  28   Tue May 10 15:26:35 2022 Rodolfo, MaxLaserLaser stabilisation runningIt is now possible to stabilize and move the
laser with both stages.
 MRC_Settings_for_Bismuth_10-05-2022.PNGMRC_hardware_Setting.PNG 
  9   Wed May 4 22:55:37 2022 Rodolfo, SebastianLaserFCU / Temp Control BugWe found a "Bug" in the control of the Sirah
dye system.
  
  10   Wed May 4 22:56:22 2022 Rodolfo, Sebastian, KonstantinLaserLaser step complete, scan complete fixed"Laser step complete" and "laser scan complete"
are working again.
  
  52   Wed May 11 23:36:24 2022 Ruijiu Chen, Carsten, GuyDAQstart NTCAP.Carrier frequency:244MHz
IQ rate: 10M Samples/seconds
Reference level: -25
  
  11   Thu May 5 15:24:34 2022 SebastianLaserlaser busy problemCurrent status: 
Laser busy signal works correctly (5V if
the grating moves, 0V if not) if the laser
  
  12   Fri May 6 10:30:23 2022 SebastianDAQlaser busy / dye out signal fixedSolution for a working laser busy / dye out
signal was to disable the FCU and set the
FCU status to "False" (see attached picture).
 CobraFix_LaserBusy.PNG 
  13   Fri May 6 10:35:06 2022 SebastianDAQlaser step complete cabling changedI changed the laser step complete cabling
from a quad coinc (see attached picture)
to a Logic Fan In / Fan out (same as laser
 stepComplete_beforeChange_QuadCoincCabling.PNG 
  26   Tue May 10 11:40:55 2022 SebastianAcceleratorstill no beam @ESRThere is still no Beam in the ESR.

The beam is currently at TE3 - between SIS
  
  33   Tue May 10 23:10:25 2022 Sebastian, Patrick, Volker, GuyAcceleratorT-Scraper ReloadedWe have established a Control to "manually"
move the T-Scrapers via a "remote desktop".
We can now move both scrapers in the horizontal
  
  62   Thu May 12 15:21:23 2022 Sebastian, SimonLaserFlash lamps changedWe (Sebastian and I) changed the flash lamps
of the Quanta Ray Laser and measured the
output energy of the laser system afterwards
  
  67   Fri May 13 11:46:51 2022 Sebastian, Simon Laser energyWe (Sebastian and I)  measured the output
energy of the laser system again. 
All values in [mJ] and over 500 laser-pulses
  
  27   Tue May 10 11:59:45 2022 Simon, Christopher, Kristian, SebastianLaserLabview VI checkupsThe lower limit (to position) of the Cobra
control VI has to be one increment smaller.
This means for moving from 601,95 to 601,74
  
  39   Wed May 11 06:41:32 2022 Simon, Christopher, ShahabRunsshiftsnew shift started.   
  40   Wed May 11 06:47:58 2022 Simon, Christopher, ShahabRunsrestart laser scan
[TABLE border="1"]
Time | SIS-intensity | ESR-intensity after
  
  41   Wed May 11 07:28:39 2022 Simon, Christopher, ShahabGeneralDAQ Loop for shiftersProcedure:

- Scanning phase ends (on LabView wait until
  
  58   Thu May 12 11:31:11 2022 Simon, Christopher, Shahab, Wilfried, Sebastian, RaganGeneralRuns morning shift
[TABLE border="1"]
Time | SIS-intensity | ESR-intensity after
  
  55   Thu May 12 10:10:35 2022 Simon, Christopher, Wilfried, SebastianDAQ1ms gated PMT north / middle added to DAQWe added on DAQ1 and DAQ2 1ms gated PMT North
and PMT Middle signals.
Now we "supress" the signal by a factor of
  
  14   Fri May 6 16:29:03 2022 Simon, PatrickDetectorsPMT preparationWe changed PMT north and PMT south and tested
them at 1500 V.
  
  23   Mon May 9 19:05:17 2022 WilfriedDAQQ-Switch signals + 0.5 ms and +1 ms in VUPROMWe have generated two additional signals which
are fed into two VUPROM channels:
Q-Switch + 0.5 ms --> channel 9
  
  94   Mon May 30 21:52:08 2022 Wilfried, Danyal, David, RodolfoGeneralPosition of T-ScrapersWe have determined the position of the H-like
208Bi ion beam using the T-Scrapers. The
position actually did not change in comparison
 MRC_208Bi_30-05-2022-10Uhr.png 