Sun Mar 25 08:24:18 125, Ken Ueberholz, Detectors, XUV-Detector Rates
|
Count rates were measured for the XUV-
detector for a full circle:
Cathode: -5V, shields: 0V, main coil 8A,
|
Wed Apr 20 10:09:50 2022, Rodolfo, Laser, Optic with spots
|
The average energy produced by the laser along the beamtime was about 10 mJ/puls. After about two weeks of beamtime some of the optics, especially the optic
placed just after the BBO crystall, got some spots, see pictures below. Here the diameter of the laser beam is about 6 mm. |
Sat Apr 16 04:19:05 2022, Patrick, Simon, Alex, Sarper, Runs, Runs night + morning shifts 16.04.
|
Runs of the night shift 16.04. after changing the wavelength range. The new range is 320.12 - 319.859 nm. The step size was decreased from 0.002 nm to 0.001
nm.
The goal was to have the same number of scans as before (261), however, the scan range in the labview program could not be set accordingly. We decided |
Fri Apr 15 20:43:18 2022, Konstantin, Laser, Dye changed
|
Dye of the laser system has been changed. UV laser power increased from ~8 mJ to 11 mJ pulse energy. |
Fri Apr 15 12:53:25 2022, Max, Danyal, Laser, checked laser pulse energy after change in wavelength range
|
previous scan range:
the laser pulse energy @ 320.62 nm was ~11 mJ
|
Fri Apr 15 12:42:18 2022, Max, Danyal, Carsten, Konstantin, Patrick, Simon, General, laser wavelength range changed: 320.12 - 319.60 nm
|
Runs of the morning/afternoon/night shift. The wavelength range is now 320.12 - 319.60 nm.
|
Thu Apr 14 22:44:40 2022, Patrick, Simon, Danyal, Max, Runs, Runs night shift 14./15.04.
|
Runs of the night shift. The wavelength range is 320.62 - 320.10 nm.
[TABLE border="1"]
|
Thu Apr 14 19:46:46 2022, Sarper & Konstantin & Rodolfo, Laser, Laser power
|
We checked the laser power before starting the new scan range.
- Seed energy before Cobra @ 532 nm: 513 mJ (before 499 mJ).
|
Thu Apr 14 17:47:55 2022, Carsten, Rodolfo, Konstantin, Sarper, Runs, Afternoon shift runs 14.04 changing the wavelength scan range
|
The new scan range is changed from [B]321.12 - 320.60 nm to 320.62 - 320.10 nm[/B]
[TABLE border="1"]
|
Thu Apr 14 12:38:38 2022, Sebastian, Detectors, XUV / Cu checkups
|
Check regularly if the connection to the XUV and Cu detectors are still active and the detectors move in (after injection) and out (after killing the ion
beam).
|
Thu Apr 14 12:27:18 2022, Sebastian, Max, Detectors, Connection to XUV / Cu reestablished!
|
The connection to the XUV and Cu were lost during the nightshift.
The XUV detector was in the park position,
|
Thu Apr 14 10:57:11 2022, Carsten, Accelerator, Maintanence UNILAC / Op training
|
We had no beam since about 1000. Also, we are still trying to find out why we have such a low intensity in the expereimnt.
Presently, there is a planned maintenance in the UNILAC, beam is expected back in 25-30 mins.
The profile grid TE3 in the transfer beamline in front of the target will be used to check if the extracted SIS beam has changed compared to before the |
Thu Apr 14 08:41:42 2022, Carsten, Accelerator, Low intensity at the experiment after series of failures during the night
|
After the recovery from the failures during the night, we see much lower intensity at the experiment (measured with the particle detector count rate).
Instead of 8000-10000 cts we see only 3000-4000cts at comparable SIS intensities.
Also ESR cooler seems to be slightly unstable (wiggling lines) during injection. However, it is visually evident from the Schottky that the injected intensity |
Thu Apr 14 06:29:18 2022, Max, Sebastian, Accelerator, lesser ESR injection intensity
|
We now have less ESR injection intensity (see screenshot attached).
This can also be seen in the count rate after "UNILAC/SIS failures".
|
Wed Apr 13 15:41:29 2022, Rodolfo & Sarper, Laser, Laser status
|
We have checked the power of the laser and optimized it "a bit".
- Seed energy for the Cobra @ 532 nm: 508 mJ (before 496 mJ)*
|
Wed Apr 13 14:29:04 2022, Carsten, Rodolfo, Sarper, David, Pierre-Michel, Max, Sebastian, Runs, Afternoon - morning shift runs 13.+14.04.
|
Changed the wavelength scan range from [B]321.62 - 321.10 nm[/B] to [B][B]321.12 - 320.60[/B] nm[/B] after one scan.
[TABLE border="1"]
|
Tue Apr 12 21:56:54 2022, Volker, David, Pierre-Michel & Danyal, Alexandre, Sebastian, General, new laser wavelength scan range 321.62 - 321.10 nm
|
New wavelength scan range is [B]321.62 - 321.10 nm[/B]. Before the new scan range (compare old scan range in [URL=https://elog.gsi.de/esr/E142/215] Entry
215[/URL]) was applied,
the cooler settings were slightly modified. Current was changed from [B]50 mA to 40 mA[/B], correction voltage was changed from [B]3943 V to 3948 V[/B]. |
Tue Apr 12 17:01:30 2022, Konstantin, Detectors, Voltage PMT North
|
Voltage of PMT North has been increased from 2500 V to 2600 V. Additionally the threshold was adapted from 25 mV to 12 mV. |
Tue Apr 12 15:05:56 2022, Rodolfo, Laser, Laser status
|
We have checked the laser power at 14:15 *
- Pump energy after Q-Ray @ 532nm: 559 mJ.
|
Tue Apr 12 14:52:19 2022, Rodolfo, Detectors, PMT Nord, Middle. Datasheets
|
Attached are the datasheets for the solar-blind PMTs.
- PMT Nord. Type 9423. SN: 650
|
Tue Apr 12 13:24:40 2022, Alexandre, Sebastian, Detectors, Sideband in Schottky
|
Sometimes we see a broad sideband at lesser schottky frequencies to the high intensity thorium beam.
Pic 1 shows the relative broad sideband on the left.
|
Tue Apr 12 14:11:52 2022, Ragan, Detectors, Sideband in Schottky
|
> Sometimes we see a broad sideband at lesser Schottky frequencies to the high-intensity thorium beam.
>
> Pic 1 shows the relative broad sideband on the left.
|
Tue Apr 12 12:41:20 2022, Alexandre, Sebastian, Accelerator, no beam in SIS18
|
|
Tue Apr 12 04:43:29 2022, Alexandre, Kristian, Sebastian, Runs, 322.01 - 321.60 nm
|
We scan the wavelength scan range is 322.01 - 321.60 nm
before the first scan of the new wavelength range we changed the high voltage divider ratio in Labview from 248514 to 248517, which is the correct value |
Mon Apr 11 22:33:16 2022, Carsten, General, Backup of data to the external RAID disks
|
|
Mon Apr 11 22:14:51 2022, Konstantin, Sarper, Laser, Changing of the wavelength range
|
It is planned to change the scan range of the laser wavelength from 322.41 - 322.0 nm to 322.01 - 321.6 nm @ 4 a.m. |
Mon Apr 11 18:57:18 2022, Rodolfo, Laser, New flashlamps
|
We (Sarper, Konstantin and me) have changed the old flashlamps (with ~ 26 Mio. shots) by new "fresh" ones. We have also exchange the dye. Here are some
parameters of the laser we measured with the new flashlamps and dye*:
|
Mon Apr 11 18:46:31 2022, Rodolfo, Sarper and Konstantin, Laser, Alignment-check
|
We cross-checked the laser-ion alignment. The horizontal position was optimized. However,
since the laser beam is much larger than the ion beam, this was not an issue for measurements, which
have been performed during the last shifts.
|
Mon Apr 11 14:45:11 2022, Paul, Konstantin, Sarper, David, Runs, Afternoon shift runs 11.04
|
We continue scan the new wavelength scan range is 322.41 - 322.00 nm
|
Mon Apr 11 12:50:22 2022, Kristian, Accelerator, Th lifetime
|
We used the counter to measure the lifetime. During the 30 min scan time dropped only slowly.
