| ID | 
Date | 
Author | 
Category | 
Subject | 
Text | 
  | 
| 
  84  
 | 
Tue Mar 25 09:25:07 110 | 
Ken Ueberholz | Detectors | XUV-Detector Rates | Count rates were measured for the XUV-
 
detector for a full circle:
 
Cathode: -5V, shields: 0V, main coil 8A, | 
      | 
| 
  237  
 | 
Wed Apr 20 10:09:50 2022 | 
Rodolfo | Laser | Optic with spots | The average energy produced by the laser along 
the beamtime was about 10 mJ/puls. After 
about two weeks of beamtime some of the optics, | 
     | 
| 
  236  
 | 
Sat Apr 16 04:19:05 2022 | 
Patrick, Simon, Alex, Sarper | Runs | Runs night + morning shifts  16.04. | Runs of the night shift 16.04. after changing 
the wavelength range. The new range is 320.12 
- 319.859 nm. The step size was decreased | 
   | 
| 
  235  
 | 
Fri Apr 15 20:43:18 2022 | 
Konstantin | Laser | Dye changed | Dye of the laser system has been changed. 
UV laser power increased from ~8 mJ to 11 
mJ pulse energy. | 
   | 
| 
  234  
 | 
Fri Apr 15 12:53:25 2022 | 
Max, Danyal | Laser | checked laser pulse energy after change in wavelength range | previous scan range: 
 
the laser pulse energy @ 320.62 nm  was ~11 
mJ 
 | 
   | 
| 
  233  
 | 
Fri Apr 15 12:42:18 2022 | 
Max, Danyal, Carsten, Konstantin, Patrick, Simon | General | laser wavelength range changed: 320.12 - 319.60 nm | 
 
Runs of the morning/afternoon/night shift. 
The wavelength range is now 320.12 - 319.60 | 
    | 
| 
  232  
 | 
Thu Apr 14 22:44:40 2022 | 
Patrick, Simon, Danyal, Max | Runs | Runs night shift 14./15.04. | Runs of the night shift. The wavelength range 
is 320.62 - 320.10 nm.
 
 | 
   | 
| 
  231  
 | 
Thu Apr 14 19:46:46 2022 | 
Sarper & Konstantin & Rodolfo | Laser | Laser power | We checked the laser power before starting 
the new scan range.
 
 | 
   | 
| 
  230  
 | 
Thu Apr 14 17:47:55 2022 | 
Carsten, Rodolfo, Konstantin, Sarper | Runs | Afternoon shift runs 14.04 changing the wavelength scan range | The new scan range is changed from [B]321.12 
- 320.60 nm to 320.62 - 320.10 nm[/B]
 
 | 
   | 
| 
  229  
 | 
Thu Apr 14 12:38:38 2022 | 
Sebastian | Detectors | XUV / Cu checkups | Check regularly if the connection to the XUV 
and Cu detectors are still active and the 
detectors move in (after injection) and out | 
   | 
| 
  228  
 | 
Thu Apr 14 12:27:18 2022 | 
Sebastian, Max | Detectors | Connection to XUV / Cu reestablished! | The connection to the XUV and Cu were lost 
during the nightshift.
 
 | 
   | 
| 
  227  
 | 
Thu Apr 14 10:57:11 2022 | 
Carsten | Accelerator | Maintanence UNILAC / Op training | We had no beam since about 1000. Also, we 
are still trying to find out why we have 
such a low intensity in the expereimnt. 
 | 
   | 
| 
  226  
 | 
Thu Apr 14 08:41:42 2022 | 
Carsten | Accelerator | Low intensity at the experiment after series of failures during the night | After the recovery from the failures during 
the night, we see much lower intensity at 
the experiment (measured with the particle | 
   | 
| 
  225  
 | 
Thu Apr 14 06:29:18 2022 | 
Max, Sebastian | Accelerator | lesser ESR injection intensity | We now have less ESR injection intensity (see 
screenshot attached).
 
 | 
    | 
| 
  224  
 | 
Wed Apr 13 15:41:29 2022 | 
Rodolfo & Sarper | Laser | Laser status | We have checked the power of the laser and 
optimized it "a bit".
 
 | 
   | 
| 
  223  
 | 
Wed Apr 13 14:29:04 2022 | 
Carsten, Rodolfo, Sarper, David, Pierre-Michel, Max, Sebastian | Runs | Afternoon - morning shift runs 13.+14.04. | Changed the wavelength scan range from [B]321.62 
- 321.10 nm[/B] to [B][B]321.12 - 320.60[/B] 
nm[/B] after one scan.
 | 
       | 
| 
  222  
 | 
Tue Apr 12 21:56:54 2022 | 
Volker, David, Pierre-Michel & Danyal, Alexandre, Sebastian | General | new laser wavelength scan range 321.62 - 321.10 nm | New wavelength scan range is [B]321.62 - 321.10 
nm[/B]. Before the new scan range (compare 
old scan range in [URL=https://elog.gsi.de/esr/E142/215] | 
     | 
| 
  221  
 | 
Tue Apr 12 17:01:30 2022 | 
Konstantin | Detectors | Voltage PMT North | Voltage of PMT North has been increased from 
2500 V to 2600 V. Additionally the threshold 
was adapted from 25 mV to 12 mV. | 
   | 
| 
  220  
 | 
Tue Apr 12 15:05:56 2022 | 
Rodolfo | Laser | Laser status | We have checked the laser power at 14:15 *
 
 
- Pump energy after Q-Ray @ 532nm: 559 mJ.
 | 
   | 
| 
  219  
 | 
Tue Apr 12 14:52:19 2022 | 
Rodolfo | Detectors | PMT Nord, Middle. Datasheets | Attached are the datasheets for the solar-blind 
PMTs.
 
 | 
    | 
| 
  218  
 | 
Tue Apr 12 14:11:52 2022 | 
Ragan | Detectors | Sideband in Schottky | > Sometimes we see a broad sideband at lesser 
Schottky frequencies to the high-intensity 
thorium beam.
 | 
   | 
| 
  217  
 | 
Tue Apr 12 13:24:40 2022 | 
Alexandre, Sebastian | Detectors | Sideband in Schottky | Sometimes we see a broad sideband at lesser 
schottky frequencies to the high intensity 
thorium beam.
 | 
     | 
| 
  216  
 | 
Tue Apr 12 12:41:20 2022 | 
Alexandre, Sebastian | Accelerator | no beam in SIS18 |   | 
   | 
| 
  215  
 | 
Tue Apr 12 04:43:29 2022 | 
Alexandre, Kristian, Sebastian | Runs | 322.01 - 321.60 nm  | We scan the wavelength scan range is  322.01 
- 321.60 nm 
 
 | 
    | 
| 
  214  
 | 
Mon Apr 11 22:33:16 2022 | 
Carsten | General | Backup of data to the external RAID disks |   | 
   | 
| 
  213  
 | 
Mon Apr 11 22:14:51 2022 | 
Konstantin, Sarper | Laser | Changing of the wavelength range | It is planned to change the scan range of 
the laser wavelength from 322.41 - 322.0 
nm to 322.01 - 321.6 nm @ 4 a.m.  | 
   | 
| 
  212  
 | 
Mon Apr 11 18:57:18 2022 | 
Rodolfo | Laser | New flashlamps | We (Sarper, Konstantin and me) have changed 
the old flashlamps (with ~ 26 Mio. shots) 
by new "fresh" ones. We have also exchange | 
    | 
| 
  211  
 | 
Mon Apr 11 18:46:31 2022 | 
Rodolfo, Sarper and Konstantin | Laser | Alignment-check | We cross-checked the laser-ion alignment. 
The horizontal position was optimized. However, 
 | 
    | 
| 
  210  
 | 
Mon Apr 11 14:45:11 2022 | 
Paul, Konstantin, Sarper, David | Runs | Afternoon shift runs 11.04 | We continue scan the new wavelength scan range 
is  322.41 - 322.00 nm 
 
