We restarted the MRC system and shifted the laser back into position with the piezo mirror, looking at the first screen (without amplifiers!).
After enabling everything again, the before found settings did not correspond to the correct beam position anymore, so we overlapped again.
Note 2: After the repositioning, the horizontal screw was at its limit and the beam stabilization failed. We moved the screw away from the limit while still hitting the screen and restarted the MRC system. The laser could then be stabilized again. |
Attachment 1: 2022-05-11_final_laser_spot.PNG
|
|
Attachment 2: 2022-05-11_final_MRC_settings_2.PNG
|
|
36
|
Wed May 11 03:10:55 2022 |
Patrick, Sebastian, Volker, Guy | Laser | Laser - bunch timing |
We added the delay and gate generator in between the bunch trigger and the CFD to move the timing of the PMT middle signal relative to the photodiodes north and south.
The center value of PD-North is at 19 bins, PD-South at 55 bins, the middle is therefore at 37 bins. The PMT peak should be half a revolution, ie 75 bins further in the histogram, giving a peak position of 112, from which then 3 bins are subtracted to account for the distance of the PMT-middle from the center of the straight section. So all in all the delay is adjusted so the PMT-middle peak appears at 109 bins (see attached plot). |
Attachment 1: 2022-05-06_Bi2023_laserscan_test.png
|
|
44
|
Wed May 11 14:59:01 2022 |
Rodolfo | Laser | Laser status |
I have measured the output energy of the laser system. All values in [mJ] and over 500 laser-pulses average
Date | Out of Quanta Ray @ 532 nm | Seed energy for Cobra @ 532 nm | Out of Cobra @ 601 nm
|
10.05.2022 | 533 | 485 | 71
|
11.05.2022 | 484 | 480 | 67
|
|
49
|
Wed May 11 22:19:17 2022 |
Max, Wilfried, Volker | Laser | laser bunch timing |
We overlapped the laser and ion bunch in the time domain. The delay has to be set on the module above the laser and NOT in the ESR Messhütte!
The values are:
PD NO: 63
PD SO: 98.5
Ion bunch: 150 |
Attachment 1: timing10_5_22.png
|
|
62
|
Thu May 12 15:21:23 2022 |
Sebastian, Simon | Laser | Flash lamps changed |
We (Sebastian and I) changed the flash lamps of the Quanta Ray Laser and measured the output energy of the laser system afterwards (still during warm up).
All values in [mJ] and over 500 laser-pulses average
Date | Out of Quanta Ray @ 532 nm | Seed energy for Cobra @ 532 nm | Out of Cobra @ 601 nm
|
10.05.2022 | 533 | 485 | 71
|
11.05.2022 | 484 | 480 | 67
|
12.05.2022 | 578 | 501 | 75
|
The lambda/2 is at 103°. Correction factor for the 532nm-monitor is 11.4 (exchanged broken monitor). |
70
|
Mon May 16 15:57:37 2022 |
Rodolfo | Laser | Wavemeter calibration |
I have calibrated our Wavemeter WS7 using the temperature stabilized He-Ne Laser (SIOS EN60 825-1:2014). In the screenshots below you can see the readout of the wavemeter before (vor) and after (nach) calibration. |
Attachment 1: WS7_VorKalibrationMitHe-Ne_16-05-2022.PNG
|
|
Attachment 2: WS7_NachKalibrationMitHe-Ne_16-05-2022.PNG
|
|
75
|
Mon May 16 21:08:48 2022 |
Rodolfo & Max | Laser | Laser power |
We checked the energy of the laser and optimized it a bit
Time | Seed energy for Cobra @ 532 nm | Out of Cobra @ 601 nm
|
20:41 | 415 | 63
|
20:48 | 506 | 73
|
The lambda/2 is now at 108°. |
82
|
Fri May 27 14:57:23 2022 |
Rodolfo | Laser | Laser Energy |
I have started up the laser system (Q-Ray & Cobra) again. Here some parameters*:
- Q-Ray's output energy (@532 nm: 547 mJ.
- Seed energy for Cobra# (@532 nm): 504 mJ.
- Cobra's output energy (@ 601.7 nm): 71 mJ.
* all values averaged over 500 pulses.
