| ID | Date | Author | Category | Subject | Text |  | 
| 79 | Wed Jun 26 11:59:12 2024 | Ruijiu, Shahab, Carsten | Analysis | Prepare plots for Thomas's presentation | data: (env_OnlineDataAnalysisSystem) litv-exp@lxbk0497:/lustre/astrum/experiment_data/2024-05_E018/OnlineDataAnalysisSystem/data/iq/IQ_2024-06-
 12_21-47-54$
 | 6x   | 
| 78 | Sat Jun 15 03:46:34 2024 | Night shift | Accelerator | GSI down | All quadrupole and dipole magnets at SIS and ESR are out of order. Repairs can be expected
 earliest tomorrow evening. No shifts will
 |  | 
| 77 | Fri Jun 14 14:44:51 2024 | Afternoon shift | Runs | Afternoon shift | Pizza shift 
 [B]the scan direction is from higher wavelength
 |  | 
| 76 | Fri Jun 14 13:14:29 2024 | Konstantin, Bernhard, Peter, Lukas | General | DAQ crashes at :10 | 10 Minutes past a full hour (XX:10) the windows PCs (sometimes) loose connection.
 This causes the labview server & the laser
 |  | 
| 74 | Fri Jun 14 11:44:23 2024 | Carsten Brandau | Accelerator | Cooling Time in SC7 increased from 9s to 12s | Cooling Time in SC7 increased from 9s to 12s |  | 
| 73 | Fri Jun 14 06:38:41 2024 | Konstantin Mohr, Bernhard Maass, Peter Micke, Lukas Kau, Dounia Boudefla | Runs | Morning runs | Very successful shift, no crashes until 11:10 Previous night shift had problems with the
 pulse laser starting around 3am.
 |  | 
| 72 | Thu Jun 13 22:08:07 2024 | Dayshift | Accelerator | Problems with Accelerator? | It seems to be that there are some problems with the injections in the ESR.
 The last 2 accumulation cycles showed that
 |     | 
| 71 | Thu Jun 13 18:48:26 2024 | Ruijiu Chen, David, Shahab, Yuri | Analysis | The ion identification of 229Th is correct. | Harmonic: 124.0 , Bp: 7.796046 [Tm] ion            fre[Hz]
 yield [pps]
 |    | 
| 70 | Thu Jun 13 18:38:24 2024 | Rodolfo Sanchez, Julian Palmes, Sebastian Klammes | Laser | Laser Dye change - new MRC settings - scraper checks of laser beam | We changed the dye at ~3pm. 
 After some fine adjustements of the laser
 |   | 
| 69 | Thu Jun 13 18:23:46 2024 | Julian | Laser | Restarting the laser controller | If the laser controller is locked up again, the little display will show an error message
 after an attempted medusa scan.
 |  | 
| 68 | Thu Jun 13 18:03:36 2024 | Afternoon shift | Runs | Continuation: First scans | [TABLE border="1"] Time | Part. Det. Int | ESR-intensity| cooler-HV
 | LMD-DAQ1 Files | LMD-DAQ2 Files | scan
 |    | 
| 67 | Thu Jun 13 14:56:09 2024 | Stefan Schippers | Accelerator | Estimated beam lifetime: About  150 s | Estimate of storage lifetime using my code 'hydrocal':
 
 |  | 
| 66 | Thu Jun 13 14:41:22 2024 | Ruijiu Chen, David | Analysis | Lifetime of beam during June 13 9:00~11:00 |  |   | 
| 65 | Thu Jun 13 14:35:27 2024 | Ruijiu Chen, David | Analysis | Lifetime of beam during June 13 7:00~9:00 |  |   | 
| 64 | Thu Jun 13 14:20:36 2024 | Ruijiu Chen, David | Analysis | Lifetime of beam during June 13 4:30:7:00 |  |   | 
| 63 | Thu Jun 13 13:29:47 2024 | Ruijiu Chen, David | Analysis | Lifetime of beam during June 12 22:00- June 13 0:00 |  |   | 
| 62 | Thu Jun 13 12:37:14 2024 | Ruijiu Chen, David | Analysis | Lifetime of beam in center of mass is 193 s. |  |   | 
| 61 | Thu Jun 13 12:05:31 2024 | Konstantin, Peter, Jonas | Accelerator | RF-amplitude | To realize the shortest bunches possible, we check the bunch length as a function of
 two different bunching amplitudes. With the
 |    | 
| 60 | Thu Jun 13 10:42:02 2024 | Danyal | General | Question | What is the ion beam lifetime, according to the DCCT and Schottky?