 | 
   | 
| 
  209  
 | 
Mon Apr 11 12:50:22 2022 | 
Kristian | Accelerator | Th lifetime | We used the counter to measure the lifetime. 
During the 30 min scan time dropped only 
slowly. 
 | 
    | 
| 
  208  
 | 
Mon Apr 11 11:05:35 2022 | 
Sebastian, Kristian | Laser | Check laser power | we checked the UV laser power at 10:15 am. 
It was only 8.5 - 9 mJ. Pumping power was 
460 mJ, which is the maximum output of the | 
   | 
| 
  207  
 | 
Mon Apr 11 10:31:07 2022 | 
Kristian, Sebastian, Alexandre, Danyal, Paul | General | new laser wavelength scan range 322.41 - 322.00 nm | new wavelength scan range is  322.41 - 322.00 
nm 
 
 | 
   | 
| 
  206  
 | 
Mon Apr 11 10:17:01 2022 | 
Danyal | Accelerator | Interlock  - reset  | 10:17 
 
Wie heutemorgen in der Sitzung (morning briefing) 
vereinbart, 
 | 
   | 
| 
  205  
 | 
Mon Apr 11 07:54:37 2022 | 
Danyal, Alexandre, Kristian, Sebastian | Accelerator | Jump in E-cooler voltage | We saw a jump in the E-cooler voltage at about 
7:50. 
 
jump from 209619 to 209610 V 
 | 
   | 
| 
  204  
 | 
Mon Apr 11 04:26:53 2022 | 
Patrick | DAQ | Next wavelength range | Measurements with the next wavelength scan 
range should be started roughly around 10:00 
am.
 | 
   | 
| 
  203  
 | 
Sun Apr 10 22:47:05 2022 | 
Patrick, Volker, Prannay, Danyal, Alexandre, Sebastian, Kristian | Runs | Runs night shift 10.04  + morning shift 11.04 | These are the measurement runs of the night 
shift (10/11.04.).
 
 | 
   | 
| 
  202  
 | 
Sun Apr 10 15:29:53 2022 | 
Rodolfo | Laser | Status - Laser system | We have checked the laser system.
 
 
- The pump laser has ~ 23Mio shots. The total | 
   | 
| 
  201  
 | 
Sun Apr 10 15:29:37 2022 | 
Carsten,Sarper | Accelerator | Ecooler failed | Ecooler failed @15:23 Carsten is fixing it
 
Fixed @16:00 | 
   | 
| 
  200  
 | 
Sun Apr 10 15:16:19 2022 | 
Carsten, Rodolfo, Sarper | Runs | Runs afternoon shifts 10.04 | These are the measurement runs of the afternoon 
shift (10.04.).
 
 | 
   | 
| 
  199  
 | 
Sun Apr 10 15:08:42 2022 | 
Carsten | DAQ | Backup of Data from 1rst Scanrange to extrenal RAID1 HDDs | I just performed a backup of all data (+log-Files) 
that were collected until now to the two 
RAID HDDs.
 | 
   | 
| 
  198  
 | 
Sun Apr 10 13:17:53 2022 | 
Kristian, Max | DAQ | Wavemeter readout | We changed the Labview code in order to reduce 
the amount of data that was overfilling the 
WM data queue. We removed the temperature | 
   | 
| 
  197  
 | 
Sun Apr 10 11:11:18 2022 | 
Kristian, Max | Laser | additional WM readout | since the WM readout still does not work reliably, 
I started a python script at atppc023 that 
saves timestamp and wavemeter frequency. | 
   | 
| 
  196  
 | 
Sun Apr 10 08:53:26 2022 | 
Kristian, Max | Laser | Timing verified | timing is still good.
 
 
we checked the relative timing between laser | 
   | 
| 
  195  
 | 
Sun Apr 10 06:39:16 2022 | 
Kristian, Marek, Max | Runs | runs morning shift 10.04. | These are the measurement runs of the morning 
shift (10.04.).
 
 | 
   | 
| 
  194  
 | 
Sun Apr 10 06:34:49 2022 | 
Kristian, Max | DAQ | Wavemeter Readout | Problem solved. Ignore this entry
 
 
 | 
   | 
| 
  193  
 | 
Sun Apr 10 04:37:29 2022 | 
Patrick | DAQ | Next wavelength range | Measurements with the next wavelength scan 
range should be started roughly around 12:00 
noon. The plan is to go to lower wavelengths | 
    | 
| 
  192  
 | 
Sun Apr 10 04:07:53 2022 | 
Patrick | DAQ | go4 shows wrong frequencies and wavelengths | [B]Update (ID:194):[/B] In labview "Stop All", 
close labview, stop python script (ID:197), 
close wavemeter. Then restart wavemeter, | 
      | 
| 
  191  
 | 
Sun Apr 10 03:56:56 2022 | 
V. Hannen | Detectors | Low rate in PMT South | PMT south shows a much lower count rate than 
observed in elog entry 170.
 
 | 
    | 
| 
  190  
 | 
Sun Apr 10 02:20:17 2022 | 
Patrick | DAQ | New mbs startup guide | Connecting to the rfio server and opening 
a file is included in '@startup' now. Since 
mbs crashes reliably every 3 or 4 runs, it | 
   | 
| 
  189  
 | 
Sat Apr  9 19:50:40 2022 | 
Carsten |   | New thresholds for ADC of DAQ2 (starting with file0117) | unsigned short caen_V785AH_threshold_mod0 
[32] = 
 
                {0x0100 , 0x0004 , 0x0003 | 
   | 
| 
  188  
 | 
Sat Apr  9 19:09:57 2022 | 
Rodolfo | Laser | Optimization - Resonator of dye laser  | I have tuned a bit the resonator of the dye 
laser. 
 
 | 
   | 
| 
  186  
 | 
Sat Apr  9 16:55:19 2022 | 
Rodolfo | Laser |   | 16:15 I have changed the dye (of both oscillator 
and amplifier).
 
 | 
   | 
| 
  185  
 | 
Sat Apr  9 15:47:21 2022 | 
Carsten, Rodolfo, Sarper, Patrick | Runs | Afternoon and night shift runs | After the morning shift, we got new increment 
(nm) for the laser scanning. We are now scanning 
2 nm of wavelength range in 5 big steps (5 | 
   | 
| 
  184  
 | 
Sat Apr  9 14:15:44 2022 | 
Kristian, Max, Konstantin | DAQ | Change in LV code | We changed the LV code so that the Cobra Set 
Wavelength is permanently sent. Before it 
was only sent once at the beginning of each | 
   | 
| 
  183  
 | 
Sat Apr  9 13:45:42 2022 | 
Max, Kristian | Runs | first scans with LV problem solved! | This is the first time with LV buffer problem 
solved (see Elog ID: 184). Before the PMT 
signals were out of sync with the wavelength | 
   | 
| 
  182  
 | 
Sat Apr  9 06:42:23 2022 | 
Kristian | Laser | Laser step size | We changed the laser step size to 0.002 nm 
which corresponds to 5.75 GHz. | 
   | 
| 
  181  
 | 
Sat Apr  9 06:25:22 2022 | 
Max, Kristian | Runs | morning shift runs | DAQ1: lxg1927:/data.local2/BeamTimeTh22/
 