During a longer break, e.g., when the laser flash lamps are changed, one could take data over a longer time. |
Mon Apr 11 11:05:35 2022, Sebastian, Kristian, Laser, Check laser power
|
we checked the UV laser power at 10:15 am. It was only 8.5 - 9 mJ. Pumping power was 460 mJ, which is the maximum output of the current system. We should
change the flashlamps soon |
Mon Apr 11 10:31:07 2022, Kristian, Sebastian, Alexandre, Danyal, Paul, General, new laser wavelength scan range 322.41 - 322.00 nm
|
new wavelength scan range is 322.41 - 322.00 nm
at 10:31
|
Mon Apr 11 10:17:01 2022, Danyal, Accelerator, Interlock - reset
|
10:17
Wie heutemorgen in der Sitzung (morning briefing) vereinbart,
führt Herr Florenkowski einen Interlock und danach einen Reset von E02MK durch.
|
Mon Apr 11 07:54:37 2022, Danyal, Alexandre, Kristian, Sebastian, Accelerator, Jump in E-cooler voltage
|
We saw a jump in the E-cooler voltage at about 7:50.
jump from 209619 to 209610 V
|
Mon Apr 11 04:26:53 2022, Patrick, DAQ, Next wavelength range
|
Measurements with the next wavelength scan range should be started roughly around 10:00 am.
The next wavelength scan range is planned to be 322.41 - 322.00 nm, again with a step size of 0.002 nm, 5s dwell time and 2 loops.
This range includes roughly a third of the first scan range. |
Sun Apr 10 22:47:05 2022, Patrick, Volker, Prannay, Danyal, Alexandre, Sebastian, Kristian, Runs, Runs night shift 10.04 + morning shift 11.04
|
These are the measurement runs of the night shift (10/11.04.).
Wavelength range is still 322.81 - 322.40 nm.
|
Sun Apr 10 15:29:53 2022, Rodolfo, Laser, Status - Laser system
|
We have checked the laser system.
- The pump laser has ~ 23Mio shots. The total output energy has dropped to about 475 mJ. Maybe it is time to install new fresh flashlamps. In case the |
Sun Apr 10 15:29:37 2022, Carsten,Sarper, Accelerator, Ecooler failed
|
Ecooler failed @15:23 Carsten is fixing it
Fixed @16:00 |
Sun Apr 10 15:16:19 2022, Carsten, Rodolfo, Sarper, Runs, Runs afternoon shifts 10.04
|
These are the measurement runs of the afternoon shift (10.04.).
Wavelength range is changed from 323.2 - 322.8 nm to 322.81 - 322.40 nm.
|
Sun Apr 10 15:08:42 2022, Carsten, DAQ, Backup of Data from 1rst Scanrange to extrenal RAID1 HDDs
|
I just performed a backup of all data (+log-Files) that were collected until now to the two RAID HDDs.
For DAQ1: /media/sdc2 - last file 229Thor7889.lmd. Please note, that the naming sceme on DAQ1 is inconsistent - please look for the filenumnbering
|
Sun Apr 10 13:17:53 2022, Kristian, Max, DAQ, Wavemeter readout
|
We changed the Labview code in order to reduce the amount of data that was overfilling the WM data queue. We removed the temperature and intensity values
and added an additional 250ms wait.
|
Sun Apr 10 11:11:18 2022, Kristian, Max, Laser, additional WM readout
|
since the WM readout still does not work reliably, I started a python script at atppc023 that saves timestamp and wavemeter frequency. in folder wavemeter
at the desktop.
|
Sun Apr 10 08:53:26 2022, Kristian, Max, Laser, Timing verified
|
timing is still good.
we checked the relative timing between laser and ions. the time difference between photodiodes, PMTs and particle detectors is still the same as original |
Sun Apr 10 06:39:16 2022, Kristian, Marek, Max, Runs, runs morning shift 10.04.
|
These are the measurement runs of the morning shift (10.04.).
|
Sun Apr 10 06:34:49 2022, Kristian, Max, DAQ, Wavemeter Readout
|
Problem solved. Ignore this entry
|
Sun Apr 10 04:37:29 2022, Patrick, DAQ, Next wavelength range
|
Measurements with the next wavelength scan range should be started roughly around 12:00 noon. The plan is to go to lower wavelengths / higher energies.
So the next wavelength scan range would be 322.81 - 322.4 nm, again with a step size of 0.002 nm and 5s dwell time.
|
Sun Apr 10 04:07:53 2022, Patrick, DAQ, go4 shows wrong frequencies and wavelengths
|
[B]Update (ID:194):[/B] In labview "Stop All", close labview, stop python script (ID:197), close wavemeter. Then restart wavemeter, python script, labview
an mbs.
|
Sun Apr 10 03:56:56 2022, V. Hannen, Detectors, Low rate in PMT South
|
PMT south shows a much lower count rate than observed in elog entry 170.
Yesterday the voltage on this PMT was increased to 2400 V and the threshold setting of the CAEN CFD to 90 units, (from 70 units before) however the low |
Sun Apr 10 02:20:17 2022, Patrick, DAQ, New mbs startup guide
|
Connecting to the rfio server and opening a file is included in '@startup' now. Since mbs crashes reliably every 3 or 4 runs, it may be preferable to use
the following procedure (which did not crash yet):
|
Sat Apr 9 19:50:40 2022, Carsten, , New thresholds for ADC of DAQ2 (starting with file0117)
|
unsigned short caen_V785AH_threshold_mod0 [32] =
{0x0100 , 0x0004 , 0x0003 , 0x0004 , 0x0005 , 0x0004 , 0x0006 , 0x0005 ,
0x0006 , 0x0005 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 ,
|
Sat Apr 9 19:09:57 2022, Rodolfo, Laser, Optimization - Resonator of dye laser
|
I have tuned a bit the resonator of the dye laser.
The total output energy after the main amplifier is 86 mJ (@ 645,6 nm)
|
Sat Apr 9 16:55:19 2022, Rodolfo, Laser,
|
16:15 I have changed the dye (of both oscillator and amplifier).
16:27 I have increased the pump energy (@ 532 nm) from 476 mJ to 507 mJ.
|
Sat Apr 9 15:47:21 2022, Carsten, Rodolfo, Sarper, Patrick, Runs, Afternoon and night shift runs
|
After the morning shift, we got new increment (nm) for the laser scanning. We are now scanning 2 nm of wavelength range in 5 big steps (5 x 0.4 nm). Started
with 323.2 nm - 322.8 nm range. Every big step (0.4 nm) is scanned with 0.002 nm of wavelength increment.
|
Sat Apr 9 14:15:44 2022, Kristian, Max, Konstantin, DAQ, Change in LV code
|
We changed the LV code so that the Cobra Set Wavelength is permanently sent. Before it was only sent once at the beginning of each laser step. This, however,
lead to too much data for the LV->MBS server.
