  proton-capture on 118Te, Page 3 of 29  Not logged in ELOG logo
ID Date Author Category Subject Yeardown
  324   Wed May 12 11:34:11 2021 Jan GloriusCalibrationrun0027 - crap2021
  325   Wed May 12 11:47:41 2021 Jan GloriusCalibrationrun0029 - Xray90 calib Pb210 d=167.5mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************

Detector: GEM3000 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 11:46:19 - 12.05.2021
Stop time:  13:25:46 - 12.05.2021

file name: e127b_run0029.lmd
avrg. rate: 50Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  326   Wed May 12 13:36:24 2021 Jan GloriusCalibrationrun0030 - Xray90 calib Am241_hi d=167.5mm2021
Efficiency calibration in the lab

AMPLIFIER JUMPED!

************************************
CAUTION! 
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************

Detector: GEM3000 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 13:35:42 - 12.05.2021
Stop time:  14:05:08 - 12.05.2021

file name: e127b_run0030.lmd
avrg. rate: 600Hz
dead-time:  4%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  327   Wed May 12 14:16:27 2021 Jan GloriusCalibrationrun0031 - Xray90 calib Am241_hi d=167.5mm2021
Efficiency calibration in the lab


************************************
CAUTION! 
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************

Detector: GEM3000 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 14:16:07 - 12.05.2021
Stop time:  14:34:00- 12.05.2021

file name: e127b_run0031.lmd
avrg. rate: 600Hz
dead-time:  4%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  328   Wed May 12 14:55:04 2021 Jan GloriusCalibrationrun0032 - Xray90 calib Cd109 d=167.5mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************

Detector: GEM3000 - 90 deg
Source: 109Cd
Distance: 167.5mm
Start time: 14:54:36 - 12.05.2021
Stop time:  15:10:18- 12.05.2021

file name: e127b_run0032.lmd
avrg. rate: 1800Hz
dead-time:  12%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  329   Wed May 12 15:28:55 2021 Jan GloriusCalibrationrun0033 - Xray145 calib Cd109 d=305mm2021
Efficiency calibration in the lab

Detector: GEM3000 - 145 deg
Source: 109Cd
Distance: 305mm
Start time: 15:28:01 - 12.05.2021
Stop time:  16:01:57 - 12.05.2021

file name: e127b_run0033.lmd
avrg. rate: 600 Hz
dead-time:  5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  330   Thu May 13 11:51:42 2021 Jan GloriusCalibrationrun0034 - Xray90 calib Am241_hi d=167.5mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************

Detector: GEM1800 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 11:49:53 - 13.05.2021
Stop time:  12:02:24 - 13.05.2021

file name: e127b_run0034.lmd
avrg. rate: 600Hz
dead-time:  5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  331   Thu May 13 12:06:51 2021 Jan GloriusCalibrationrun0035 - Xray90 calib Cd109 d=167.5mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************

Detector: GEM1800 - 90 deg
Source: 109Cd
Distance: 167.5mm
Start time: 12:06:24 - 13.05.2021
Stop time:  12:16:46 - 13.05.2021

file name: e127b_run0035.lmd
avrg. rate: 1800Hz
dead-time:  12%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  332   Thu May 13 12:21:41 2021 Jan GloriusCalibrationrun0036 - Xray90 calib Ba133_lo d=167.5mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************

Detector: GEM1800 - 90 deg
Source: 133Ba_lo
Distance: 167.5mm
Start time: 12:21:21 - 13.05.2021
Stop time:  13:22:04 - 13.05.2021

file name: e127b_run0036.lmd
avrg. rate: 300Hz
dead-time:  2%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  333   Thu May 13 13:24:44 2021 Jan GloriusCalibrationrun0037 - Xray90 calib Pb210 d=167.5mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 13:24:26 - 13.05.2021
Stop time:  14:06:35 - 13.05.2021

file name: e127b_run0037.lmd
avrg. rate: 70Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  334   Thu May 13 14:16:58 2021 Jan GloriusCalibrationrun0038 - Xray145 calib Pb210 d=305mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Rate is increasing slowly, from ~70 to ~220 Hz in ~7 hours.
Also the threshold seems to be lower now.
Also possible source holder movement!
Resolution still good.
Take this run with care!