DAQ2: lxg0155:/data.local2/E142/
 
LMD file name DAQ1:229Thor
 | 
   | 
| 
  180  
 | 
Sat Apr  9 05:26:25 2022 | 
Patrick | Laser | Wavemeter | The wavemeter still shows large spikes from 
time to time. The exposure time is still 
at 34ms. | 
    | 
| 
  179  
 | 
Sat Apr  9 05:00:11 2022 | 
Carsten | General | Draft Docs and cheat sheet | Please check the attached files,
 
 
and correct or enhance them.
 | 
     | 
| 
  178  
 | 
Sat Apr  9 04:40:01 2022 | 
Patrick | DAQ | Bugged Trending plots? | The problem from yesterday (ID:147) still 
persists. All the plots (not only the labview 
stuff) comes in too late in go4. However, | 
    | 
| 
  177  
 | 
Sat Apr  9 04:01:35 2022 | 
Carsten |   | 2nd DAQ | Switched the CAEN TDC back on and changed 
the threshoolds for the ADC to: 
 
 | 
   | 
| 
  176  
 | 
Sat Apr  9 03:57:10 2022 | 
Carsten | Accelerator | New Cooler voltage after the crash of the cooler | After the restart of the cooler the Ohmlabs 
divider showed a voltage 16V lower than before. 
This could also be seen as a slight shift | 
   | 
| 
  175  
 | 
Sat Apr  9 03:46:31 2022 | 
Carsten | Accelerator | Slits in TE line | Changed GTEDSAH Rechts to 10mm | 
   | 
| 
  174  
 | 
Sat Apr  9 03:27:47 2022 | 
Patrick | DAQ | Laser scanning | DAQ1: lxg1927:/data.local2/BeamTimeTh22/
 
DAQ2: lxg0155:/data.local2/E142/
 
LMD file name DAQ1:229Thor
 | 
    | 
| 
  173  
 | 
Sat Apr  9 02:55:38 2022 | 
Carsten | Accelerator | Kühlerausfall und Gründe dazu | Wir hatten heute Abend eine Kühlerausfall, 
der uns mehrere Stunden beschäftigt hat, 
weil  gem. Anzeige in Device Control kein | 
   | 
| 
  172  
 | 
Fri Apr  8 23:56:07 2022 | 
Patrick | DAQ | DAQ improvement | As requested in ID:146 and ID:147, the laserscan 
tab was improved:
 
[LIST]
 | 
   | 
| 
  171  
 | 
Fri Apr  8 22:26:44 2022 | 
Patrick | DAQ | Spike in all detectors | In the previous shift, there was a spike in 
all detectors in the same scan step. Apparently, 
this was NOT a resonance since in the next | 
   | 
| 
  170  
 | 
Fri Apr  8 17:57:45 2022 | 
Konstantin | DAQ | Laser scanning | 
 
 
[TABLE border="1"]
 | 
    | 
| 
  169  
 | 
Fri Apr  8 17:44:25 2022 | 
Konstantin | Calibration | Final Timing | We measured the timing for the laser bunch 
synchronization. A summary is given below.
 
 | 
   | 
| 
  168  
 | 
Fri Apr  8 15:45:24 2022 | 
YL |   | Spectrum manipulation 5 | See attached | 
    | 
| 
  167  
 | 
Fri Apr  8 15:42:20 2022 | 
Yuri | Accelerator | Slits in the TE line | At around 15:00, The slits in the TE line 
were changed to :
 
TE4DS5HL -12 mm -> -20 mm
 | 
   | 
| 
  166  
 | 
Fri Apr  8 15:39:36 2022 | 
all | Runs | first data after timing set up | started taking data again
 
 
started at 15:40  
 | 
   | 
| 
  165  
 | 
Fri Apr  8 14:46:02 2022 | 
Danyal & Sebastian & Volker & Ken | Detectors | high dark countrate Cu PMT solved | The problem - "discovered" by Volker and Ken 
- of the high dark countrate of the Cu PMT 
has been solved. 
 | 
     | 
| 
  164  
 | 
Fri Apr  8 14:43:50 2022 | 
Wilfried | DAQ | laser timing - Q-switch start | below are my notes concerning laser timing 
/ Q-switch start | 
    | 
| 
  163  
 | 
Fri Apr  8 14:14:28 2022 | 
Kristian | DAQ | Delay of bunch RF | The bunch HF is now fed through a delay to 
shift the stop trigger relative to the PMT 
signal. The delay generator is placed directly | 
   | 
| 
  162  
 | 
Fri Apr  8 14:12:57 2022 | 
morning shift | DAQ | timing for h=1 | The timing for h=1 has been set up 
 
 
The time window is now twice as large as | 
    | 
| 
  161  
 | 
Fri Apr  8 12:04:13 2022 | 
Ragan | Runs | NTCAP started | NTCAP is started for the data taking.
 
 
Carrier Frequency: 242.02 M
 | 
    | 
| 
  160  
 | 
Fri Apr  8 11:16:57 2022 | 
Kristian | Laser | Wavemeter | We changed the fiber to the wavemeter. The 
third fiber worked best and we could operate 
the WM at low exposure times, i.e., below | 
     | 
| 
  159  
 | 
Fri Apr  8 11:02:04 2022 | 
DW, YL, RSS, SL |   | Change h=2 to h=1 | After changing to h=1
 
The frequency manipulation #5: f=242.021487MHz 
(124th harmonic) [no bunching]
 | 
   | 
| 
  158  
 | 
Fri Apr  8 10:44:55 2022 | 
Ragan, Yuri, Danyal, Sergey | Runs | NTCAP stopped | NTCAP is stopped. Sergey will now set an ion 
bunch instead of two in the ESR.
 
 | 
   | 
| 
  157  
 | 
Fri Apr  8 09:39:04 2022 | 
Ragan | Runs | NTCAP started | Carrier Frequency: 242.2 M
 
IQ Rate (S/s): 10 M
 
Reference level (dBm): -30
 | 
   | 
| 
  156  
 | 
Fri Apr  8 09:33:52 2022 | 
Dayshift 1 8.4.2022 |   | Reference Measurements | DAQ1: lxg1297:/data.local2/BeamTimeTh22/
 
DAQ2: lxg0155:/data.local2/E142/
 
LMD file name DAQ1:229Thor
 | 
   | 
| 
  155  
 | 
Fri Apr  8 09:32:32 2022 | 
Ragan | Runs | NTCAP stopped | Since there is no beam, NTCAP is stopped.
 
 
Carrier Frequency: 242.2 M
 | 
   | 
| 
  154  
 | 
Fri Apr  8 09:13:38 2022 | 
Ragan, Danyal | Accelerator | No beam | Cathode is being changed at the ion source. 
So no beam for the moment.
 
 | 
   | 
| 
  153  
 | 
Fri Apr  8 08:40:06 2022 | 
Sebastian | DAQ | DAQ improvement | Concerning ID 146 a new histogram for accumulating 
data (no cleaning histograms) can be found 
under
 | 
   | 
| 
  152  
 | 
Fri Apr  8 08:17:52 2022 | 
Sebastian | Laser | Laser Ext Trig is working | We (Max, Kristian, Danyal, me) fixed the "external 
trigger" problem.
 
 | 
       | 
| 
  151  
 | 
Fri Apr  8 06:53:29 2022 | 
Ragan | Runs | NTCAP data taking | NTCAP is taking data since 07.04.2022 23:53:47
 