This was fixed by inserting a 250 ms wait in LV (in Krypton data VI). Basically, all LV values are now updated every 250 ms, which works fine with MBS |
Sat Apr 9 13:45:42 2022, Max, Kristian, Runs, first scans with LV problem solved!
|
This is the first time with LV buffer problem solved (see Elog ID: 184). Before the PMT signals were out of sync with the wavelength data!
We concluded to scan with step sizes of 0.002nm, scan range of 0.5nm (f
nm), dwelltime = 150(5s), loops = 2 (approx.: 30min measurements).
|
Sat Apr 9 06:42:23 2022, Kristian, Laser, Laser step size
|
We changed the laser step size to 0.002 nm which corresponds to 5.75 GHz. |
Sat Apr 9 06:25:22 2022, Max, Kristian, Runs, morning shift runs
|
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
Sat Apr 9 05:26:25 2022, Patrick, Laser, Wavemeter
|
The wavemeter still shows large spikes from time to time. The exposure time is still at 34ms. |
Sat Apr 9 05:00:11 2022, Carsten, General, Draft Docs and cheat sheet
|
Please check the attached files,
and correct or enhance them.
|
Sat Apr 9 04:40:01 2022, Patrick, DAQ, Bugged Trending plots?
|
The problem from yesterday (ID:147) still persists. All the plots (not only the labview stuff) comes in too late in go4. However, the time axis is correct,
see picture. |
Sat Apr 9 04:01:35 2022, Carsten, , 2nd DAQ
|
Switched the CAEN TDC back on and changed the threshoolds for the ADC to:
unsigned short caen_V785AH_threshold_mod0 [32] =
|
Sat Apr 9 03:57:10 2022, Carsten, Accelerator, New Cooler voltage after the crash of the cooler
|
After the restart of the cooler the Ohmlabs divider showed a voltage 16V lower than before. This could also be seen as a slight shift in the Schottky lines
w.r.t to the bunching frequency and the marker. Corrected
the voltages at injection to DeltaU +3963V and in Sc 5/7 to + 3943V. Now perfect match with the old Schottky frequency.
|
Sat Apr 9 03:46:31 2022, Carsten, Accelerator, Slits in TE line
|
Changed GTEDSAH Rechts to 10mm |
Sat Apr 9 03:27:47 2022, Patrick, DAQ, Laser scanning
|
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
Sat Apr 9 02:55:38 2022, Carsten, Accelerator, Kühlerausfall und Gründe dazu
|
Wir hatten heute Abend eine Kühlerausfall, der uns mehrere Stunden beschäftigt hat, weil gem. Anzeige in Device Control kein Netzteil mit „rot“ angezeigt/ausgefallen
war (siehe unten).
Nach Rufbereitschaft Serge, dann Regina sowie Controls konnte zunächst gefunden werden, dass das Netzteil einer Clearingelektrode (GECEKD2D) keinen korrekten |
Fri Apr 8 23:56:07 2022, Patrick, DAQ, DAQ improvement
|
As requested in ID:146 and ID:147, the laserscan tab was improved:
[LIST]
[*] The laserscan tab shows each observable plotted against the predefined wavelength setpoints now.
|
Fri Apr 8 22:26:44 2022, Patrick, DAQ, Spike in all detectors
|
In the previous shift, there was a spike in all detectors in the same scan step. Apparently, this was NOT a resonance since in the next step it appeared
again but at a different wavelength.
The current hypothesis is, that this is a disruption signal from one of the machines. |
Fri Apr 8 17:57:45 2022, Konstantin, DAQ, Laser scanning
|
[TABLE border="1"]
|
Fri Apr 8 17:44:25 2022, Konstantin, Calibration, Final Timing
|
We measured the timing for the laser bunch synchronization. A summary is given below.
[TABLE border="1"]
|
Fri Apr 8 15:45:24 2022, YL, , Spectrum manipulation 5
|
See attached |
Fri Apr 8 15:42:20 2022, Yuri, Accelerator, Slits in the TE line
|
At around 15:00, The slits in the TE line were changed to :
TE4DS5HL -12 mm -> -20 mm
TE5DSAH 0 mm -> +15 mm |
Fri Apr 8 15:39:36 2022, all, Runs, first data after timing set up
|
started taking data again
started at 15:40
|
Fri Apr 8 14:46:02 2022, Danyal & Sebastian & Volker & Ken, Detectors, high dark countrate Cu PMT solved
|
The problem - "discovered" by Volker and Ken - of the high dark countrate of the Cu PMT has been solved.
(when the Cu mirror was moved in, the dark countrate was very high ... also without ion beam)
|
Fri Apr 8 14:43:50 2022, Wilfried, DAQ, laser timing - Q-switch start
|
below are my notes concerning laser timing / Q-switch start |
Fri Apr 8 14:14:28 2022, Kristian, DAQ, Delay of bunch RF
|
The bunch HF is now fed through a delay to shift the stop trigger relative to the PMT signal. The delay generator is placed directly after the CFG and before
splitting the signal for the two DAQs. The relative timing of both DAQs was not changed. |
Fri Apr 8 14:12:57 2022, morning shift, DAQ, timing for h=1
|
The timing for h=1 has been set up
The time window is now twice as large as before!
|
Fri Apr 8 12:04:13 2022, Ragan, Runs, NTCAP started
|
NTCAP is started for the data taking.
Carrier Frequency: 242.02 M
|
Fri Apr 8 11:16:57 2022, Kristian, Laser, Wavemeter
|
We changed the fiber to the wavemeter. The third fiber worked best and we could operate the WM at low exposure times, i.e., below the laser repetition rate.
We used the neutral density filter to adapt the laser power that only one laser pulse is measured by the WM each time (Exposure time =34ms)
|
Fri Apr 8 11:02:04 2022, DW, YL, RSS, SL, , Change h=2 to h=1
|
After changing to h=1
The frequency manipulation #5: f=242.021487MHz (124th harmonic) [no bunching]
Manipulation #7: f=242.021812 MHz [with bunching] |
Fri Apr 8 10:44:55 2022, Ragan, Yuri, Danyal, Sergey, Runs, NTCAP stopped
|
NTCAP is stopped. Sergey will now set an ion bunch instead of two in the ESR.
SC_2022-04-08_09-39-33
|
Fri Apr 8 09:39:04 2022, Ragan, Runs, NTCAP started
|
Carrier Frequency: 242.2 M
IQ Rate (S/s): 10 M
Reference level (dBm): -30
|
Fri Apr 8 09:33:52 2022, Dayshift 1 8.4.2022, , Reference Measurements
|
DAQ1: lxg1297:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
Fri Apr 8 09:32:32 2022, Ragan, Runs, NTCAP stopped
|
Since there is no beam, NTCAP is stopped.
Carrier Frequency: 242.2 M
|
Fri Apr 8 09:13:38 2022, Ragan, Danyal, Accelerator, No beam
|
Cathode is being changed at the ion source. So no beam for the moment.
Sergey has been contacted to change from two ion bunches to a single ion bunch in the ESR. |
Fri Apr 8 08:40:06 2022, Sebastian, DAQ, DAQ improvement
|
Concerning ID 146 a new histogram for accumulating data (no cleaning histograms) can be found under
Step2_Anl/TDC/TDC***_accumulated_*** |
Fri Apr 8 08:17:52 2022, Sebastian, Laser, Laser Ext Trig is working
|
We (Max, Kristian, Danyal, me) fixed the "external trigger" problem.