 Can somebody measure it, please?
 |  | 
| 59 | Thu Jun 13 10:32:29 2024 | Konstantin | Accelerator | (Proper) Cooler settings | We change the acceleration voltage of the electron cooler and observe the platform
 high voltage and the count rate at the particle
 |  | 
| 58 | Thu Jun 13 10:06:35 2024 | Constantine, Peter, Lukas, Jonas, Dounia, Carsten | Accelerator | Suspected Beam Troubles (is it really 229-Thorium?) | From experience, the lifetime of the beam in the storage ring should be on the order
 of magnitude of hours, however, we observer
 |  | 
| 57 | Thu Jun 13 02:05:04 2024 | Carsten | Accelerator | Short Lifetime at 100 mA | Lifetime still is short |   | 
| 55 | Wed Jun 12 22:48:55 2024 | Carsten |  | Beam Halo | The seamingly short live time reported in [URL=https://elog.gsi.de/esr/E0052/53]entry
 53[/URL] is due to a halo from a different
 |  | 
| 54 | Wed Jun 12 21:21:15 2024 | Yuri | Detectors | NTCap started | Fcenter - 244.6 MHz BW - 10 MHz
 |  | 
| 53 | Wed Jun 12 18:48:43 2024 | Afternoon shift | Runs | First laser scans | [TABLE border="1"] Time | Part. Det. Int | ESR-intensity| cooler-HV
 | LMD-DAQ1 Files | LMD-DAQ2 Files | scan
 |   | 
| 51 | Wed Jun 12 17:12:13 2024 | Julian, Danyal, Rodolfo | Laser | Timing für 229Th89+ | We did the timing Ion-Beam Laser-beam time overlap.
 
 |   | 
| 49 | Wed Jun 12 12:12:44 2024 | Konstantin Mohr, Julien Spahn, Jonas Koedel, Bernhard Maass | Accelerator | First Scraper Scan | Thorium is in the ESR, we have time to check the beam position with the scrapers.
 Beam Intensity in the ring 6E-5
 |   | 
| 48 | Wed Jun 12 09:43:57 2024 | Carsten | Accelerator | Stopped NTCAP at 9:43 | Setup at ESR is carried on - stopped the long-time NTCAP run from overnight,
 |  | 
| 47 | Wed Jun 12 09:17:43 2024 | Carsten |  | Tests of background with ionization pumps on (from monday) | On monday the ionization pumps were switch on to test the background, and if it is possible
 to run with the pumps switched on.
 |      | 
| 46 | Wed Jun 12 01:51:51 2024 | RuiJiu, David | Analysis | ID | The preliminary identification has been thoroughly checked using the RionID software. We have
 confirmed that the identification is accurate.
 |      | 
| 45 | Tue Jun 11 23:46:06 2024 | Carsten )))) | Accelerator | ID | ---- The frequency of the primary beam 238U82+
 at the injection:
 |  | 
| 44 | Tue Jun 11 22:59:06 2024 | Carsten | Accelerator | Identification / NTCAP started | ESR is set-up in most parts.We are trying to identify our 229Th89+. It is more cumbersome
 then exspected because of the thick target
 |  | 
| 43 | Tue Jun 11 21:38:35 2024 | Julian, Imke, Rodolfo | Laser | Wavemeter HeNe long time stability | The HeNe was running over night and the measured laser frequency of the wavemeter was logged,
 see picture.
 |   | 
| 42 | Tue Jun 11 21:23:11 2024 | Julian, Imke | General | How to read out voltage | 1.) If not already opened, open the Medusa Laser Control software, see picture:
 2.) In the "view" submenu, choose "proteus
 |     | 
| 41 | Tue Jun 11 09:17:46 2024 | Julien, Konstantin, Lukas | DAQ | Important commands (DAQ1) | To connect to the DAQ1 pc, open a terminal and type:
 ssh atplaser@r4l-41
 |       | 
| 40 | Mon Jun 10 11:59:12 2024 | Ruijiu Chen | Analysis | The ion identification of spectrum recorded on 2022 | The simulated frequency is given in simulation_result.out. The yield is calculated by LISE file. The
 LISE file is attached here also.