 
Carrier Frequency: 242.2 M
 | 
    | 
| 
  150  
 | 
Fri Apr  8 05:46:35 2022 | 
V. Hannen | Detectors | High background rate in Cu PMT points to light leak in ESR or active pressure gauge | The background rate of the Cu mirror PMT is 
only about 300 cps when the mirror is in 
the parking position but ca. 7000 cps when | 
   | 
| 
  149  
 | 
Fri Apr  8 03:54:39 2022 | 
Patrick | Laser | Q-switch window | The solution of increasing the Q-switch window 
from 2us to 5us described in [I]ID:148[/I] 
fulfills its purpose, but the window is unstable.
 | 
    | 
| 
  148  
 | 
Fri Apr  8 02:14:05 2022 | 
Wilfried | Laser | Laser timings adjusted  | Since we use the middle window as a reference 
for the laser pulse timing, the fact that 
the window is 2.25 m behind the target has | 
   | 
| 
  147  
 | 
Fri Apr  8 01:48:57 2022 | 
Patrick | DAQ | Bugged Trending plots? | The wavelengths/frequencies shown in the graph 
in the labview window do not match the wavelengths/frequencies 
in go4.
 | 
    | 
| 
  146  
 | 
Fri Apr  8 01:18:51 2022 | 
Ken Ueberholz | DAQ | TODO: DAQ improvements  | For somebody who knows the DAQ1 code: Go4 
resets the histograms with each laser step. 
Can there be additional histograms where | 
   | 
| 
  145  
 | 
Fri Apr  8 00:06:37 2022 | 
Patrick | Laser | Connection to the laser pc | To login to the laser pc use the command
 
 
xfreerdp -g 1920x1080 -u atplaser atppc023.campus.gsi.de | 
   | 
| 
  144  
 | 
Thu Apr  7 22:18:06 2022 | 
Night shift 7.4-8.4 |   | Next scans | DAQ1: lxg1927:/data.local2/BeamTimeTh22/
 
DAQ2: lxg0155:/data.local2/E142/
 
LMD file name DAQ1:229Thor
 | 
   | 
| 
  143  
 | 
Thu Apr  7 19:13:26 2022 | 
Konstantin, Patrick | Calibration | How to find the correct laser position | In order to achieve the correct laser ion 
beam superposition one needs to do the following 
steps:
 | 
   | 
| 
  142  
 | 
Thu Apr  7 16:50:22 2022 | 
Konstantin | Calibration | Spacial laser-ion overlap | Laser-ion-beam overlap was achieved after 
measuring the ion beam positions (compare 
[URL=https://elog.gsi.de/esr/E142/83]Entry | 
   | 
| 
  141  
 | 
Thu Apr  7 16:45:10 2022 | 
  | Detectors | Detector distances | Relativ distances of the detector  | 
    | 
| 
  140  
 | 
Thu Apr  7 15:18:25 2022 | 
Danyal & Sebastian | DAQ | DAQ channels | We checked again the cables for both DAQs 
 
(on Carsten's kind request) 
 | 
   | 
| 
  139  
 | 
Thu Apr  7 14:13:27 2022 | 
Danyal | Detectors | new PMT south | I have replaced the PMT south (looks at the 
gas target). 
 
in the hope that this will produce less dark | 
   | 
| 
  138  
 | 
Thu Apr  7 14:04:36 2022 | 
Paul Indelicato | General | no beam until 17:00 | 10:10  There is a problem with the UNILAC. 
 
The repair will take several hours. (4 hrs?) | 
   | 
| 
  137  
 | 
Thu Apr  7 14:01:22 2022 | 
Sebastian, Max | Detectors | Mesytec reset | The Mesytec crate was turned off and on again.
 
 
It is again possible to communicate with | 
   | 
| 
  136  
 | 
Thu Apr  7 11:11:31 2022 | 
Max, Sebastian | Laser | dye changed, pulse energy measurements | We changed the dye at 10am.
 
 
We increased the pump power for the dye Laser | 
   | 
| 
  135  
 | 
Thu Apr  7 10:10:31 2022 | 
Ragan, Yuri, Danyal | Accelerator | Problem in UNILAC | The power supply of the UNILAC magnet is broken. 
The repair will take up to 6-7 hours. | 
   | 
| 
  134  
 | 
Thu Apr  7 09:58:55 2022 | 
Ragan, Yuri, Danyal | Accelerator | Kicker problem solved | The kicker problem is solved and the beam 
is back in the ESR. | 
   | 
| 
  133  
 | 
Thu Apr  7 09:23:18 2022 | 
Sebastian | DAQ | DAQ1 & DAQ2 - 1kHz and delayed cs removed | Questionable 1kHz and delayed common stop 
signals has been removed.
 
 | 
   | 
| 
  132  
 | 
Thu Apr  7 08:43:19 2022 | 
Danyal, Yuri, Ragan | Accelerator | again no injection in ESR - ESR kicker problem | we had no injections in the ESR again 
 
 
it turned out that the injection kicker of | 
   | 
| 
  131  
 | 
Thu Apr  7 08:14:03 2022 | 
Yuri & Danyal | General | 229Th89+ line in Schottky  | GLOBAL calculations for the "content" of the 
Th line in the Schottky spectrum
 
(380 MeV/u exit energy, equilibrium charge | 
   | 
| 
  130  
 | 
Thu Apr  7 07:40:09 2022 | 
Danyal, Max, Sebastian | DAQ | No beam and DAQs crashed | since shortly after 7 AM we have no beam in 
the ESR
 
(reason unclear) 
 | 
   | 
| 
  129  
 | 
Thu Apr  7 05:55:17 2022 | 
Ken Ueberholz | Detectors | XUV-Detector Mesytec crash | When the XUV-detector has to shut down unexpectedly, 
it can happen that the Mesytek power supply 
crashes and can't be 
 | 
   | 
| 
  128  
 | 
Thu Apr  7 05:44:40 2022 | 
Pierre-Michel, Ragan | DAQ | NTCAP settings | Carrier Frequency: 244M
 
IQ Rate: 10M
 
reference Level: -30
 | 
   | 
| 
  127  
 | 
Thu Apr  7 05:23:50 2022 | 
Patrick | DAQ | How To Start MBS for DAQ2 (Carstens DAQ) | (MBS server should be running at the LXG0155 
machine. If not you will get an error message. 
Then start the server - last command in console)
 | 
   | 
| 
  126  
 | 
Thu Apr  7 04:48:58 2022 | 
  |   | ToDo's for 07.04. | Some ToDo's for the next shift:
 
- Work on  Laser stabilisation 2nd stage 
in the ESR
 | 
   | 
| 
  125  
 | 
Thu Apr  7 04:10:18 2022 | 
Ken Ueberholz | DAQ | Corrected cabling DAQ 2 | A lot of cables were wrongly or not 
 
connected to the DAQ 2. 
 
 | 
    | 
| 
  124  
 | 
Thu Apr  7 03:44:21 2022 | 
Patrick | DAQ | Measurement procedure | [U][B]Measurement procedure[/B][/U]
 
 
 | 
   | 
| 
  123  
 | 
Thu Apr  7 03:09:36 2022 | 
Carsten | DAQ | NTCAP running since 02:40 |   | 
   | 
| 
  122  
 | 
Thu Apr  7 03:06:15 2022 | 
Patrick | DAQ | DAQ file sizes | The file sizes of DAQ1 (229Thor) and DAQ2 
(229Th, "Carstens DAQ") are different.
 