1.) The delay between lamp sync and Q-switch trigger was too long. The delay must be 200µs (3rd picture)!
|
Fri Apr 8 06:53:29 2022, Ragan, Runs, NTCAP data taking
|
NTCAP is taking data since 07.04.2022 23:53:47
Carrier Frequency: 242.2 M
|
Fri Apr 8 05:46:35 2022, V. Hannen, Detectors, High background rate in Cu PMT points to light leak in ESR or active pressure gauge
|
The background rate of the Cu mirror PMT is only about 300 cps when the mirror is in the parking position but ca. 7000 cps when in the beam position, even
if no beam is present in the ESR. This was noticed when we had an "empty injection" into the ring, causing the Cu mirror to drive into the beam position
with no ions present.
|
Fri Apr 8 03:54:39 2022, Patrick, Laser, Q-switch window
|
The solution of increasing the Q-switch window from 2us to 5us described in [I]ID:148[/I] fulfills its purpose, but the window is unstable.
We saw the laser jumping across the screen which was because the Laser had a lot of hickups.
The hickups come from the signal which determines the duration of the window.
|
Fri Apr 8 02:14:05 2022, Wilfried, Laser, Laser timings adjusted
|
Since we use the middle window as a reference for the laser pulse timing, the fact that the window is 2.25 m behind the target has to be taken into account.
From here the ions have 10 ns less time of flight to the middle of the electron cooler than they need for half a revolution. The signals were still symmetrically
aligned around the reference bunch as it was done yesterday. Therefore, the delay of the laser pulse was reduced by 10-ns on the oscilloscope, which led |
Fri Apr 8 01:48:57 2022, Patrick, DAQ, Bugged Trending plots?
|
The wavelengths/frequencies shown in the graph in the labview window do not match the wavelengths/frequencies in go4.
Compare the picture which shows an incomplete laser scan although the scans in the labview window are completed (The measurement was started at an 'elapsed
MBS time' of 2700s and took 1700s/28min).
|
Fri Apr 8 01:18:51 2022, Ken Ueberholz, DAQ, TODO: DAQ improvements
|
For somebody who knows the DAQ1 code: Go4 resets the histograms with each laser step. Can there be additional histograms where the data gets accumulated?
And possibly a histogram where important detectors are plotted against the laser steps? The read back wavelength of the wavemeter is not very precise. |
Fri Apr 8 00:06:37 2022, Patrick, Laser, Connection to the laser pc
|
To login to the laser pc use the command
xfreerdp -g 1920x1080 -u atplaser atppc023.campus.gsi.de |
Thu Apr 7 22:18:06 2022, Night shift 7.4-8.4, , Next scans
|
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
Thu Apr 7 19:13:26 2022, Konstantin, Patrick, Calibration, How to find the correct laser position
|
In order to achieve the correct laser ion beam superposition one needs to do the following steps:
[LIST=1]
|
Thu Apr 7 16:50:22 2022, Konstantin, Calibration, Spacial laser-ion overlap
|
Laser-ion-beam overlap was achieved after measuring the ion beam positions (compare [URL=https://elog.gsi.de/esr/E142/83]Entry 83[/URL]).
Screenshots are comprised in the following table.
|
Thu Apr 7 16:45:10 2022, , Detectors, Detector distances
|
Relativ distances of the detector |
Thu Apr 7 15:18:25 2022, Danyal & Sebastian, DAQ, DAQ channels
|
We checked again the cables for both DAQs
(on Carsten's kind request)
|
Thu Apr 7 14:13:27 2022, Danyal, Detectors, new PMT south
|
I have replaced the PMT south (looks at the gas target).
in the hope that this will produce less dark counts.
Thus far, the dark countrate is low (few Hz).
|
Thu Apr 7 14:04:36 2022, Paul Indelicato, General, no beam until 17:00
|
10:10 There is a problem with the UNILAC.
The repair will take several hours. (4 hrs?)
|
Thu Apr 7 14:01:22 2022, Sebastian, Max, Detectors, Mesytec reset
|
The Mesytec crate was turned off and on again.
It is again possible to communicate with the XUV detector via python to apply voltage and current. |
Thu Apr 7 11:11:31 2022, Max, Sebastian, Laser, dye changed, pulse energy measurements
|
We changed the dye at 10am.
We increased the pump power for the dye Laser to 480mJ (before 450mJ). New calibration factor is 29 for Quanta Ray monitor signal.
|
Thu Apr 7 10:10:31 2022, Ragan, Yuri, Danyal, Accelerator, Problem in UNILAC
|
The power supply of the UNILAC magnet is broken. The repair will take up to 6-7 hours. |
Thu Apr 7 09:58:55 2022, Ragan, Yuri, Danyal, Accelerator, Kicker problem solved
|
The kicker problem is solved and the beam is back in the ESR. |
Thu Apr 7 09:23:18 2022, Sebastian, DAQ, DAQ1 & DAQ2 - 1kHz and delayed cs removed
|
Questionable 1kHz and delayed common stop signals has been removed.
These two cablings were for some previous tests. |
Thu Apr 7 08:43:19 2022, Danyal, Yuri, Ragan, Accelerator, again no injection in ESR - ESR kicker problem
|
we had no injections in the ESR again
it turned out that the injection kicker of the ESR has a problem |
Thu Apr 7 08:14:03 2022, Yuri & Danyal, General, 229Th89+ line in Schottky
|
GLOBAL calculations for the "content" of the Th line in the Schottky spectrum
(380 MeV/u exit energy, equilibrium charge state distribution)
|
Thu Apr 7 07:40:09 2022, Danyal, Max, Sebastian, DAQ, No beam and DAQs crashed
|
since shortly after 7 AM we have no beam in the ESR
(reason unclear)
|
Thu Apr 7 05:55:17 2022, Ken Ueberholz, Detectors, XUV-Detector Mesytec crash
|
When the XUV-detector has to shut down unexpectedly, it can happen that the Mesytek power supply crashes and can't be
communicated with. If this happens the Mesytec PS needs to be restarted from hand inside of the ESR. This can be done by
restarting the crate located at the bottom of the XUV-detector.
|
Thu Apr 7 05:44:40 2022, Pierre-Michel, Ragan, DAQ, NTCAP settings
|
Carrier Frequency: 244M
IQ Rate: 10M
reference Level: -30
|
Thu Apr 7 05:23:50 2022, Patrick, DAQ, How To Start MBS for DAQ2 (Carstens DAQ)
|
(MBS server should be running at the LXG0155 machine. If not you will get an error message. Then start the server - last command in console)
Just enter the following commands in the console of any PC
|
Thu Apr 7 04:48:58 2022, , , ToDo's for 07.04.
|
Some ToDo's for the next shift:
- Work on Laser stabilisation 2nd stage in the ESR
- Mesytec power supply of the xuv-detector crashed and needs to be restarted in the ESR. Simply turning it the crate off and on again)!!
|
Thu Apr 7 04:10:18 2022, Ken Ueberholz, DAQ, Corrected cabling DAQ 2
|
A lot of cables were wrongly or not
connected to the DAQ 2.
|
Thu Apr 7 03:44:21 2022, Patrick, DAQ, Measurement procedure
|
[U][B]Measurement procedure[/B][/U]
|
Thu Apr 7 03:09:36 2022, Carsten, DAQ, NTCAP running since 02:40
|
|
Thu Apr 7 03:06:15 2022, Patrick, DAQ, DAQ file sizes
|
The file sizes of DAQ1 (229Thor) and DAQ2 (229Th, "Carstens DAQ") are different.
During 20min of laser scanning DAQ1 saves 4 files while DAQ2 saves 2 files which was 3.3 GB and 2 GB, respecively.
