Entry  Sat May 22 16:29:07 2021, Laszlo et al., General, coolers are back, 2021 
we can measure again.

(please write a detailed description if needed)
Entry  Wed Jul 14 14:46:05 2021, Jan, Sophia, Detectors, collimator geometry, 2021 10x
We have measured the collimator geometry. The annotations of the variables can be found in the attached sketch  

[TABLE border="1"]
Entry  Sat Mar 21 22:57:28 2020, Laszlo, Jan, DAQ, changing cables of the Xray 90 and 145,  
Energy signals of 90 and 145 degree detector have been checked and exchanged
They were connected to the wrong ADC channel.
Entry  Sat May 29 13:43:39 2021, Tino, Rene, Yuri, Laszlo, Markus, Runs, change of settings, 2021 
from now on, everal beam setting will be different. We could not get the beam back to were it was before. When the scraper was pulled back by 2 cm, the
loe/energy beam was visible again via cap. 
Schottky. However, no x-rays could be observerved during the target-on phase. 
Entry  Thu Apr 29 20:32:52 2021, Jan Glorius, Calibration, calibration sources, 2021 
We use the following list of sources for calibration of the Xray detectors:

- 210Pb [40.1 kBq (4%), 01.10.2020 12:00 UTC] [URL=https://elog.gsi.de/esr/X-Ray+Lab/201006_131944/20-006_210Pb_40kBq.pdf]SpecSheet[/URL]
Entry  Thu Jan 23 16:35:47 2020, Laszlo, DAQ, cable documentation,  cable_documentation_e127.pdfcable_documentation_e127.ods
 
Entry  Fri Feb 21 09:39:40 2020, Laszlo, DAQ, cable documentation,  cable_documentation_e127.pdfcable_documentation_e127.ods
 
Entry  Wed Mar 11 14:33:44 2020, Laszlo, General, beam time manual,  E127_experimental_manual.pdfmanual.odt
Here you can find the digital version of the E127 beam time manual.
Entry  Mon Apr 6 23:00:51 2020, Laszlo, Calibration, beam energies for 124Xe and 118Te measurements,  SP54120040623030.pdf
"The e- cooler settings were the same for the 124Xe and 118Te beams" - Sergey --> speed of the ions is the same --> same MeV/u

E_beam = 10.0606MeV/u
Entry  Thu Jul 8 14:07:19 2021, Jan, Calibration, beam energies, 2021 CodeCogsEqn.pngCodeCogsEqn(1).png
Table of ion energies calculated from cooler parameters

[TABLE border="1"]
Entry  Thu May 20 17:31:50 2021, Jan, Laszlo, Beatriz, General, beam & target alignement, 2021 
beam and target alignement was checked by looking at the Xray rates. Best case when:
Target bump (position) SC 9: -18.5mm
Entry  Thu Aug 29 10:02:22 2019, Jan, Detectors, bakeout at ESR setup,  T_set_bakeout.png
The bakeout for our detector setup has started on the evening of 27.08.2019 
We have a temperature sensor in the vacuum next to the detector.
The goal is to maximize bakeout temperature, while not going above 125°C at the detector/sensor.
Entry  Mon Mar 23 07:56:51 2020, Laszlo, General, Yuri made magick and we operate again,  
DSSSD was moved out (HV switched off)
also the scraper was mistakenly moved.
Entry  Sun Mar 22 19:49:12 2020, Michael, Laszlo, Yuri, , Yuri is still tuning FRS,  
We filled the Xray detectors.
Yuri is sill tuning FRS for primary beam.
Entry  Wed May 26 14:28:23 2021, Jan, Analysis, Xrays from 118Te & 124Xe, 2021 xray90_124Xe_118Te.png
Attached is a plot of the uncalibrated energy spectra taken with the 90 degree xray detector for 118Te (red) and 124Xe (blue).
Shown is the region covering the K-alpha and K-REC peaks.
Entry  Fri Mar 20 16:58:02 2020, Jan, DAQ, Xray2 & Xray3 cabling,  
We found that Xray2 and Xray3 the cabling is not consistent:

90 degree detector 
Entry  Tue May 4 14:34:51 2021, Jan Glorius, Detectors, Xray weight control induces signals, 2021 
When the readout/control of the det. weight control via serial connection is active, there is a clear signal induced in the Xray detectors for each command
sent.
Entry  Sun May 16 11:54:12 2021, Jan Glorius, Calibration, Xray performance in lab, 2021 
These are the current Xray detector performances in the lab using data from run0046

[TABLE border="1"]
Entry  Fri Apr 30 10:55:34 2021, Jan Glorius, DAQ, Xray multi-triggers - blocking gate, 2021 
From the CFD of Xray1 (90deg) we get multiple trigger signals every ~5sec for unknown reasons. 
No hint from the preamp signals.
Entry  Mon May 3 14:55:44 2021, Jan Glorius, Calibration, Xray energy calib. table, 2021 
Here is the table of energy calibration lines from Am241 for all Xray detectors for several measurements.
Also the linear fit parameters [ E(ch) = A*ch + B ] are given at the bottom of the table.