During 20min of laser scanning DAQ1 saves | 
   | 
| 
  121  
 | 
Thu Apr  7 02:32:29 2022 | 
all on shift | Runs | First Scans 322-320nm | Laser scanning from 322-320 steps 0.01nm/s
 
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
 
DAQ2: lxg0155:/data.local2/E142/
 | 
   | 
| 
  120  
 | 
Thu Apr  7 01:19:05 2022 | 
Patrick | Laser | Laser scanning | We did some laser scans. The wavemeter shows 
spikes during scans. This may be just while 
the frequency changes. See picture. | 
       | 
| 
  119  
 | 
Thu Apr  7 00:36:36 2022 | 
Uwe | DAQ | Threshholds and shaping Mesytec for ADC |   | 
    | 
| 
  118  
 | 
Thu Apr  7 00:15:11 2022 | 
Patrick | Laser | Laser timing adjusted | We have adjusted the laser timing using PMT-middle. 
Bunching freq. is 3.90362 MHz.
 
 | 
      | 
| 
  117  
 | 
Wed Apr  6 22:29:55 2022 | 
Patrick | Accelerator | ECooler voltages | ECooler voltages:
 
 
[B]06.04:[/B]
 | 
   | 
| 
  116  
 | 
Wed Apr  6 21:03:25 2022 | 
Ken Ueberholz  | Detectors | Xuv-Detektor, further information  | Some further information regarding the 
 
xuv-detector and its components. | 
    | 
| 
  115  
 | 
Wed Apr  6 19:39:08 2022 | 
Ken Ueberholz  | Detectors | LED installed  | The UV-LED of Münster was connected over 
 
channel 8 to the Messhütte. 6V are 
 
supplied by the power supply (picture 2). | 
     | 
| 
  113  
 | 
Wed Apr  6 18:51:33 2022 | 
Danyal & Carsten | Accelerator | target bump | This afternoon, we found out that Sergey has 
also implemented a "target bump". 
 
Although we do not use a target, this bump | 
   | 
| 
  112  
 | 
Wed Apr  6 16:01:07 2022 | 
BL | Accelerator | ESR settings before adjusting laser: 2022-04-06_14-37-30_ESR_EXP22_02_238U92_Brandau_30March22.C1.txt  | # Version: 1.2
 
# Comment: ParamModi Export at 2022-04-06T14:37:35.957978+02:00[Europe/Berlin]
 
# Context: ESR_EXP22_02_238U92_Brandau_30March22.C1
 | 
   | 
| 
  111  
 | 
Wed Apr  6 15:19:29 2022 | 
Svea, Shahab | Accelerator | 5s delay implemented before injection | Operators have informed us that for better 
stability and operation (and cooperation 
with HHT), and after discussing with Sergey, | 
   | 
| 
  110  
 | 
Wed Apr  6 14:52:29 2022 | 
Rodolfo | Runs | Scraping the ion beam | We are T-scraping the ion beam.  | 
   | 
| 
  108  
 | 
Wed Apr  6 13:32:57 2022 | 
Max, Sebastian | DAQ | LabView Cobra Channel List | channel 0: linux time (epoch time)
 
channel 1: ms timer
 
channel 2: E-Cooler Voltage(Ohmlabs)
 | 
   | 
| 
  107  
 | 
Wed Apr  6 09:30:53 2022 | 
Danyal | Accelerator | ESR is blue again - errors are fixed | The ESR is "blue" again. 
 
 
The errors (scu related) have been fixed. | 
   | 
| 
  106  
 | 
Wed Apr  6 08:31:54 2022 | 
Danyal & Regina | General | ESR had some errors | Regina noticed that the ESR was "red" 
 
 
This means that there are some errors. 
 | 
   | 
| 
  105  
 | 
Wed Apr  6 08:06:02 2022 | 
Danyal | DAQ | stopped NTCAP data acquisition | I stopped the NTCAP data acquisition at 8 
AM.
 
 | 
   | 
| 
  104  
 | 
Tue Apr  5 23:55:11 2022 | 
Carsten | Runs | Long time measurment with NTCAP | Settings:
 
Center: 2242.5MHz
 
RefLevel: -25dB
 | 
   | 
| 
  103  
 | 
Tue Apr  5 23:07:34 2022 | 
Carsten | Accelerator | 229Th89+ identification after re-tuning of the ESR | The ESR was retuned this aftenoon since the 
it was unclear if we still sit the 229Th89+ 
line.
 | 
   | 
| 
  102  
 | 
Tue Apr  5 20:58:53 2022 | 
Wilfried | Laser | Trigger for Laser | Electronic for the timing of the laser puls 
optimized. The problem with the broad photodiode 
histograms is solved.
 | 
     | 
| 
  101  
 | 
Tue Apr  5 20:16:13 2022 | 
Rodolfo | Laser | Discriminator for HF reference signal | For the laser trigger we need to generate 
a few ns NIM-puls from the HF reference signal. 
Therefore we (Wilfried, Konstantin and me) | 
     | 
| 
  100  
 | 
Tue Apr  5 15:25:47 2022 | 
Sebastian, Kristian, Pierre-Michel | DAQ | Go4 Conditions for "TDC_stacked_Sig/Ref" histogram | First version of TDC_stacked histograms implemented.
 
 
Under folder "Conditions" you can now integrate | 
   | 
| 
  99  
 | 
Tue Apr  5 14:51:51 2022 | 
Kristian | DAQ | dark counts | We shielded the PMTs, i.e., the N2 tube, which 
improved the Cu PMT background rate.
 
 | 
   | 
| 
  98  
 | 
Tue Apr  5 13:28:06 2022 | 
Kristian | DAQ | Spikes in PMT signal | Occasionally, we noticed a spike in the PMT 
signal. No ions in the ring and at PMT Cu 
was not even supplied with HV. Has to be | 
    | 
| 
  97  
 | 
Tue Apr  5 13:24:40 2022 | 
Kristian | Detectors | PMT Cu | HV cable was in wrong channel in power supply. 
Now it is back to the correct position. Checked 
connection at the PMT. | 
   | 
| 
  96  
 | 
Tue Apr  5 12:17:12 2022 | 
Ragan, Rui-Jiu | DAQ | Instructions to start and stop NTCAP | Attached are the instructions to start and 
stop data taking using the NTCAP. | 
       | 
| 
  95  
 | 
Tue Apr  5 11:14:52 2022 | 
Sebastian, Kristian, Pierre-Michel | DAQ | PMT over wavelength histogram v2 | Today we finalized the PMT over wavelength 
histogram in:
 
 | 
   | 
| 
  94  
 | 
Tue Apr  5 09:45:47 2022 | 
Danyal & Yuri | Accelerator | beam is back in the ESR - a slit was moved in | And now also in English :) 
 
 
The ion beam is back in the ESR! 
 | 
   | 
| 
  93  
 | 
Tue Apr  5 09:39:01 2022 | 
Danyal & Yuri | Accelerator | Strahl ist wieder im ESR - Schlitz war reingefahren | Der Strahl ist wieder im ESR! 
 
Angeblich war einen Schlitz, oder sogar 2, 
reingefahren worden. 
 | 
   | 
| 
  92  
 | 
Tue Apr  5 08:00:50 2022 | 
Kristian & Sebastian | DAQ | MBS Channellist (so far) | MBS Channellist (time 5.4.2022)
 
 
0 Anode
 | 
   | 
| 
  91  
 | 
Mon Apr  4 19:05:57 2022 | 
Carsten | DAQ | Cooler Frequency Signals Ion current not connected until now | Up to now the signals for 
 
Ion Current
 
Cooler elctron current
 | 
   | 
| 
  90  
 | 
Mon Apr  4 17:26:48 2022 | 
Jan Glorius | DAQ | MSCF shaper settings | attached are the current shaper settings.
 
 
list of currently connected channels:
 | 
    | 
| 
  89  
 | 
Mon Apr  4 16:26:04 2022 | 
Carsten | DAQ | Backup Harddisk /External Disk | Since we cannot write on lustre, as a temproal 
solution two RAID 1 (mirror) USB hard disks 
 with cpacity 6 TB each are installed.
 | 
   | 
| 
  88  
 | 
Mon Apr  4 15:32:38 2022 | 
Kristiand & Sebastian | DAQ | Go4 Backup Code | Due to some code changes (PMT over wavelength 
historgram) Kristian and I backuped a working 
Go4 code in the current go4 folder on the | 
   | 
| 
  87  
 | 
Mon Apr  4 15:27:31 2022 | 
Sebastian | DAQ | PMT over wavelength histogram | Kristian and I tried to create a "PMT over 
wavelength" histogram under step2Anl/LabView.
 