[B]The file sizes of DAQ2 should always be greater than those of DAQ1, because of extra ADC data![/B]
|
Thu Apr 7 02:32:29 2022, all on shift, Runs, First Scans 322-320nm
|
Laser scanning from 322-320 steps 0.01nm/s
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
|
Thu Apr 7 01:19:05 2022, Patrick, Laser, Laser scanning
|
We did some laser scans. The wavemeter shows spikes during scans. This may be just while the frequency changes. See picture. |
Thu Apr 7 00:36:36 2022, Uwe, DAQ, Threshholds and shaping Mesytec for ADC
|
|
Thu Apr 7 00:15:11 2022, Patrick, Laser, Laser timing adjusted
|
We have adjusted the laser timing using PMT-middle. Bunching freq. is 3.90362 MHz.
PD NO: 7.79
|
Wed Apr 6 22:29:55 2022, Patrick, Accelerator, ECooler voltages
|
ECooler voltages:
[B]06.04:[/B]
|
Wed Apr 6 21:03:25 2022, Ken Ueberholz , Detectors, Xuv-Detektor, further information
|
Some further information regarding the
xuv-detector and its components. |
Wed Apr 6 19:39:08 2022, Ken Ueberholz , Detectors, LED installed
|
The UV-LED of Münster was connected over
channel 8 to the Messhütte. 6V are
supplied by the power supply (picture 2).
|
Wed Apr 6 18:51:33 2022, Danyal & Carsten, Accelerator, target bump
|
This afternoon, we found out that Sergey has also implemented a "target bump".
Although we do not use a target, this bump is still convenient because
we can move the ion beam such that is passes perfectly through the Cu mirror.
|
Wed Apr 6 16:01:07 2022, BL, Accelerator, ESR settings before adjusting laser: 2022-04-06_14-37-30_ESR_EXP22_02_238U92_Brandau_30March22.C1.txt
|
# Version: 1.2
# Comment: ParamModi Export at 2022-04-06T14:37:35.957978+02:00[Europe/Berlin]
# Context: ESR_EXP22_02_238U92_Brandau_30March22.C1
|
Wed Apr 6 15:19:29 2022, Svea, Shahab, Accelerator, 5s delay implemented before injection
|
Operators have informed us that for better stability and operation (and cooperation with HHT), and after discussing with Sergey, they have introduced 5s
delay "Wartezeit RESET" before every injection.
|
Wed Apr 6 14:52:29 2022, Rodolfo, Runs, Scraping the ion beam
|
We are T-scraping the ion beam. |
Wed Apr 6 13:32:57 2022, Max, Sebastian, DAQ, LabView Cobra Channel List
|
channel 0: linux time (epoch time)
channel 1: ms timer
channel 2: E-Cooler Voltage(Ohmlabs)
|
Wed Apr 6 09:30:53 2022, Danyal, Accelerator, ESR is blue again - errors are fixed
|
The ESR is "blue" again.
The errors (scu related) have been fixed.
|
Wed Apr 6 08:31:54 2022, Danyal & Regina, General, ESR had some errors
|
Regina noticed that the ESR was "red"
This means that there are some errors.
|
Wed Apr 6 08:06:02 2022, Danyal, DAQ, stopped NTCAP data acquisition
|
I stopped the NTCAP data acquisition at 8 AM.
Let's hope that the file contains good data. |
Tue Apr 5 23:55:11 2022, Carsten, Runs, Long time measurment with NTCAP
|
Settings:
Center: 2242.5MHz
RefLevel: -25dB
|
Tue Apr 5 23:07:34 2022, Carsten, Accelerator, 229Th89+ identification after re-tuning of the ESR
|
The ESR was retuned this aftenoon since the it was unclear if we still sit the 229Th89+ line.
This lead to new ring settings (data with all the settings will be put in tomorrow).
|
Tue Apr 5 20:58:53 2022, Wilfried, Laser, Trigger for Laser
|
Electronic for the timing of the laser puls optimized. The problem with the broad photodiode histograms is solved.
The photo of the oscilloscope picture shows in magenta the pulses from the (discriminated RF)/2, in yellow the pulses that are generated to open the Q-Switch.
The latter are generated by the coincidence of the 30-Hz trigger (2 microsec long) to
|
Tue Apr 5 20:16:13 2022, Rodolfo, Laser, Discriminator for HF reference signal
|
For the laser trigger we need to generate a few ns NIM-puls from the HF reference signal. Therefore we (Wilfried, Konstantin and me) have changed the constant
fraction descriminator (CFD) to generate this kind of pulse.
|
Tue Apr 5 15:25:47 2022, Sebastian, Kristian, Pierre-Michel, DAQ, Go4 Conditions for "TDC_stacked_Sig/Ref" histogram
|
First version of TDC_stacked histograms implemented.
Under folder "Conditions" you can now integrate over user defined Limits (Xmin and Xmax). |
Tue Apr 5 14:51:51 2022, Kristian, DAQ, dark counts
|
We shielded the PMTs, i.e., the N2 tube, which improved the Cu PMT background rate.
Without beam and ESR lights off:
|
Tue Apr 5 13:28:06 2022, Kristian, DAQ, Spikes in PMT signal
|
Occasionally, we noticed a spike in the PMT signal. No ions in the ring and at PMT Cu was not even supplied with HV. Has to be some electronic pickup |
Tue Apr 5 13:24:40 2022, Kristian, Detectors, PMT Cu
|
HV cable was in wrong channel in power supply. Now it is back to the correct position. Checked connection at the PMT. |
Tue Apr 5 12:17:12 2022, Ragan, Rui-Jiu, DAQ, Instructions to start and stop NTCAP
|
Attached are the instructions to start and stop data taking using the NTCAP. |
Tue Apr 5 11:14:52 2022, Sebastian, Kristian, Pierre-Michel, DAQ, PMT over wavelength histogram v2
|
Today we finalized the PMT over wavelength histogram in:
Analysis/Histograms/Step2_Anl/LabView/laserscan
|
Tue Apr 5 09:45:47 2022, Danyal & Yuri, Accelerator, beam is back in the ESR - a slit was moved in
|
And now also in English :)
The ion beam is back in the ESR!
|
Tue Apr 5 09:39:01 2022, Danyal & Yuri, Accelerator, Strahl ist wieder im ESR - Schlitz war reingefahren
|
Der Strahl ist wieder im ESR!
Angeblich war einen Schlitz, oder sogar 2, reingefahren worden.
Es könnte entweder der GEEXDR11S und/oder den GEEXDR1AS gewesen sein.
|
Tue Apr 5 08:00:50 2022, Kristian & Sebastian, DAQ, MBS Channellist (so far)
|
MBS Channellist (time 5.4.2022)
0 Anode
|
Mon Apr 4 19:05:57 2022, Carsten, DAQ, Cooler Frequency Signals Ion current not connected until now
|
Up to now the signals for
Ion Current
Cooler elctron current
|
Mon Apr 4 17:26:48 2022, Jan Glorius, DAQ, MSCF shaper settings
|
attached are the current shaper settings.
list of currently connected channels:
|
Mon Apr 4 16:26:04 2022, Carsten, DAQ, Backup Harddisk /External Disk
|
Since we cannot write on lustre, as a temproal solution two RAID 1 (mirror) USB hard disks with cpacity 6 TB each are installed.
The disk on the lxg1297 the disk can be accessd at /meida/sdd2
an on the lxg0155 at /media/sdc2
|
Mon Apr 4 15:32:38 2022, Kristiand & Sebastian, DAQ, Go4 Backup Code
|
Due to some code changes (PMT over wavelength historgram) Kristian and I backuped a working Go4 code in the current go4 folder on the atplaser linux account.
It can be found in the subfolder "working go4 code 4.4.2020"
|
Mon Apr 4 15:27:31 2022, Sebastian, DAQ, PMT over wavelength histogram
|
Kristian and I tried to create a "PMT over wavelength" histogram under step2Anl/LabView.