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 14:16:18 - 13.05.2021
Stop time:  21:09:12 - 13.05.2021

file name: e127b_run0038.lmd
avrg. rate: 70Hz > 220Hz!!!!
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  335   Thu May 13 17:38:13 2021 Jacobus SwartzCalibrationRunlist for X-ray detector calibrations2021
Here is a run summary of all the calibration measurements performed so far on the three X-ray detectors. 
Attachment 1: E127_Runlist-calibrations-4-3.xlsx
  336   Thu May 13 21:22:31 2021 Jan GloriusCalibrationrun0039 - Xray145 calib Pb210 d=305mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Rate is increasing slowly 200 > 400 Hz, more noise coming in.
Also possible source holder movement!
Resolution still good.
Take this run with care!

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 21:22:00 - 13.05.2021
Stop time:  08:56:31 - 14.05.2021

file name: e127b_run0039.lmd
avrg. rate: 220Hz > 400Hz !!!!
dead-time:  3%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  337   Fri May 14 09:06:55 2021 Jan GloriusCalibrationrun0040 - Xray145 calib Ba133_hi d=305mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Threshold slightly increased

Detector: GEM1800 - 145 deg
Source: 133Ba_hi
Distance: 305mm
Start time: 09:06:33 - 14.05.2021
Stop time:  09:31:32 - 14.05.2021

file name: e127b_run0040.lmd
avrg. rate: 850Hz
dead-time:  6%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  338   Fri May 14 09:35:13 2021 Jan GloriusCalibrationrun0041 - Xray145 calib Cd109 d=305mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Threshold slightly increased

Detector: GEM1800 - 145 deg
Source: 109Cd
Distance: 305mm
Start time: 09:34:52 - 14.05.2021
Stop time:  09:47:42 - 14.05.2021

file name: e127b_run0041.lmd
avrg. rate: 600Hz
dead-time:  5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  339   Fri May 14 10:28:49 2021 Jan GloriusCalibrationrun0042 - Xray145 calib 241Am_hi d=305mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Threshold slightly increased

Detector: GEM1800 - 145 deg
Source: 241Am_hi
Distance: 305mm
Start time: 10:27:58 - 14.05.2021
Stop time:  11:14:03 - 14.05.2021

file name: e127b_run0042.lmd
avrg. rate: 200Hz
dead-time:  1.5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  340   Fri May 14 11:17:48 2021 Jan GloriusCalibrationrun0043 - Xray145 calib 210Pb d=305mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Threshold slightly increased

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 11:17:23 - 14.05.2021
Stop time:  13:47:42 - 14.05.2021

file name: e127b_run0043.lmd
avrg. rate: 40Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  341   Fri May 14 13:58:52 2021 Jan GloriusCalibrationrun0044 - Xray145 calib 210Pb d=305mm2021
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

90deg TFA diff set from 50 to 200.
Also threshold is slightly increased.

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 13:58:33 - 14.05.2021
Stop time:  16:46:00 - 14.05.2021

file name: e127b_run0044.lmd
avrg. rate: 40Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  342   Sat May 15 20:21:59 2021 Jan GloriusCalibrationrun0045 - Si pulser run2021
Detector: DSSD
Source: CAEN Detector Emulator

Start time: 20:24:14 - 15.05.2021
Stop time:  21:01:34 - 15.05.2021

file name: e127b_run0045.lmd
avrg. rate: 1000Hz
dead-time:  7.5%


Si bias test cycle:
0V > 50V/200uA > 100V/280uA > 150V/370uA > 120V/320uA > 90V/260uA > 60V/210uA > 30V/160uA > 0V
This cycle is run twice i the first 15 minutes and shall be used for calibration of the scaler channels.

Also ESR trafo is running with beam.
  343   Sun May 16 11:25:25 2021 Jan GloriusCalibrationrun0046 - Xray E-calib 241Am all dets2021
Detectors: Xray1 (GEM1800), Xray2 (GLP), Xray3 (GEM3000)
Sources: different Am241 at each detector

Start time: 11:25:04 - 16.05.2021
Stop time:  11:29:45 - 16.05.2021

file name: e127b_run0046.lmd
avrg. rate: 6000Hz
dead-time:  30%