Unfortunately, the code ist still in progress | 
   | 
| 
  86  
 | 
Mon Apr  4 09:10:44 2022 | 
Ragan, Yuri, Rui-Jiu | Runs | Measurement | NTCAP is taking data with the beam with a 
stop at manipulation 5 (taking all the fragments).
 
 | 
   | 
| 
  85  
 | 
Mon Apr  4 00:19:16 2022 | 
Ken Ueberholz | DAQ | FirstTests | New measurement:
 
Start: About 00:12
 
Stop: About 00:43
 | 
     | 
| 
  83  
 | 
Sun Apr  3 18:52:30 2022 | 
Konstantin | Calibration | Laser ion beam overlap | We measured vertical and horizontal ion beam 
positions. Unfortunately, the horizontal 
scraper positions can not be manipulated
 | 
   | 
| 
  82  
 | 
Sun Apr  3 18:41:11 2022 | 
Konstantin and Rodolfo | Laser | Beam Position Stabilization | We (Konstantin and I) have optimized the TEM 
"laser-beam position stabilization".
 
 | 
       | 
| 
  81  
 | 
Sun Apr  3 15:34:34 2022 | 
Ken Ueberholz  |   | Xuv-Detektor CFD-treshhold | I co controlled all detector signals and 
 
adjusted the threshold to be just above 
 
the noise. I will double check this when | 
   | 
| 
  80  
 | 
Sun Apr  3 13:05:08 2022 | 
Carsten | Laser | Distance of Partilce detecotrs wr.t. to electron cooler | As of Yuris post in the elog the positions 
(Info from H. Weick):
 
 | 
   | 
| 
  79  
 | 
Sun Apr  3 12:35:17 2022 | 
Max | DAQ | Go 4 manual | How to start go4?:
 
 
- go to console
 | 
      | 
| 
  78  
 | 
Sun Apr  3 12:31:10 2022 | 
Rui-Jiu Chen, Yuri, Carsten, Ragan | Runs | Measurement of beam with different circles | Parameters of NTCAP:
 
 
  1. IQ rate 20 MSamples/s
 | 
 6x   | 
| 
  77  
 | 
Sun Apr  3 11:40:48 2022 | 
Shahab, Yuri | Detectors | 245 MHz Resonator tuned to new frequency | resonance frequency changed from 243.2 MHz 
to 244.692 MHz by reducing the value of both 
of the plungers from 395 to 325. Now after | 
    | 
| 
  76  
 | 
Sun Apr  3 11:01:57 2022 | 
Rodolfo | Accelerator | Scrapers @ E-Cooler | We the mid- and night shift (Wilfried, Simon, 
Patrick, Ken and me) we not able to manipulate 
the scrapers in the E-Cooler.
 | 
   | 
| 
  75  
 | 
Sun Apr  3 10:35:20 2022 | 
Yuri (Sergey) | Accelerator | Dipole correction | The ramp for moving the beam to the central 
orbit
 
To be inserted into manipulation, GE01MU1 | 
   | 
| 
  74  
 | 
Sun Apr  3 10:28:25 2022 | 
Yuri (Helmut Weick) | Accelerator | Distance cooler - detector | Calculation from Helmut Weick
 
Distance e-cooler to particle detectors chamber:
 
“This is half of ESR plus distance target | 
   | 
| 
  73  
 | 
Sun Apr  3 00:41:45 2022 | 
Ken Ueberholz  |   | Xuv-Detektor Signale  | The Xuv-detector cabel channels seem to be 
 
wrongly labeled in the Messhütte (see 
 | 
    | 
| 
  72  
 | 
Sat Apr  2 23:16:55 2022 | 
Carsten |   | Cu mirror and CsI moved in and out of the ring | Ken tested the Cu-mirror and CsI disk Detector. 
They move without a problem and don't scrape 
the ion beam. | 
   | 
| 
  71  
 | 
Sat Apr  2 22:43:36 2022 | 
Carsten |   | Ring set-up for 229/89+, scraped and bunched, Iel=500mA at injection and 100mA during measurment | The ring SIS/ESR were set up for 229/89+. 
Pölease note that this not only 1e- Th but 
ponteially also the isobars of Ac (0e-1)/Pa(2e-, | 
   | 
| 
  70  
 | 
Sat Apr  2 18:28:47 2022 | 
Markus/Bernd | Accelerator | 	SIS setting complete for fragments in ESR: 2022-04-02_18-09-10_SIS18_FAST_E142_20220325_110213.C1.txt  | # Version: 1.2 # Comment: ParamModi Export 
at 2022-04-02T18:09:36.402593+02:00[Europe/Berlin] 
# Context: SIS18_FAST_E142_20220325_110213.C1 | 
   | 
| 
  69  
 | 
Sat Apr  2 18:19:47 2022 | 
Yuri | Accelerator | Scraping |   | 
    | 
| 
  68  
 | 
Sat Apr  2 18:17:31 2022 | 
Yuri | Accelerator | Photo-Identification | Just a photo of the ID | 
    | 
| 
  67  
 | 
Sat Apr  2 17:47:16 2022 | 
Ken Ueberholz  |   | Telefon, Gästehaus | My guesthouse phone number: 06159-71-2957 | 
   | 
| 
  66  
 | 
Sat Apr  2 16:10:22 2022 | 
Markus, Bernd | Accelerator | SIS setting complete, U90+ mit Be target: 2022-04-02_16-04-00_SIS18_FAST_E142_20220325_110213.C1.txt  | # Version: 1.2
 
# Comment: ParamModi Export at 2022-04-02T16:04:10.778077+02:00[Europe/Berlin]
 
# Context: SIS18_FAST_E142_20220325_110213.C1
 | 
   | 
| 
  65  
 | 
Sat Apr  2 15:54:32 2022 | 
Carsten | Accelerator | ESR setup with primary beam and 1cm Be plate + Schottky data on NTCAP | MArkus and Bernd have setup the ESR with the 
thick stripper foil and put the ion beam 
(238U92+) on the central orbit.
 | 
   | 
| 
  64  
 | 
Sat Apr  2 10:22:21 2022 | 
Rui-Jiu Chen, Ragan | General | Beam is in the ESR | Primary beam is in the ESR now. Markus and 
the team are tuning the ESR. | 
   | 
| 
  63  
 | 
Sat Apr  2 05:08:00 2022 | 
Kristian | General | Morning shift | I see no updates in the log. Is the morning 
shift taking place?
 
 | 
   | 
| 
  62  
 | 
Sat Apr  2 01:27:25 2022 | 
Jan Glorius | Runs | test runs 02.04.2022 | Beam settings:
 
primary beam 
 
copper stripper
 | 
    | 
| 
  Draft  
 | 
Sat Apr  2 01:02:39 2022 | 
Carsten |   | Test measurements with two DAQs | We do two measurements:
 
 
One measurement using the  | 
   | 
| 
  60  
 | 
Fri Apr  1 23:47:45 2022 | 
Carsten | Detectors | PMT thresholds / PMT north does not allow voltage abow ~ -1kV | Checked thresholds of PMT with beam and without 
beam.
 