Unfortunately, the code ist still in progress and does not work yet.
|
Mon Apr 4 09:10:44 2022, Ragan, Yuri, Rui-Jiu, Runs, Measurement
|
NTCAP is taking data with the beam with a stop at manipulation 5 (taking all the fragments).
1)
|
Mon Apr 4 00:19:16 2022, Ken Ueberholz, DAQ, FirstTests
|
New measurement:
Start: About 00:12
Stop: About 00:43
|
Sun Apr 3 18:52:30 2022, Konstantin, Calibration, Laser ion beam overlap
|
We measured vertical and horizontal ion beam positions. Unfortunately, the horizontal scraper positions can not be manipulated
by the "Device Control" software. Robert Boywitt has supported us a lot to get the drives moving by using the expert software on
a local control PC, which can only be remotely accessed from beapc032 with the usual experiment account "atplaser". In the first
|
Sun Apr 3 18:41:11 2022, Konstantin and Rodolfo, Laser, Beam Position Stabilization
|
We (Konstantin and I) have optimized the TEM "laser-beam position stabilization".
- I took some pictures (screenshots) about the current setting*.
|
Sun Apr 3 15:34:34 2022, Ken Ueberholz , , Xuv-Detektor CFD-treshhold
|
I co controlled all detector signals and
adjusted the threshold to be just above
the noise. I will double check this when
|
Sun Apr 3 13:05:08 2022, Carsten, Laser, Distance of Partilce detecotrs wr.t. to electron cooler
|
As of Yuris post in the elog the positions (Info from H. Weick):
The postion of the particle counter behind the cooler is 21.6 m behind the middle of the cooler.
|
Sun Apr 3 12:35:17 2022, Max, DAQ, Go 4 manual
|
How to start go4?:
- go to console
|
Sun Apr 3 12:31:10 2022, Rui-Jiu Chen, Yuri, Carsten, Ragan, Runs, Measurement of beam with different circles 6x
|
Parameters of NTCAP:
1. IQ rate 20 MSamples/s
|
Sun Apr 3 11:40:48 2022, Shahab, Yuri, Detectors, 245 MHz Resonator tuned to new frequency
|
resonance frequency changed from 243.2 MHz to 244.692 MHz by reducing the value of both of the plungers from 395 to 325. Now after movement, the peak is
on the resonance. |
Sun Apr 3 11:01:57 2022, Rodolfo, Accelerator, Scrapers @ E-Cooler
|
We the mid- and night shift (Wilfried, Simon, Patrick, Ken and me) we not able to manipulate the scrapers in the E-Cooler.
- Carsten contacted Markus concerning this issue.
|
Sun Apr 3 10:35:20 2022, Yuri (Sergey), Accelerator, Dipole correction
|
The ramp for moving the beam to the central orbit
To be inserted into manipulation, GE01MU1 (SC 5): 1.05969755 rad
This can be done only when one stays in the manipulation 5
|
Sun Apr 3 10:28:25 2022, Yuri (Helmut Weick), Accelerator, Distance cooler - detector
|
Calculation from Helmut Weick
Distance e-cooler to particle detectors chamber:
“This is half of ESR plus distance target - detector. This even is on the web page”
|
Sun Apr 3 00:41:45 2022, Ken Ueberholz , , Xuv-Detektor Signale
|
The Xuv-detector cabel channels seem to be
wrongly labeled in the Messhütte (see
picture) and are propably also wrongly
|
Sat Apr 2 23:16:55 2022, Carsten, , Cu mirror and CsI moved in and out of the ring
|
Ken tested the Cu-mirror and CsI disk Detector. They move without a problem and don't scrape the ion beam. |
Sat Apr 2 22:43:36 2022, Carsten, , Ring set-up for 229/89+, scraped and bunched, Iel=500mA at injection and 100mA during measurment
|
The ring SIS/ESR were set up for 229/89+. Pölease note that this not only 1e- Th but ponteially also the isobars of Ac (0e-1)/Pa(2e-, strongest) and U(3e-).
SIS eenrgy plus 3.5 MeV/u w.r.t. to the 238U92+ settings. We nearly don't see any of the primary beam in the ring
Cooling energy is 208998 KeV (read from the ESR multimeter).
|
Sat Apr 2 18:28:47 2022, Markus/Bernd, Accelerator, SIS setting complete for fragments in ESR: 2022-04-02_18-09-10_SIS18_FAST_E142_20220325_110213.C1.txt
|
# Version: 1.2 # Comment: ParamModi Export at 2022-04-02T18:09:36.402593+02:00[Europe/Berlin] # Context: SIS18_FAST_E142_20220325_110213.C1 # Columns: PARAMETERNAME,VALUE,UNIT,BEAMPROCESSPURPOSE
LOGICAL.GTK7MU5/HKICK,0.0,rad,TRANSFER_INJECTION LOGICAL.GTK7MU5/HKICK_THEO,0.0018,rad,TRANSFER_INJECTION LOGICAL.GTK7BC1L/HKICK,0.0,rad,TRANSFER_INJECTION |
Sat Apr 2 18:19:47 2022, Yuri, Accelerator, Scraping
|
|
Sat Apr 2 18:17:31 2022, Yuri, Accelerator, Photo-Identification
|
Just a photo of the ID |
Sat Apr 2 17:47:16 2022, Ken Ueberholz , , Telefon, Gästehaus
|
My guesthouse phone number: 06159-71-2957 |
Sat Apr 2 16:10:22 2022, Markus, Bernd, Accelerator, SIS setting complete, U90+ mit Be target: 2022-04-02_16-04-00_SIS18_FAST_E142_20220325_110213.C1.txt
|
# Version: 1.2
# Comment: ParamModi Export at 2022-04-02T16:04:10.778077+02:00[Europe/Berlin]
# Context: SIS18_FAST_E142_20220325_110213.C1
|
Sat Apr 2 15:54:32 2022, Carsten, Accelerator, ESR setup with primary beam and 1cm Be plate + Schottky data on NTCAP
|
MArkus and Bernd have setup the ESR with the thick stripper foil and put the ion beam (238U92+) on the central orbit.
SIS Energy is 536 MeV/u
|
Sat Apr 2 10:22:21 2022, Rui-Jiu Chen, Ragan, General, Beam is in the ESR
|
Primary beam is in the ESR now. Markus and the team are tuning the ESR. |
Sat Apr 2 05:08:00 2022, Kristian, General, Morning shift
|
I see no updates in the log. Is the morning shift taking place?
For now I will stay at home and be on call. I can be at GSI in 20 min. Please let me know the status.
|
Sat Apr 2 01:27:25 2022, Jan Glorius, Runs, test runs 02.04.2022
|
Beam settings:
primary beam
copper stripper
|
Sat Apr 2 01:02:39 2022, Carsten, , Test measurements with two DAQs
|
We do two measurements:
One measurement using the |
Fri Apr 1 23:47:45 2022, Carsten, Detectors, PMT thresholds / PMT north does not allow voltage abow ~ -1kV
|
Checked thresholds of PMT with beam and without beam.
The [B]PMT "south"[/B] which is now at the gas-jet target has burst-like pulses with a train of many short pulses of duration of about 1us. The signals |
Fri Apr 1 23:12:37 2022, Patrick, Detectors, Scraper positions "only primary beam"
|
Scraper positions where there is only primary beam for detector tests. Use only one outer and one inner scraper at the same time.