 |     | 
| 39 | Mon Jun 10 09:17:38 2024 | Carsten | DAQ | Updated Channel List |  |    | 
| 38 | Mon Jun 10 08:57:35 2024 | Carsten/Konstantin | DAQ | Further Tests from Sunday 2024-06-09 - new voltages and tresholds for PMTs - testdata in files | Voltages and thresholds increased for PMT Middle and North (in order to account for
 the ADC measurement in DAQ2
 |  | 
| 37 | Sun Jun  9 10:20:38 2024 | Carsten/Danyal | Detectors | Recap from yesterdays work  with test beam (Detectors / beam /DAQ) | In the ESR we had thw whole day (saturday) a 400MeV/u bunched 238U92+ test beam (about
 2 MHz). The RF signal is cabled as input
 |  | 
| 36 | Sun Jun  9 10:08:24 2024 | Carsten/Danyal | Detectors | Particle detectors - Recap of yesterdays work | [B]South (behind cooler) - "SA_SZ" - DD2...[/B] We have put the particle detecotr North to
 the poistion behind the cooler (north) to
 |  | 
| 35 | Sat Jun  8 13:11:52 2024 | Carsten | General | Put the go4login in atplaser .bashrc | Put the go4login in atplaser .bashrc in order to avoid that one always has to go to the
 go4 directory to set the environment variable.
 |  | 
| 34 | Sat Jun  8 12:45:05 2024 | Carsten | DAQ | Temporary change of RF signal | The setting was 1.540974 MHz and division factor 68 (for a 200 MeV/u) beam, and the
 signal was taken from a signal generator.
 |  | 
| 33 | Sat Jun  8 12:02:20 2024 | Carsten | Detectors | Test Particle Detectors - Det South not working | With the test beam from ESR I started to have a look at the PartDets.:
 North outside: Works properly with a voltage
 |  | 
| 32 | Fri Jun  7 17:10:45 2024 | Sebastian | General | XUV LED | Maximum XUV LED Voltage: 6.5V |  | 
| 31 | Fri Jun  7 15:29:50 2024 | Rodolfo | Detectors | Stored beam @ ESR | One of the first PMT signals from the stored-ion beam @ ESR
 |   | 
| 30 | Fri Jun  7 11:14:04 2024 | Rodolfo | Laser | Trigger-signal for MRC | We are using a 2-Channel Function Generator (from Textronik) to generate the trigger
 signal for the MRC Beam Stabilization.
 |      | 
| 29 | Fri Jun  7 09:15:31 2024 | Carsten | DAQ | Updated Channel List |  |    | 
| 28 | Thu Jun  6 19:51:11 2024 | Carsten | Accelerator | ESR crew left after persisting problems with ESR magnets | No beam today.--. |  | 
| 27 | Thu Jun  6 19:47:01 2024 | Carsten | DAQ | Exchange of Mesytec CFD/shaper - noise on signal lines from ESR magnets | Jan exchanged the Mesytec CFD/shper since it could not be "remote controlled" anymore.
 The new one has different shaping times and
 |  | 
| 26 | Thu Jun  6 19:42:20 2024 | Carsten | Detectors | Exchange of amplifer channels / cables channels / HV voltage cables for PMT south | In order to check if the problem of regular burst of pulses on the PMT south is related
 to amplfier channels and or the HV cable,
 |  | 
| 25 | Thu Jun  6 09:39:23 2024 | Carsten | DAQ | Preliminary Channel Table (DAQ1 and DAQ2) |  |    | 
| 24 | Thu Jun  6 09:23:53 2024 | Konstantin Mohr | DAQ | Timing | The best way to adjust the bunch-laser timing is to start go4 (mbs must be running) from
 atppc023. Therefore use x2goclient and use
 |     | 
| 22 | Thu Jun  6 08:16:36 2024 | Danyal | Accelerator | scraper info by Robert Boywitt | attched is more info about the T-scrapers (E-cooler) by Robert Boywitt (Tel: 2067 or
 3357)
 |      | 
| Draft | Thu Jun  6 07:54:15 2024 | Danyal | General | pressure gauges and ion pumps | yesterday (05.06.2024), firstly, the two ion pumps (underneath the fluorescence chamber)
 were switched off by the vacuum group (Björn
 |  | 
| 20 | Thu Jun  6 07:12:49 2024 | Konstantin Mohr | Laser | T-Scraper | The T-Scrapers can be driven by DeviceControl, which must be started from a console. To
 do so, first open the app-launcher and then
 | 6x   | 
| 19 | Wed Jun  5 19:22:43 2024 | Carsten, Konstantin, Rodolfo | Detectors | PMT tests and preliminary thresholds | For all channels of the CAEN CFD the width of the logic signal is set to 255ns (maximum)
 to mitigate potential double pulses. For
 |    | 
| 18 | Wed Jun  5 14:17:36 2024 | Carsten | Accelerator | Distance Cooler Detector | Calculation from Helmut Weick (copied from E142)
 Distance e-cooler to particle detectors chamber:
 |  | 
| 17 | Tue Jun  4 22:54:02 2024 | Rodolfo | Detectors | PMT - Manufacture - Test Result Data |  |     | 
| 16 | Mon Jun  3 19:42:06 2024 | Rodolfo | Laser | Laser-beam through E-Cooler | UV-laser beam is now going through the Electron-Cooler. 
