  Nuclear hyperfine mixing in 229Th89+ 2024, Page 1 of 4  Not logged in ELOG logo
Entry  Wed Jun 12 18:48:43 2024, Afternoon shift, Runs, First laser scans 2024_06_12_229Th_Lifetime_at_190MeV.png
[TABLE border="1"]
Time | Part. Det. Int | ESR-intensity| cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|-
Entry  Thu Jun 13 18:03:36 2024, Afternoon shift, Runs, Continuation: First scans 2024_06_13_229Th_Lifetime_at_190MeV_afterscraping.png2024-06-13_23-13-55-242.png
[TABLE border="1"]
Time | Part. Det. Int | ESR-intensity| cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|-
Entry  Fri Jun 14 06:38:41 2024, Konstantin Mohr, Bernhard Maass, Peter Micke, Lukas Kau, Dounia Boudefla, Runs, Morning runs 
Very successful shift, no crashes until 11:10
Previous night shift had problems with the pulse laser starting around 3am.
Something was loose at the FCU. Rodolfo was called in and was able to fix it.
Entry  Fri Jun 14 14:44:51 2024, Afternoon shift, Runs, Afternoon shift 
Pizza shift

[B]the scan direction is from higher wavelength to lower wavelength, even though it is noted otherwise here in the software.
Entry  Thu Apr 25 16:34:05 2024, Rodolfo, Laser, Efficiency of Fluorescein27 Fluorescein27_Efficiency.png
The figure below shows the efficiency of Fluorescein27 vs. Wavelength. Each point was averaged over 500 laser shots.

Dye Laser: Sirah, Cobra Strech. 
Entry  Wed May 29 09:49:30 2024, Rodolfo, Laser, Laser-beam profile Laserstrahl-at-EKühler_2024-05-03_#014.jpgLaserstrahl-at-EKühler_2024-05-03_Photo.jpg
Using a telescope (Plan-concave lens -500 mm and Plan-convex lens +1000 mm) after the pulsed-dye laser system we got the following laser-beam profile at
the Electron-Cooler's position. Dye Laser @ 550 nm / 275 nm  
Entry  Mon Jun 3 19:42:06 2024, Rodolfo, Laser, Laser-beam through E-Cooler LaserBeam-at-Screen_ESR-NO_08-06-2024.JPG
UV-laser beam is now going through the Electron-Cooler. 

Telescope settings:
Entry  Thu Jun 6 07:12:49 2024, Konstantin Mohr, Laser, T-Scraper 6x
The T-Scrapers can be driven by DeviceControl, which must be started from a console. To do so, first open the app-launcher and then select "Device Control",
which is found under "Betrieb & Steuerung". Then select the appropriate context (should be the name of the experiment). After a short time the device icons
will be present. The scrapers can be found in the last four columns and the second last row (see screenshot 1). By default, they should be colored like
Entry  Fri Jun 7 11:14:04 2024, Rodolfo, Laser, Trigger-signal for MRC MRC_TriggerSignal_T1_06-06-2024.jpgMRC_TriggerSignal_T2_Tmin_06-06-2024.jpgGeneratorSettings_for_MRC_06-06-2024.jpgMRC_Parameters_06-06-2024_18-30.JPG
We are using a 2-Channel Function Generator (from Textronik) to generate the trigger signal for the MRC Beam Stabilization. 

The triggers are (according to the MRC nomenclature):
Entry  Tue Jun 11 21:38:35 2024, Julian, Imke, Rodolfo, Laser, Wavemeter HeNe long time stability HeNe_LangeMessung_11-06-2024.JPG
The HeNe was running over night and the measured laser frequency of the wavemeter was logged, see picture.
A significant drift of around 300 MHz is visible during the night.
It is possible, that the laser lab got too cold during the night, as the pulsed laser was not running.
Entry  Wed Jun 12 17:12:13 2024, Julian, Danyal, Rodolfo, Laser, Timing für 229Th89+ Timing_229Th89_12-06-2024.JPG
We did the timing Ion-Beam Laser-beam time overlap. 

The ion revolution frequency (and therefore the bunching frequency) is: 1.5429 MHz.
Entry  Thu Jun 13 18:23:46 2024, Julian, Laser, Restarting the laser controller 
If the laser controller is locked up again, the little display will show an error message after an attempted medusa scan.
To restart it, follow these steps:
1.) Unlock the beam stabilization.
Entry  Thu Jun 13 18:38:24 2024, Rodolfo Sanchez, Julian Palmes, Sebastian Klammes, Laser, Laser Dye change - new MRC settings - scraper checks of laser beam MRC-settings-13-06-2024-after-dye-change-at-3pm.JPG
We changed the dye at ~3pm.

After some fine adjustements of the laser we had to change the MRC settings, because the laser position changed.
Entry  Tue Apr 16 14:42:37 2024, Jan, General, G-PAC Proposal G-22-00052.pdf
 
Entry  Tue Apr 16 16:28:21 2024, Danyal, General, Great 
Thanks Jan!
Entry  Wed May 29 10:54:21 2024, Shahab, Jan, General, Gas Target Events with new timing system 2024-05-29_11-07-40-259.pngIMG_0465.JPGIMG_0466.JPGIMG_0467.JPG
Using the new timing system, it is possible to grab any events as pulses, but also as long inter event gates under GENESYS. Many instances of GENESYS can
be running on ACO computers, so that there is no conflict, if an instance is running in main control room or another in ESR-Messhütte.
Entry  Thu Jun 6 07:54:15 2024, Danyal, General, pressure gauges and ion pumps  
yesterday (05.06.2024), firstly, the two ion pumps (underneath the fluorescence chamber) were switched off by the vacuum group (Björn Mai, Tel: 3438).
Secondly, the two pressure gauges (located near the target) were switched off.  The pressure gauge at the fluo chamber is called E02VM3x and the pressure
was 5.5e-11 mbar.  The other gauge is called IG2 and located further south, it measured 3.7e-11 mbar.  That was when ion pumps were still switched on.
Entry  Fri Jun 7 17:10:45 2024, Sebastian, General, XUV LED 
Maximum XUV LED Voltage: 6.5V
Entry  Sat Jun 8 13:11:52 2024, Carsten, General, Put the go4login in atplaser .bashrc 
Put the go4login in atplaser .bashrc in order to avoid that one always has to go to the go4 directory to set the environment variable.
go4login64 is now executed automaticlally and the environments are now set every time you open a (bash) shell.
A fixed version 6.3 is used. The login file might be changed back after the beamtime.
Entry  Tue Jun 11 21:23:11 2024, Julian, Imke, General, How to read out voltage medusa1.JPGmedusa3.pngmedusa4.png
1.) If not already opened, open the Medusa Laser Control software, see picture:
2.) In the "view" submenu, choose "proteus status".
3.) At the bottom of the list, the K250 shows the scaled voltage.