[B]Move detectors out before new beam![/B]
|
Wed Mar 23 11:20:55 2022, Simon, Detectors, PMT operating voltages
|
During the testing of the PMTs the following operating voltages were applied:
PMT Cu: -1300V
|
Fri Apr 1 22:01:14 2022, Konstantin, Detectors, PMT operating voltages
|
> During the testing of the PMTs the following operating voltages were applied:
>
> PMT Cu: -1300V
|
Fri Apr 1 18:30:38 2022, Carsten, Detectors, Positions / Voltages threshlds for particle detectors
|
[B]DC beam 238U92+ (205kV, e-beam)[/B]
We used the detector NA_SZ to determine the central primary-beam (238U92+) position by observing the Schottky-signal while moving the detector inside.
|
Fri Apr 1 18:19:58 2022, Kristian, Calibration, HV divider
|
We checked the strange behavior of the HV dividers again. The Ohmlabs and the PTB divider are in good agreement at 120 kV and 150 kV. There, both dividers
deviated by 0.35% from the value on the HKR multimeter. Ohmlabs divider yields higher voltage.
|
Fri Apr 1 16:46:33 2022, Carsten, Detectors, All pressuzied air drives are tested now and are working properly
|
All pressuzied air drives are tested now and are working properly |
Fri Apr 1 15:52:23 2022, Rodolfo, Detectors, UV PMT South
|
The "Solar Blind PMT" labeled as "PMT South" has been installed by Simon and Max in the gas target chamber on a CaF Viewport. The PMT is pointing directly
to the center chamber.
|
Fri Apr 1 15:27:00 2022, Rodolfo, Detectors, Nitrogen for UV-PMTs
|
There was a question concerning the nitrogen we use to rinse the "Solar Blind PMTs".
- I have checked the nitrogen flow rate this afternoon (14:00) and it was a level of 0.2 LPM.
|
Fri Apr 1 11:19:57 2022, Rui-Jiu Chen, DAQ, S.M.A.R.T check on NTCAP
|
S.M.A.R.T. check has been done on NTCAP on March 28th. All 12 hard disk drivers are normal. |
Fri Apr 1 11:00:41 2022, Rui-Jiu Chen, DAQ, Start NTCAP
|
Start taking data with NTCAP.
Carrier Frequency: 245MHz
IQ rate:5 MSamples/s
|
Fri Apr 1 02:29:40 2022, Carsten, , ESR Cooler crashed
|
ESR cooler crashed. We could not restart the cooler. Since we had several other problems, we stop now. |
Fri Apr 1 01:57:12 2022, Kristian, DAQ, Timing
|
with bunched primary beam (bunching frequency 1.94070 MHz) we have strong PMT (mid) and particle detector signal. Time between both detectors was measured
(in 3.3ns bins)
|
Fri Apr 1 00:27:27 2022, Carsten, , Settings Particle Detectors /Tests with primary beam
|
All the beam settings will be given in a separate post.
We have 238U92+ centered in the ring with typically a few times 10⁷ ions.
|
Thu Mar 31 22:41:35 2022, Kristian, Calibration, HV divider
|
We noticed a difference in the electron cooler high voltage:
The set voltage is ~205 kV, the Ohmlabs divider measures 207 kV and the PTB200 shows a value of ~212 kV. The latter is doing some crazy jumps of up to |
Thu Mar 31 22:19:14 2022, Kristian, DAQ, How To Start MBS
|
(MBS server should be running at the LXG1297 machine. If not you will get an error message. Then start the server - last command in console)
Just enter the following commands in the console of any PC
|
Thu Mar 31 20:07:47 2022, Shahab, DAQ, Schottky Display in AP Messhütte (or anywhere else)
|
[B]UPDATE: 16.05.2022[/B]
[I]Please use ATPPC029 instead of ATPPC019 for the 245 MHz Schottky
|
Thu Mar 31 17:56:32 2022, Konstantin, DAQ, Photodiode test signal
|
We used a signal generator to provide a 4MHz dummy-RF signal. This was divided by a factor of 2 for laser-"bunch" synchronization and by a factor of 200
for to provide the stop-tigger attached to the DAQ. With this Dummy-RF we were able to measure the relative arrival times of the laser at the photodiodes
located in the laser towers. Note that the signal of the southern photodiode arrives first. This is because the scale of the time axis is given in (10/3ns)-bins |
Wed Mar 23 16:16:42 2022, V. Hannen, Detectors, Control software for XUV and copper mirror detector
|
Attached you find a manual for the software used to control the movement of the XUV detector and the copper mirror detector.
The software is also used to set the electrode voltages and coil currents of the XUV detector. |
Thu Mar 31 15:53:50 2022, Konstantin, Detectors, Control software for XUV and copper mirror detector
|
> Attached you find a manual for the software used to control the movement of the XUV detector and the copper mirror detector.
> The software is also used to set the electrode voltages and coil currents of the XUV detector.
|
Thu Mar 31 13:28:17 2022, Jan Glorius, DAQ, DAQ-2 cabling docu
|
newest version V2 - valid since Mo 04.April 2022 |
Thu Mar 31 11:49:54 2022, Konstantin, Detectors, HV-Supply of particle detectors
|
We cross-checked the cabeling of the particle detectors north and south. Particle detector south is connected to channel [B]HV5[/B] at the yellow "ESR Messplatz
III" patch-panel. Particle detector north is connected to channel [B]K5[/B] at the green "ESR Messplatz II" patch-panel. Red high voltage power-supply
is used to control the voltage. |
Wed Mar 30 09:27:15 2022, Rodolfo, Accelerator, Issue? Scrapers at E-Cooler
|
Carsten* forwarded me an E-Mail from Markus yesterday. "Betreff: Scraper im ESR Elektronenkühler"
- Markus has tried to get moving the scrapers at E-Cooler last week. But he got some issues.
|
Tue Mar 29 21:37:13 2022, Rodolfo, Laser, Pump Laser
|
The gold "reflectors and end plates" of our pump laser (Quanta-Ray PRO) have been exchanged by new ones.
- Output power after Oscillator and Amplifier (@ 1064 nm): 41 W.
|
Tue Mar 29 12:47:26 2022, Ken Ueberholz, , Parameter setting for XUV and copper mirror detector
|
Some Default Settings for the XUV
Detector. The copper
mirror should always move fully in.
|
Tue Mar 29 12:01:08 2022, Rodolfo, Laser, Laser targets
|
I have installed two targets* (size: 2 inches) to monitor the position of the laser beam.
- The internet cameras can be accessed from any browser, but only inside the GSI-Campus network. Benutzername: admin. Kennwort: 1234
|
Fri Mar 25 14:31:24 2022, Rodolfo, Laser, Photodiode signal SO in the Messhütte
|
Measured at the Panel "Messhütte - Laserhütte", BNC "Fast PD / SO".
- Average over 256 samples |
Thu Mar 24 17:10:33 2022, Carsten, DAQ, DAQ / MBS VUPROM TDC buffer errors
|
VUPROM TDC buffer errors potentially due to long TDC common Stop gate (>1ms from div 200).
Signal piped through coinc. unit gicv es shorter sginal (~200ns). This signal used as common stop seems to have cured the TDC buffer errors.. |
Thu Mar 24 16:59:35 2022, Rodolfo, Laser, Photodiode signal SO
|
Fast Photodiode signal at the "SO-Bogen" (South-east arc).
- Average made over 250 samples.
|
Thu Mar 24 13:27:47 2022, Rodolfo, DAQ, Disk space - LXG1297
|
In our Linux computer (LXG1297: Laser spectroscopy) we have a Drive with 1.8 TB total disk space. 300 GB are at the moment in use, but I can clean this.
|
Wed Mar 23 16:58:33 2022, V. Hannen, Detectors, Status copper mirror and XUV detectors
|
Over the past days, the parabolic copper mirror and XUV detectors haven been prepared and tested for the upcoming beamtime (see first attached photo for
an overview of the detectors).
|
Wed Mar 23 15:38:15 2022, Carsten, General, Not all lights in the ESR can be switched off
|
Three of the lights in the ESR cannot be switched off. It seems that these lights are now "emergency lights" (with battery?). We will need to find dark
plastik foilor similar to cover those lights during our beam time |
Mon Mar 21 21:48:46 2022, Jan, DAQ, MesyTec Shaper Control
|
The MesyTec Shaper & Constant Fraction (MSCF) is used to process the PMT/det signals for the ADC in the second DAQ.