 | 
   | 
| 
  59  
 | 
Fri Apr  1 23:12:37 2022 | 
Patrick | Detectors | Scraper positions "only primary beam" | 
 
Scraper positions where there is only primary 
beam for detector tests. Use only one outer | 
   | 
| 
  58  
 | 
Fri Apr  1 22:01:14 2022 | 
Konstantin | Detectors | PMT operating voltages | > During the testing of the PMTs the following 
operating voltages were applied:
 
> 
 | 
   | 
| 
  57  
 | 
Fri Apr  1 18:30:38 2022 | 
Carsten | Detectors | Positions / Voltages threshlds for particle detectors | [B]DC beam 238U92+ (205kV, e-beam)[/B]
 
We used the detector NA_SZ to determine the 
central primary-beam (238U92+) position by | 
   | 
| 
  56  
 | 
Fri Apr  1 18:19:58 2022 | 
Kristian | Calibration | HV divider | We checked the strange behavior of the HV 
dividers again. The Ohmlabs and the PTB divider 
are in good agreement at 120 kV and 150 kV. | 
   | 
| 
  55  
 | 
Fri Apr  1 16:46:33 2022 | 
Carsten | Detectors | All pressuzied air drives are tested now and are working properly | All pressuzied air drives are tested now and 
are working properly | 
   | 
| 
  54  
 | 
Fri Apr  1 15:52:23 2022 | 
Rodolfo | Detectors | UV PMT South | The "Solar Blind PMT" labeled as "PMT South" 
has been installed by Simon and Max in the 
gas target chamber on a CaF Viewport. The | 
    | 
| 
  53  
 | 
Fri Apr  1 15:27:00 2022 | 
Rodolfo | Detectors | Nitrogen for UV-PMTs | There was a question concerning the nitrogen 
we use to rinse the "Solar Blind PMTs".
 
 | 
     | 
| 
  52  
 | 
Fri Apr  1 11:19:57 2022 | 
Rui-Jiu Chen | DAQ | S.M.A.R.T check on NTCAP | S.M.A.R.T. check has been done on NTCAP on 
March 28th. All 12 hard disk drivers  are 
normal. | 
    | 
| 
  51  
 | 
Fri Apr  1 11:00:41 2022 | 
Rui-Jiu Chen | DAQ | Start NTCAP | Start taking data with NTCAP.
 
Carrier Frequency: 245MHz
 
IQ rate:5 MSamples/s
 | 
   | 
| 
  50  
 | 
Fri Apr  1 02:29:40 2022 | 
Carsten |   | ESR Cooler crashed | ESR cooler crashed. We could not restart the 
cooler. Since we had several other problems, 
we stop now. | 
   | 
| 
  49  
 | 
Fri Apr  1 01:57:12 2022 | 
Kristian | DAQ | Timing | with bunched primary beam (bunching frequency 
1.94070 MHz) we have strong PMT (mid) and 
particle detector signal. Time between both | 
      | 
| 
  47  
 | 
Fri Apr  1 00:27:27 2022 | 
Carsten |   | Settings Particle Detectors /Tests with primary beam | All the beam settings will be given in a separate 
post.
 
 | 
   | 
| 
  46  
 | 
Thu Mar 31 22:41:35 2022 | 
Kristian | Calibration | HV divider | We noticed a difference in the electron cooler 
high voltage:
 
 | 
    | 
| 
  45  
 | 
Thu Mar 31 22:19:14 2022 | 
Kristian | DAQ | How To Start MBS | (MBS server should be running at the LXG1297 
machine. If not you will get an error message. 
Then start the server - last command in console)
 | 
   | 
| 
  44  
 | 
Thu Mar 31 20:07:47 2022 | 
Shahab | DAQ | Schottky Display in AP Messhütte (or anywhere else) | [B]UPDATE: 16.05.2022[/B]
 
 
[I]Please use ATPPC029 instead of ATPPC019 | 
    | 
| 
  43  
 | 
Thu Mar 31 17:56:32 2022 | 
Konstantin | DAQ | Photodiode test signal | We used a signal generator to provide a 4MHz 
dummy-RF signal. This was divided by a factor 
of 2 for laser-"bunch" synchronization and | 
    | 
| 
  42  
 | 
Thu Mar 31 15:53:50 2022 | 
Konstantin | Detectors | Control software for XUV and copper mirror detector | > Attached you find a manual for the software 
used to control the movement of the XUV detector 
and the copper mirror detector. 
 | 
    | 
| 
  41  
 | 
Thu Mar 31 13:28:17 2022 | 
Jan Glorius | DAQ | DAQ-2 cabling docu | newest version V2 - valid since Mo 04.April 
2022 | 
      | 
| 
  40  
 | 
Thu Mar 31 11:49:54 2022 | 
Konstantin | Detectors | HV-Supply of particle detectors  | We cross-checked the cabeling of the particle 
detectors north and south. Particle detector 
south is connected to channel [B]HV5[/B] | 
   | 
| 
  39  
 | 
Wed Mar 30 09:27:15 2022 | 
Rodolfo | Accelerator | Issue? Scrapers at E-Cooler | Carsten* forwarded me an E-Mail from Markus 
yesterday. "Betreff: Scraper im ESR Elektronenkühler"
 
 | 
    | 
| 
  38  
 | 
Tue Mar 29 21:37:13 2022 | 
Rodolfo | Laser | Pump Laser | The gold "reflectors and end plates" of our 
pump laser (Quanta-Ray PRO) have been exchanged 
by new ones.
 | 
    | 
| 
  37  
 | 
Tue Mar 29 12:47:26 2022 | 
Ken Ueberholz |   | Parameter setting for XUV and copper mirror detector  | Some Default Settings for the XUV 
 
Detector. The copper 
 
mirror should always move fully in.
 | 
   | 
| 
  36  
 | 
Tue Mar 29 12:01:08 2022 | 
Rodolfo | Laser | Laser targets | I have installed two targets* (size: 2 inches) 
to monitor the position of the laser beam. 
 | 
      | 
| 
  35  
 | 
Fri Mar 25 14:31:24 2022 | 
Rodolfo | Laser | Photodiode signal SO in the Messhütte | Measured at the Panel "Messhütte - Laserhütte", 
BNC "Fast PD / SO".
 
 | 
    | 
| 
  34  
 | 
Thu Mar 24 17:10:33 2022 | 
Carsten | DAQ | DAQ / MBS VUPROM TDC buffer errors | VUPROM TDC buffer errors potentially due to 
long TDC common Stop gate (>1ms from div 
200).
 | 
   | 
| 
  33  
 | 
Thu Mar 24 16:59:35 2022 | 
Rodolfo | Laser | Photodiode signal SO | Fast Photodiode signal at the "SO-Bogen" (South-east 
arc).
 