 
 |   | 
| 15 | Mon Jun  3 18:24:44 2024 | Rodolfo | Detectors | PMT arrangement | The picture shows the actual configuration of the PMTs.
 
 |    | 
| 14 | Sun Jun  2 00:01:46 2024 | Carsten | DAQ | 25ps TDC CAEN V1290N DAQ2 lead to crash in DAQ2 due to a hardware issue | On the CAEN V1290N TDC the piggy board for the lower 8 channels is not fixed well to
 the motherboard.
 |  | 
| 13 | Sat Jun  1 19:19:07 2024 | Carsten | Detectors | Reminder: Check signals in scope and in DAQ of detectors with voltages switched off (vs. CFD threshold) | When the ion pumps and vacuum gauges are switched off the scaler data of
 PMT_south, PMT_middle, PMT_north, PD_NO,
 |  | 
| 12 | Sat Jun  1 19:00:57 2024 | Carsten | DAQ | Change of (offline) frequency for common stop (CS) of  VUPROMs (and 2nd TDC) and scaling factor | Temporarily, for tests of the second DAQ I changed the frequency from the frequency
 generator and scaling factor  that is used
 |  | 
| 11 | Sat Jun  1 18:55:18 2024 | Carsten | Analysis | Go4 Analysis Version for DAQ1 (atplaser) | Small enhancements and minor corrections / bug fixes are made to the original version
 of the online GO$ for DAQ1.
 |  | 
| 10 | Sat Jun  1 18:14:01 2024 | Carsten | Analysis | DAQ and analysis directories | [U][B]DAQ1 (original):[/B][/U] -user atplaser
 -runs on lxg1297
 |  | 
| 9 | Sat Jun  1 18:00:20 2024 | Carsten | DAQ | ReadOut Trigger im Timer Scaler changed because of faulty signal from frequency generator | Old readout trigger (~200 Hz) in 1MHz Scaler showed two nearby peaks indicating a jitter
 of the readout frequency or same faulty output
 |    | 
| 8 | Fri May 31 16:49:52 2024 | Carsten | DAQ | Wiring DAQ 1 | Started to Check the wiring. 
 Status as of  2024-05-31
 |  | 
| 7 | Fri May 31 15:53:06 2024 | Ruijiu Chen | DAQ | run NTCAP system | NTCAP is working again. Connect the RF cable and the scaler cables
 to NTCAP system.
 |  | 
| 6 | Wed May 29 10:54:21 2024 | Shahab, Jan | General | Gas Target Events with new timing system | Using the new timing system, it is possible to grab any events as pulses, but also as
 long inter event gates under GENESYS. Many
 |      | 
| 5 | Wed May 29 09:49:30 2024 | Rodolfo | Laser | Laser-beam profile | Using a telescope (Plan-concave lens -500 mm and Plan-convex lens +1000 mm) after the
 pulsed-dye laser system we got the following
 |    | 
| 4 | Mon May 27 13:37:31 2024 | Jan | DAQ | MesyTec Shaper Control | For the 2nd DAQ the MesyTec Shaper ([URL=https://elog.gsi.de/esr/NucAR+lab/230418_110112/MesyTec_MSCF16-F-V.pdf]MSCF manual[/URL]) is used as a spectroscopy amplifier
 and CFD.
 |    | 
| 3 | Thu Apr 25 16:34:05 2024 | Rodolfo | Laser | Efficiency of Fluorescein27 | The figure below shows the efficiency of Fluorescein27 vs. Wavelength. Each point was averaged over
 500 laser shots.
 |   | 
| 2 | Tue Apr 16 16:28:21 2024 | Danyal | General | Great | Thanks Jan! |  | 
| 1 | Tue Apr 16 14:42:37 2024 | Jan | General | G-PAC Proposal |  |   |