It can be setup on its front panel, but single channel setup is very tedious.
Remote control of the settings is possible using an epics server and client.
|
Mon Mar 21 16:33:23 2022, , DAQ, Changing code of go4
|
We agreed to indicate changes in the go4 analysis code according to the
following scheme
|
Mon Mar 21 17:03:47 2022, , DAQ, Changing code of go4
|
Correction: Comments in C/C++ start with // , i.e. forward slash
[B]//--2022-mm-dd HH:MM comment on the respective change[/B].
|
Mon Mar 21 10:46:18 2022, Rodolfo, DAQ, TDC & Scaler description channels
|
For the First Vuprom we have the following channel configuration (Stand:21.03.2021).
Please note 1.
|
Fri Mar 18 16:58:26 2022, Rodolfo, Accelerator, Pressure at North-East Arc
|
The pressure at the north-east arc of ESR, where the laser window is located, looks suspicious. The sensor E01LF02 shows a value of 7.26E-8 mbar. For comparison
I have also upload a photo taken on 07.02.2022 where the pressure was 4.62E-10. We have contacted the vacuum expert and asked for his opinion. Waiting
for the reply ...
|
Mon Mar 21 09:18:28 2022, Rodolfo, Accelerator, Pressure at North-East Arc
|
Our vacuum expert has reply. The "IZ Pump" in this region was off.
[quote]The pressure at the north-east arc of ESR, where the laser window is located, looks suspicious. The sensor E01LF02 shows a value of 7.26E-8 mbar. |
Fri Mar 18 16:51:27 2022, Carsten, DAQ, Main DAQ /MBS error
|
MBS error "TDC data ready wait" solved. Common stop trigger (20 kHz) was missing. (according to Sebastian) |
Fri Mar 18 09:18:38 2022, Rodolfo, Accelerator, Vacuum leak, ODR
|
After the beaking out process we got a vacuum leak in one of the MgF2 windows (Referenz: PMT South) of the optical detection region (ODR) Stand 17.03.2022.
In order to avoid breaking the vacuum again and a second beaking out process we had to spread vacuum seal on the window.
|
Thu Mar 17 17:56:33 2022, Kristian, Calibration, HV dividers
|
PTB HV dividers arrived. Divider ratios are:
PT120: 12003,341
|
Thu Mar 17 17:37:49 2022, Rodolfo, Accelerator, High voltage dividers for E-Cooler
|
Two dividers, from PTB-Braunschweig, have been deliver today. We will use them to monitor the electron-cooler voltage in situ.
- PT120 (up to -120 kV) with scaling factor: 12003.341, and
|
Wed Mar 16 12:04:40 2022, Kristian, Detectors, HV cable to particle detector South
|
Checked HV cable to particle detector south. Label is HV2 Scraper.
Connected to Messplatz III HV5 |
Wed Mar 16 08:31:52 2022, Kristian, DAQ, Check cables and timing
|
Konstantin and I generated a signal in the Messhütte and sent it to the ESR.
There, we used a 44m BNC cable that was sufficently long to reach all detectors.
Our signal was connected instead of the PMT/PD. Hence, it was a roundtrip measurement.
|
Thu Mar 10 15:20:26 2022, Jan , DAQ, analog branch signal patches to Messhütte
|
The additional signals from PMTs and XUV detector needed for spectroscopic analysis in the second DAQ are patched to AP Messhütte (MH) via the red patch
panel at the inside of the target region. The patch channels F1 to F8 (also labeled A to H in red on white) are used and extended by a cable bundle to
reach below the mirror chamber. Inside the MH the signals arrive at the red patch panel and are forwarded to the blue patch panel close to the laser DAQ |
Thu Mar 3 21:39:21 2022, Rodolfo, Laser, Dye solution, DCM in Ethanol and DMSO
|
I have prepared a new dye solution of DCM in Ethanol and DMSO (1:1). The concentration of the "oscillator + preamplifier" is 0,3 g/l. The concentration
of the "main amplifier" is 0,0375 g/l or "1/8" the concentration of the "oscillator".
|
Thu Mar 3 17:34:42 2022, Rodolfo, Laser, Pump Laser
|
We have issues with our Pump Laser (Spectra Physics - Quanta Ray Pro). After the service made on mid Dezember (2021) the laser does not get started. After
contacting the service many times, the technicians think it can be an issue with the flashlamps.
|
Thu Mar 3 17:21:26 2022, Rodolfo, Detectors, MgF2 Windows
|
New reentrance viewports made of MgF2 have been installed at the ESR "Optical Detection Region". In the middle the old window (also MgF2) was kept into
place. |
Thu Mar 3 13:50:45 2022, Rodolfo, Laser, Dye cell cracket
|
The "20 mm" dye cell used for the Oscillator and Pramplifier got damage. I supposed it was produced by the pump beam hitting on a very "old" dye solution
(DCM / Ethanol, concentration 0,3 g/l). The solution in the dye reservoir was more 12 months old. |
Fri Nov 19 15:50:44 2021, Rodolfo, Accelerator, Laser windows at E-Cooler side
|
New laser viewports have been installed at the E-Cooler side of the ESR.
[IMG]elog:10/1[/IMG]
|
Fri Nov 5 17:23:23 2021, Rodolfo, DAQ, Dye trigger out
|
The dye laser offers different scan settings (see pages 65 - 67 of the user manual). We are using the following setting, see attached screenshot, to produce
the status "laser busy" for the DAQ.
|
Thu Oct 14 10:21:46 2021, Rodolfo, Detectors, Dismounting Cu-Mirror
|
We have dismounted the parabolic Cu-Mirror at ESR on 13.10.2021 |
Wed Sep 8 21:44:55 2021, Rodolfo, General, Preparation
|
[TABLE border="1"]
[B]Status[/B]|[B]Arbeitspaket[/B]|[COLOR=chocolate][B]Kommentare[/B][/COLOR]
|-
|
Fri Aug 6 10:40:54 2021, Rodolfo, DAQ, mbs in atplaser
|
The "MBS program" (data acquisition) is now located within atplaser@r4l-41, in the folder "th21":
|
Fri Aug 6 10:23:00 2021, Rodolfo, DAQ, atplaser on r4l-41
|
The user [B]atplaser[/B] is now existing on the [B]R4L-XY[/B] nodes for the ESR group.
R4 = [B]R[/B]IO[B]4[/B] Computer
|
Mon Aug 2 10:54:44 2021, Rodolfo, Laser, Quanta Ray - Water leak in the power supply
|
After fixing the broken water connector in the laser head a new water leak appeared in the Quanta-Ray power supply.
We found the leak in the water pump flow line (arrow up). We took the water connectors appart, rebuilt the teflon seeling and put the water connectors |
Wed Jul 28 13:31:22 2021, Rodolfo, General, Air filters
|
Filters (Hepa-) inside the flow boxes have been replaced. |
Mon Jul 26 17:58:52 2021, Rodolfo, Laser, Water Filters
|
The water filters (deioninzing and particle) have been replaced. |
Mon Jul 26 17:15:47 2021, Rodolfo, Laser, Quanta Ray - Water leak
|
On Thursday afternoon (22.07.2021) a water connector has broken and water has been spilled in the laser head. |