# angle of lambda/2-plate: 107° |
84
|
Mon May 30 00:56:33 2022 |
Max, Patrick | Laser | Beam overlap |
The ion beam positions are (+- 1mm):
device | move in dir. | scratch (mm) | gone (mm) | gone +- 4.5 (mm) | x0 (mm)
|
GECEDS1VU | up | -1.5 | 0.7 | 5.2 |
|
| | | | |
|
GECEDS1HA | right | -26 | -22 | -17.5 | -4.7
|
| left | -10.0 | -13.0 | -17.5 |
|
| | | | |
|
GECEDS2VU | up | -2.0 | 0.0 | 4.5 |
|
| | | | |
|
GECEDS2HA | right | -22.0 | -18.5 | -14.0 | 3.1
|
| left | -6.0 | -9.5 | -14.0 |
|
Notes:
- In between a measurement the scrapers stopped working in the device control. Therefore, we had to go into the electronics room.
- Position to scratch charge transferred beam: S1HA at -35.
- Beam diameter is roughly 6-7 mm (horizontally) and 4 mm vertically (as opposed to 1-2 mm).
- The charge transferred ion beam seems to shift the position of the main beam over time (1cm / 2h) as the main beam "transfers its intensity".
|
Attachment 1: 2022-05-30_laser_beam.png
|
|
Attachment 2: 2022-05-30_beamstab_settings.png
|
|
92
|
Mon May 30 19:42:20 2022 |
Rodolfo & Kristian | Laser | Changing laser wavelength |
We have changed the laser wavelength, e.g. the dye for the Cobra to Fluorescein 27.
We get about 84 mJ out the main amplifier @ 548 nm.
We also changed the optics and put into operation the MRC stabilization system for this new wavelength regime.
Next step is to T-scrape the ion- and laser-beams, e.g. to make the overlap of both beams at the E-Cooler. |
97
|
Tue May 31 03:18:52 2022 |
Patrick, Konstantin | Laser | Laser timing |
The final timings are (in bins +-1):
PD NO: 103
PD SO: 138
PMT middle: 44
e-cooler = (PD NO + PD SO) / 2: 121
e-cooler - PMT middle: 77
roundtrip / 2: 75
roundtrip / 2 + 10ns: 78 |
104
|
Tue May 31 21:58:55 2022 |
Max, Rodolfo | Laser | MRC Settings for 209Bi |
We have change the laser settings for the spectroscopy of 209Bi.
We have now T-scraped the laser beam to the old position of the ion beam. The settings for the MRC system are shown in the screenshot below. |
Attachment 1: MRC_209Bi_31-05-2022_2157.png
|
|
108
|
Wed Jun 1 17:20:04 2022 |
Rodolfo | Laser | HeNe @ WS7 |
|
Attachment 1: HeNe_WS7_01-06-2022_1.png
|
|
111
|
Fri Jan 13 17:05:22 2023 |
Rodolfo | Laser | Testing WS6 |
Testing the new Wavemeter (HighFinesse, WS6, SN3950) with the QRAY laser light @ 532 nm. Multimode fiber von High Finesse:FC/PC S (Corning, MM, OM2) |
Attachment 1: WS6_3950_QRayWavelength_13-01-2022.png
|
|
112
|
Wed Jan 18 13:02:12 2023 |
Rodolfo | Laser | Testing WS6 with Cobra |
Testing the WS6 (S.N. 3950) together with the Cobra.
Cobra operated with dye DCM in Ethanol + DMSO (Concentration 1:1)
- Oscillator concentration 0.3 g/l.
- Main amplifier concentration 0.0375 g/l (1/8 oscillator concentration).
- Laser linewidth: 6 GHz. Expansion factor: 10x. Grating: 2400 g/mm (40/2013)
Seed energy: 504 mJ @ 532 nm.
Cobra output energy: 112 mJ @ 640 nm.
Scan settings: 640.0 nm - 639.8 nm @ 0.005 nm / step.
Wavemeter settings: Pulsed (ext. trigg. 1 = QRay-Output: Lamp Sync). Precision: Fine.
- Wavemeter calibration. Reference source: Neon lamp.
- LTA-Data under ATPPC023 D:\WLM6 LTA data\D:\WLM6 (18.01.2023, 11.23, 640,10656 nm.lta) |
Attachment 1: WS6_640-639.8_0.005_18-01-2023_11.23.png
|
|