 | 
     | 
| 
  32  
 | 
Thu Mar 24 13:27:47 2022 | 
Rodolfo | DAQ | Disk space - LXG1297 | In our Linux computer (LXG1297: Laser spectroscopy) 
we have a Drive with 1.8 TB total disk space. 
300 GB are at the moment in use, but I can | 
    | 
| 
  31  
 | 
Wed Mar 23 16:58:33 2022 | 
V. Hannen | Detectors | Status copper mirror and XUV detectors | Over the past days, the parabolic copper mirror 
and XUV detectors haven been prepared and 
tested for the upcoming beamtime (see first | 
      | 
| 
  30  
 | 
Wed Mar 23 16:16:42 2022 | 
V. Hannen | Detectors | Control software for XUV and copper mirror detector | Attached you find a manual for the software 
used to control the movement of the XUV detector 
and the copper mirror detector. 
 | 
    | 
| 
  29  
 | 
Wed Mar 23 15:38:15 2022 | 
Carsten | General | Not all lights in the ESR can be switched off | Three of the lights in the ESR cannot be switched 
off. It seems that these lights are now "emergency 
lights" (with battery?). We will need to | 
   | 
| 
  28  
 | 
Wed Mar 23 11:20:55 2022 | 
Simon | Detectors | PMT operating voltages | During the testing of the PMTs the following 
operating voltages were applied:
 
 | 
   | 
| 
  27  
 | 
Mon Mar 21 21:48:46 2022 | 
Jan | DAQ | MesyTec Shaper Control | The MesyTec Shaper & Constant Fraction (MSCF) 
is used to process the PMT/det signals for 
the ADC in the second DAQ.
 | 
   | 
| 
  26  
 | 
Mon Mar 21 17:03:47 2022 | 
  | DAQ | Changing code of go4 | Correction: Comments in C/C++ start with // 
, i.e. forward slash
 
 | 
   | 
| 
  25  
 | 
Mon Mar 21 16:33:23 2022 | 
  | DAQ | Changing code of go4 | We agreed to indicate changes in the go4 analysis 
code according to the
 
following scheme
 | 
   | 
| 
  24  
 | 
Mon Mar 21 10:46:18 2022 | 
Rodolfo | DAQ | TDC & Scaler description channels | For the First Vuprom we have the following 
channel configuration (Stand:21.03.2021). 
 | 
    | 
| 
  23  
 | 
Mon Mar 21 09:18:28 2022 | 
Rodolfo | Accelerator | Pressure at North-East Arc  | Our vacuum expert has reply. The "IZ Pump" 
in this region was off.
 
 | 
   | 
| 
  22  
 | 
Fri Mar 18 16:58:26 2022 | 
Rodolfo | Accelerator | Pressure at North-East Arc  | The pressure at the north-east arc of ESR, 
where the laser window is located, looks 
suspicious. The sensor E01LF02 shows a value | 
     | 
| 
  21  
 | 
Fri Mar 18 16:51:27 2022 | 
Carsten | DAQ | Main DAQ /MBS error | MBS error "TDC data ready wait" solved. Common 
stop trigger (20 kHz) was missing. (according 
to Sebastian) | 
   | 
| 
  20  
 | 
Fri Mar 18 09:18:38 2022 | 
Rodolfo | Accelerator | Vacuum leak, ODR | After the beaking out process we got a vacuum 
leak in one of the MgF2 windows (Referenz: 
PMT South) of the optical detection region | 
    | 
| 
  19  
 | 
Thu Mar 17 17:56:33 2022 | 
Kristian | Calibration | HV dividers | PTB HV dividers arrived. Divider ratios are:
 
 
PT120:          12003,341 
 | 
    | 
| 
  18  
 | 
Thu Mar 17 17:37:49 2022 | 
Rodolfo | Accelerator | High voltage dividers for E-Cooler | Two dividers, from PTB-Braunschweig, have 
been deliver today. We will use them to monitor 
the electron-cooler voltage in situ.
 | 
    | 
| 
  17  
 | 
Wed Mar 16 12:04:40 2022 | 
Kristian | Detectors | HV cable to particle detector South | Checked HV cable to particle detector south. 
Label is HV2 Scraper. 
 
 | 
   | 
| 
  16  
 | 
Wed Mar 16 08:31:52 2022 | 
Kristian | DAQ | Check cables and timing | Konstantin and I generated a signal in the 
Messhütte and sent it to the ESR. 
 
There, we used a 44m BNC cable that was sufficently | 
   | 
| 
  15  
 | 
Thu Mar 10 15:20:26 2022 | 
Jan  | DAQ | analog branch signal patches to Messhütte  | The additional signals from PMTs and XUV detector 
needed for spectroscopic analysis in the 
second DAQ are patched to AP Messhütte (MH) | 
        | 
| 
  14  
 | 
Thu Mar  3 21:39:21 2022 | 
Rodolfo | Laser | Dye solution, DCM in Ethanol and DMSO | I have prepared a new dye solution of DCM 
in Ethanol and DMSO (1:1). The concentration 
of the "oscillator + preamplifier" is 0,3 | 
    | 
| 
  13  
 | 
Thu Mar  3 17:34:42 2022 | 
Rodolfo | Laser | Pump Laser  | We have issues with our Pump Laser (Spectra 
Physics - Quanta Ray Pro). After the service 
made on mid Dezember (2021) the laser does | 
    | 
| 
  12  
 | 
Thu Mar  3 17:21:26 2022 | 
Rodolfo | Detectors | MgF2 Windows | New reentrance viewports made of MgF2 have 
been installed at the ESR "Optical Detection 
Region". In the middle the old window (also | 
    | 
| 
  11  
 | 
Thu Mar  3 13:50:45 2022 | 
Rodolfo | Laser | Dye cell cracket | The "20 mm" dye cell used for the Oscillator 
and Pramplifier got damage. I supposed it 
was produced by the pump beam hitting on | 
    | 
| 
  10  
 | 
Fri Nov 19 15:50:44 2021 | 
Rodolfo | Accelerator | Laser windows at E-Cooler side | New laser viewports have been installed at 
the E-Cooler side of the ESR.
 
 | 
     | 
| 
  9  
 | 
Fri Nov  5 17:23:23 2021 | 
Rodolfo | DAQ | Dye trigger out  | The dye laser offers different scan settings 
(see pages 65 - 67 of the user manual). We 
are using the following setting, see attached | 
    | 
| 
  8  
 | 
Thu Oct 14 10:21:46 2021 | 
Rodolfo | Detectors | Dismounting Cu-Mirror | We have dismounted the parabolic Cu-Mirror 
at ESR on 13.10.2021  | 
    | 
| 
  Draft  
 | 
Wed Sep  8 21:44:55 2021 | 
Rodolfo | General | Preparation | [TABLE border="1"]
 
[B]Status[/B]|[B]Arbeitspaket[/B]|[COLOR=chocolate][B]Kommentare[/B][/COLOR]
 
|-
 | 
   | 
| 
  6  
 | 
Fri Aug  6 10:40:54 2021 | 
Rodolfo | DAQ | mbs in atplaser | The "MBS program" (data acquisition) is now 
located within atplaser@r4l-41, in the folder 
"th21":
 | 
   | 
| 
  5  
 | 
Fri Aug  6 10:23:00 2021 | 
Rodolfo | DAQ | atplaser on r4l-41 | The user [B]atplaser[/B] is now existing on 
the [B]R4L-XY[/B] nodes for the ESR group.
 
 | 
   | 
| 
  4  
 | 
Mon Aug  2 10:54:44 2021 | 
Rodolfo | Laser | Quanta Ray - Water leak in the power supply | After fixing the broken water connector in 
the laser head a new water leak appeared 
in the Quanta-Ray power supply. 
 | 
    | 
| 
  3  
 | 
Wed Jul 28 13:31:22 2021 | 
Rodolfo | General | Air filters | Filters (Hepa-) inside the flow boxes have 
been replaced. | 
    | 
| 
  2  
 | 
Mon Jul 26 17:58:52 2021 | 
Rodolfo | Laser | Water Filters | The water filters (deioninzing and particle) 
have been replaced.  | 
    | 
| 
  1  
 | 
Mon Jul 26 17:15:47 2021 | 
Rodolfo | Laser | Quanta Ray - Water leak | On Thursday afternoon (22.07.2021) a water 
connector has broken and water has been spilled 
in the laser head.  | 
    |